Roland Schwab, Christoph Kabbasch, Lukas Goertz, Marius Kaschner, Daniel Weiss, Christian Loehr, Hauke Wensing, Maxim Bester, Andreas Simgen, André Kemmling, Christina Wendl, Erelle Fuchs, Maximilian Thormann, Daniel Behme, Hannes Nordmeyer
{"title":"The DERIVO 2 Heal Embolization Device in the Treatment of Ruptured and Unruptured Intracranial Aneurysms: a Retrospective Multicenter Study.","authors":"Roland Schwab, Christoph Kabbasch, Lukas Goertz, Marius Kaschner, Daniel Weiss, Christian Loehr, Hauke Wensing, Maxim Bester, Andreas Simgen, André Kemmling, Christina Wendl, Erelle Fuchs, Maximilian Thormann, Daniel Behme, Hannes Nordmeyer","doi":"10.1007/s00062-024-01446-8","DOIUrl":null,"url":null,"abstract":"<p><strong>Backround: </strong>The use of flow diverting stents in the treatment of intracranial aneurysms is associated with a risk of neurological morbidity due to their thrombogenicity. To reduce this risk different surface modifications have been developed. The Derivo 2 Embolization Device (Acandis, Pforzheim, Germany) has proven to be a safe and effective flow diverter. To overcome the risk of thrombo-embolism, the device was modified by adding an anti-thrombogenic fibrin-heparin coating. We aimed to assess the safety and effectiveness of the Derivo 2 heal Embolization Device.</p><p><strong>Methods: </strong>Retrospective multicenter data from nine German neurovascular centers between February 2022 until December 2023 were used. Patients treated with the Derivo 2 heal Embolization Device for unruptured or ruptured intracranial aneurysms were included. Peri- and postprocedural adverse events, clinical outcomes, and angiographic follow-up results were evaluated.</p><p><strong>Results: </strong>84 patients (73.8% female; mean age 58.7 years) with 89 aneurysms (mean size 9.8 mm) were included. 87.6% were located in the anterior circulation. Most of them were sidewall aneurysms (88.8%). 96 flow diverters were used. 99.0% were successfully implanted. An in-stent balloon angioplasty was performed in 6.0% of the cases. An additional coiling was performed in 28.6%. Technical difficulties were present in 12.0% of the cases. Thrombotic events occurred in 4.8% with no neurological sequelae. Mortality and morbidity were 0 and 1.2% respectively. Adequate aneurysm occlusion was achieved in 80.7% with a mean follow-up time of 6.6 months.</p><p><strong>Conclusion: </strong>The Derivo 2 heal Embolization Device showed a satisfying aneurysm occlusion and safety with a low rate of neurological morbidity.</p>","PeriodicalId":49298,"journal":{"name":"Clinical Neuroradiology","volume":" ","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2024-08-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Neuroradiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00062-024-01446-8","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Backround: The use of flow diverting stents in the treatment of intracranial aneurysms is associated with a risk of neurological morbidity due to their thrombogenicity. To reduce this risk different surface modifications have been developed. The Derivo 2 Embolization Device (Acandis, Pforzheim, Germany) has proven to be a safe and effective flow diverter. To overcome the risk of thrombo-embolism, the device was modified by adding an anti-thrombogenic fibrin-heparin coating. We aimed to assess the safety and effectiveness of the Derivo 2 heal Embolization Device.
Methods: Retrospective multicenter data from nine German neurovascular centers between February 2022 until December 2023 were used. Patients treated with the Derivo 2 heal Embolization Device for unruptured or ruptured intracranial aneurysms were included. Peri- and postprocedural adverse events, clinical outcomes, and angiographic follow-up results were evaluated.
Results: 84 patients (73.8% female; mean age 58.7 years) with 89 aneurysms (mean size 9.8 mm) were included. 87.6% were located in the anterior circulation. Most of them were sidewall aneurysms (88.8%). 96 flow diverters were used. 99.0% were successfully implanted. An in-stent balloon angioplasty was performed in 6.0% of the cases. An additional coiling was performed in 28.6%. Technical difficulties were present in 12.0% of the cases. Thrombotic events occurred in 4.8% with no neurological sequelae. Mortality and morbidity were 0 and 1.2% respectively. Adequate aneurysm occlusion was achieved in 80.7% with a mean follow-up time of 6.6 months.
Conclusion: The Derivo 2 heal Embolization Device showed a satisfying aneurysm occlusion and safety with a low rate of neurological morbidity.
期刊介绍:
Clinical Neuroradiology provides current information, original contributions, and reviews in the field of neuroradiology. An interdisciplinary approach is accomplished by diagnostic and therapeutic contributions related to associated subjects.
The international coverage and relevance of the journal is underlined by its being the official journal of the German, Swiss, and Austrian Societies of Neuroradiology.