Investigation of hand tremor suppression by a customized passive surgical support robot during internal limiting membrane peeling.

IF 4.6 Q2 MATERIALS SCIENCE, BIOMATERIALS
Soichiro Yamamoto, Yuta Taniguchi, Yohei Kuroki, Hidenori Nagahama, Shin Tanaka, Kazuaki Kadonosono, Fumihito Arai, Hiroshi Enaida
{"title":"Investigation of hand tremor suppression by a customized passive surgical support robot during internal limiting membrane peeling.","authors":"Soichiro Yamamoto, Yuta Taniguchi, Yohei Kuroki, Hidenori Nagahama, Shin Tanaka, Kazuaki Kadonosono, Fumihito Arai, Hiroshi Enaida","doi":"10.1097/IAE.0000000000004226","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To construct a quantitative evaluation system for hand tremor during internal limiting membrane (ILM) peeling and investigate changes in hand tremor attributable to the use of the customized passive surgical support robot.</p><p><strong>Methods: </strong>This analytical and experimental study developed a hand tremor evaluation system that synchronizes three elements: surgical microscope images, an artificial eye module with a force sensor to simulate ILM peeling, and microforceps with an inertial measurement unit. Two surgeons used this system to measure hand tremor during ILM peeling with and without the robot.</p><p><strong>Results: </strong>The horizontal, vertical, and combined vertical and horizontal components of hand tremor were 8.1 ± 6.1, 1.7 ± 1.8, and 8.5 ± 6.2 mG, respectively, in the no robot group. These components decreased to 7.2 ± 6.0, 1.5 ± 1.7, and 7.5 ± 6.1 mG, respectively, in the robot group. In particular, hand tremor was significantly suppressed by 11.9% for the horizontal component using the robot (p = 0.0006).</p><p><strong>Conclusion: </strong>The newly constructed system helps to quantitatively evaluate hand tremor during ILM peeling. The customized passive surgical support robot enables to decrease hand tremor during ILM peeling.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-08-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/IAE.0000000000004226","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To construct a quantitative evaluation system for hand tremor during internal limiting membrane (ILM) peeling and investigate changes in hand tremor attributable to the use of the customized passive surgical support robot.

Methods: This analytical and experimental study developed a hand tremor evaluation system that synchronizes three elements: surgical microscope images, an artificial eye module with a force sensor to simulate ILM peeling, and microforceps with an inertial measurement unit. Two surgeons used this system to measure hand tremor during ILM peeling with and without the robot.

Results: The horizontal, vertical, and combined vertical and horizontal components of hand tremor were 8.1 ± 6.1, 1.7 ± 1.8, and 8.5 ± 6.2 mG, respectively, in the no robot group. These components decreased to 7.2 ± 6.0, 1.5 ± 1.7, and 7.5 ± 6.1 mG, respectively, in the robot group. In particular, hand tremor was significantly suppressed by 11.9% for the horizontal component using the robot (p = 0.0006).

Conclusion: The newly constructed system helps to quantitatively evaluate hand tremor during ILM peeling. The customized passive surgical support robot enables to decrease hand tremor during ILM peeling.

研究定制的被动式手术支持机器人在内缘膜剥离过程中抑制手部震颤的效果。
目的:构建内缘膜(ILM)剥离过程中手部震颤的定量评估系统,并研究使用定制的被动手术支持机器人后手部震颤的变化:本分析和实验研究开发了一种手部震颤评估系统,该系统由三个元素同步组成:手术显微镜图像、带有力传感器的人工眼模块(用于模拟ILM剥离)以及带有惯性测量单元的微钳。两名外科医生使用该系统测量了在使用和不使用机器人的情况下剥离ILM过程中的手部震颤:结果:在无机器人组中,手部震颤的水平、垂直以及垂直和水平组合分量分别为 8.1 ± 6.1、1.7 ± 1.8 和 8.5 ± 6.2 mG。在机器人组中,这些分量分别降至 7.2 ± 6.0、1.5 ± 1.7 和 7.5 ± 6.1 mG。使用机器人后,手部震颤的水平分量明显减少了 11.9% (p = 0.0006):结论:新构建的系统有助于定量评估 ILM 剥离过程中的手部震颤。定制的被动式手术支持机器人可减少ILM剥离过程中的手部震颤。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
ACS Applied Bio Materials
ACS Applied Bio Materials Chemistry-Chemistry (all)
CiteScore
9.40
自引率
2.10%
发文量
464
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信