A compact ion source combining electron-impact and thermal ionization for multiple-reflection time-of-flight mass spectrometry.

IF 1.3 4区 工程技术 Q3 INSTRUMENTS & INSTRUMENTATION
Jiajun Yu, Ali Mollaebrahimi, Samuel Ayet San Andrés, Timo Dickel, Wolfgang R Plaß, Heinrich Wilsenach, Sönke Beck, Zhuang Ge, Hans Geissel, Christine Hornung, Andrew Jacobs, Gabriella Kripko-Koncz, Anna A Kwiatkowski, Meetika Narang, Christoph Scheidenberger, Jaden Sequeira, Coulter Walls
{"title":"A compact ion source combining electron-impact and thermal ionization for multiple-reflection time-of-flight mass spectrometry.","authors":"Jiajun Yu, Ali Mollaebrahimi, Samuel Ayet San Andrés, Timo Dickel, Wolfgang R Plaß, Heinrich Wilsenach, Sönke Beck, Zhuang Ge, Hans Geissel, Christine Hornung, Andrew Jacobs, Gabriella Kripko-Koncz, Anna A Kwiatkowski, Meetika Narang, Christoph Scheidenberger, Jaden Sequeira, Coulter Walls","doi":"10.1063/5.0213443","DOIUrl":null,"url":null,"abstract":"<p><p>A compact ion source combining electron impact and thermal ionization has been developed and commissioned in two Multiple-Reflection Time-Of-Flight Mass Spectrometer (MR-TOF-MS) setups at the Fragment Separator Ion Catcher at the GSI Helmholtz Centre for Heavy Ion Research, Darmstadt, Germany, and at TRIUMF's Ion Trap for Atomic and Nuclear science at TRIUMF Canada's particle accelerator center, Vancouver, Canada. The ion source is notable for its compact dimensions of 50 mm in height and 68 mm in diameter. The ion source is currently in daily operation at both facilities. Design, simulations, and results of combining ions from thermal and electron-impact ionization of different gases (perfluoropropane and sulfur hexafluoride) are presented in this work. The systematic effects of heating power on the thermal source were studied in detail. The source has demonstrated stable and long-term production of reference ions over a wide mass range for the MR-TOF-MS. This versatile ion source has also been used to optimize and investigate the transport of ions with different chemical reactivity and ionization potentials.</p>","PeriodicalId":21111,"journal":{"name":"Review of Scientific Instruments","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Review of Scientific Instruments","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1063/5.0213443","RegionNum":4,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

Abstract

A compact ion source combining electron impact and thermal ionization has been developed and commissioned in two Multiple-Reflection Time-Of-Flight Mass Spectrometer (MR-TOF-MS) setups at the Fragment Separator Ion Catcher at the GSI Helmholtz Centre for Heavy Ion Research, Darmstadt, Germany, and at TRIUMF's Ion Trap for Atomic and Nuclear science at TRIUMF Canada's particle accelerator center, Vancouver, Canada. The ion source is notable for its compact dimensions of 50 mm in height and 68 mm in diameter. The ion source is currently in daily operation at both facilities. Design, simulations, and results of combining ions from thermal and electron-impact ionization of different gases (perfluoropropane and sulfur hexafluoride) are presented in this work. The systematic effects of heating power on the thermal source were studied in detail. The source has demonstrated stable and long-term production of reference ions over a wide mass range for the MR-TOF-MS. This versatile ion source has also been used to optimize and investigate the transport of ions with different chemical reactivity and ionization potentials.

用于多次反射飞行时间质谱分析的电子撞击和热电离相结合的紧凑型离子源。
在德国达姆施塔特 GSI 赫尔姆霍兹重离子研究中心的碎片分离器离子捕获器和加拿大温哥华 TRIUMF 粒子加速器中心的 TRIUMF 原子与核科学离子阱的两个多反射飞行时间质谱仪(MR-TOF-MS)装置中,开发并试运行了结合电子撞击和热电离的紧凑型离子源。该离子源以其高度 50 毫米、直径 68 毫米的紧凑尺寸而著称。该离子源目前在这两个设施中进行日常运行。这项工作介绍了将不同气体(全氟丙烷和六氟化硫)的热电离和电子撞击电离产生的离子结合起来的设计、模拟和结果。详细研究了加热功率对热源的系统影响。该离子源已证明能在很宽的质量范围内为 MR-TOF-MS 长期稳定地产生参考离子。这种多功能离子源还被用于优化和研究具有不同化学反应性和电离电位的离子的传输。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Review of Scientific Instruments
Review of Scientific Instruments 工程技术-物理:应用
CiteScore
3.00
自引率
12.50%
发文量
758
审稿时长
2.6 months
期刊介绍: Review of Scientific Instruments, is committed to the publication of advances in scientific instruments, apparatuses, and techniques. RSI seeks to meet the needs of engineers and scientists in physics, chemistry, and the life sciences.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信