{"title":"Characterisation of Intraocular Lens Injectors.","authors":"Maximilian Friedrich, Donald J Munro, Gerd U Auffarth, Ramin Khoramnia","doi":"10.1055/a-2349-2158","DOIUrl":null,"url":null,"abstract":"<p><p>In modern ophthalmic surgery, an intraocular lens (IOL) is commonly implanted into the patient's eye with an IOL injector. Many injectors are available, showing various technological differences, from the early manually loaded injector systems to the modern preloaded injectors. This review aims to give a concise overview of the defining characteristics of injector models and draws attention to complications that may occur during IOL implantation. One can differentiate injectors according to their preoperative preparation (manually loaded or preloaded), their implantation mechanism (push-type or screw-type or combined or automated), the size of the nozzle tip, the presence of an insertion depth control feature, and the injector's reusability. Potential complications are IOL misconfigurations such as a haptic-optic adhesion, adherence of the IOL to the injector plunger, an overriding plunger, uncontrolled IOL rotation, a trapped trailing haptic, or damage to the IOL. Additionally, during IOL implantation, the nozzle can become damaged with scratches, extensions, cracks, or bursts to the tip. While these complications rarely produce long-term consequences, manufacturers should try to prevent them by further improving their devices. Similarly, surgeons should evaluate new injectors carefully to ensure the highest possible surgical safety.</p>","PeriodicalId":17904,"journal":{"name":"Klinische Monatsblatter fur Augenheilkunde","volume":"241 8","pages":"905-916"},"PeriodicalIF":0.8000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Klinische Monatsblatter fur Augenheilkunde","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2349-2158","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/8/15 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

In modern ophthalmic surgery, an intraocular lens (IOL) is commonly implanted into the patient's eye with an IOL injector. Many injectors are available, showing various technological differences, from the early manually loaded injector systems to the modern preloaded injectors. This review aims to give a concise overview of the defining characteristics of injector models and draws attention to complications that may occur during IOL implantation. One can differentiate injectors according to their preoperative preparation (manually loaded or preloaded), their implantation mechanism (push-type or screw-type or combined or automated), the size of the nozzle tip, the presence of an insertion depth control feature, and the injector's reusability. Potential complications are IOL misconfigurations such as a haptic-optic adhesion, adherence of the IOL to the injector plunger, an overriding plunger, uncontrolled IOL rotation, a trapped trailing haptic, or damage to the IOL. Additionally, during IOL implantation, the nozzle can become damaged with scratches, extensions, cracks, or bursts to the tip. While these complications rarely produce long-term consequences, manufacturers should try to prevent them by further improving their devices. Similarly, surgeons should evaluate new injectors carefully to ensure the highest possible surgical safety.

眼内透镜注射器的特性。
在现代眼科手术中,通常使用人工晶体注射器将人工晶体植入患者眼内。从早期的手动装载注射器系统到现代的预装注射器,有许多注射器可供选择,并显示出各种技术差异。本综述旨在简明扼要地概述注射器型号的定义特征,并提请注意人工晶体植入过程中可能出现的并发症。我们可以根据注射器的术前准备(手动加载或预载)、植入机制(推动式或螺钉式或组合式或自动式)、喷嘴尖端的大小、是否具有插入深度控制功能以及注射器的可重复使用性来区分注射器。潜在的并发症包括人工晶体配置不当,如触觉与光学粘连、人工晶体与注射器柱塞粘连、柱塞过高、人工晶体旋转失控、拖曳触觉受困或人工晶体受损。此外,在人工晶体植入过程中,喷嘴可能会因划痕、延伸、裂缝或尖端爆裂而损坏。虽然这些并发症很少产生长期后果,但制造商应通过进一步改进其设备来尽量避免这些并发症的发生。同样,外科医生也应仔细评估新的注射器,以确保尽可能高的手术安全性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
1.30
自引率
0.00%
发文量
235
审稿时长
4-8 weeks
期刊介绍: -Konzentriertes Fachwissen aus Klinik und Praxis: Die entscheidenden Ergebnisse der internationalen Forschung - für Sie auf den Punkt gebracht und kritisch kommentiert, Übersichtsarbeiten zu den maßgeblichen Themen der täglichen Praxis, Top informiert - breite klinische Berichterstattung. -CME-Punkte sammeln mit dem Refresher: Effiziente, CME-zertifizierte Fortbildung, mit dem Refresher, 3 CME-Punkte pro Ausgabe - bis zu 36 CME-Punkte im Jahr!. -Aktuelle Rubriken mit echtem Nutzwert: Kurzreferate zu den wichtigsten Artikeln internationaler Zeitschriften, Schwerpunktthema in jedem Heft: Ausführliche Übersichtsarbeiten zu den wichtigsten Themen der Ophthalmologie – so behalten Sie das gesamte Fach im Blick!, Originalien mit den neuesten Entwicklungen, Übersichten zu den relevanten Themen.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信