Guidelines for power and time variables for microwave ablation in porcine lung in vitro.

Hongchao Cai, Guanglian Shan, Zhigang Wei, Wenhua Zhao, Guoliang Xue, Chao Zhang, Xin Ye
{"title":"Guidelines for power and time variables for microwave ablation in porcine lung in vitro.","authors":"Hongchao Cai, Guanglian Shan, Zhigang Wei, Wenhua Zhao, Guoliang Xue, Chao Zhang, Xin Ye","doi":"10.4103/jcrt.jcrt_817_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Determination of the appropriate ablative parameters is the key to the success and safety of microwave ablation (MWA) of lung tumors. The purpose of this study was to provide guidelines and recommendations for the optimal time and power for lung tumor MWA.</p><p><strong>Material and methods: </strong>MWA using a 2450-MHz system was evaluated in a porcine lung. The independent variables were power (30, 40, 50, 60, 70, and 80 W) and time (2, 4, 6, 8, 10, and 12 min), and the outcome variable was the volume of ablation. Lung tissues were procured after MWA for measurement and histological evaluation. Analysis of variance was used for statistical analysis, followed by least significant difference (LSD) t-tests where appropriate. A P value of <0.05 was considered statistically significant.</p><p><strong>Results: </strong>The outcome variable (ablative volume) was significantly affected by time, power, and time/power interaction (P < 0.05). When the total output energy was kept constant, the combination of higher power and shorter time obtained a larger ablative volume, especially in the low- and medium-energy groups (P < 0.01).</p><p><strong>Conclusions: </strong>We propose guidelines for ablative volume based on different time and power variables to provide a reference for clinical applications.</p>","PeriodicalId":94070,"journal":{"name":"Journal of cancer research and therapeutics","volume":"20 4","pages":"1150-1156"},"PeriodicalIF":0.0000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of cancer research and therapeutics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jcrt.jcrt_817_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/8/29 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Determination of the appropriate ablative parameters is the key to the success and safety of microwave ablation (MWA) of lung tumors. The purpose of this study was to provide guidelines and recommendations for the optimal time and power for lung tumor MWA.

Material and methods: MWA using a 2450-MHz system was evaluated in a porcine lung. The independent variables were power (30, 40, 50, 60, 70, and 80 W) and time (2, 4, 6, 8, 10, and 12 min), and the outcome variable was the volume of ablation. Lung tissues were procured after MWA for measurement and histological evaluation. Analysis of variance was used for statistical analysis, followed by least significant difference (LSD) t-tests where appropriate. A P value of <0.05 was considered statistically significant.

Results: The outcome variable (ablative volume) was significantly affected by time, power, and time/power interaction (P < 0.05). When the total output energy was kept constant, the combination of higher power and shorter time obtained a larger ablative volume, especially in the low- and medium-energy groups (P < 0.01).

Conclusions: We propose guidelines for ablative volume based on different time and power variables to provide a reference for clinical applications.

猪肺体外微波消融的功率和时间变量指南。
目的:确定适当的消融参数是肺部肿瘤微波消融(MWA)成功和安全的关键。本研究旨在为肺部肿瘤微波消融的最佳时间和功率提供指导和建议:在猪肺中评估了使用 2450-MHz 系统的微波消融。自变量为功率(30、40、50、60、70 和 80 W)和时间(2、4、6、8、10 和 12 分钟),结果变量为消融体积。MWA 后采集肺组织进行测量和组织学评估。统计分析采用方差分析,适当时采用最小显著性差异(LSD)t 检验。结果的 P 值为 0:结果变量(烧蚀体积)受时间、功率和时间/功率交互作用的显著影响(P < 0.05)。当总输出能量保持不变时,较高功率和较短时间的组合可获得较大的烧蚀体积,尤其是在低能量和中等能量组(P < 0.01):我们提出了基于不同时间和功率变量的烧蚀体积指南,为临床应用提供参考。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信