Zhihui Wan, Wei Liu, Yanhong Zhai, Zhijun Ma, Zheng Cao
{"title":"Performance Validation of the NeoBase 2 Non-Derivatized MSMS Assay Kit and Cutoff Values Establishment of Term and Preterm Neonates.","authors":"Zhihui Wan, Wei Liu, Yanhong Zhai, Zhijun Ma, Zheng Cao","doi":"10.1080/15513815.2024.2386659","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>NeoBase 2 Non-derivatized MSMS assay kit (NeoBase 2 kit) was used for newborn screening, the performance of the NeoBase 2 kit should be validated before its implementation in clinical diagnostic laboratories.</p><p><strong>Methods: </strong>Leftover dried blood spot samples, quality control materials in the NeoBase 2 kit, and proficiency testing materials received from the NSQAP were used. Precision, accuracy, LOD, LLOQ, recovery, and stability were carried out to verify the performance of the Waters ACQUITY TQD MS/MS system with the NeoBase 2 kit for newborn screening. Cutoffs were determined and analytes requiring different cutoffs in preterm neonates were investigated.</p><p><strong>Results: </strong>Within-run and between-run precisions ranged from 3.95% to 14.41%. The accuracy and stability were within 15%. All analytes demonstrated acceptable LOD, LLOQ, and recoveries. Cutoffs for term and preterm neonates were established.</p><p><strong>Conclusions: </strong>The performance of the NeoBase 2 kit is acceptable and can be implemented in clinical diagnostic laboratories.</p>","PeriodicalId":50452,"journal":{"name":"Fetal and Pediatric Pathology","volume":" ","pages":"366-375"},"PeriodicalIF":0.7000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Fetal and Pediatric Pathology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/15513815.2024.2386659","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/8/6 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PATHOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: NeoBase 2 Non-derivatized MSMS assay kit (NeoBase 2 kit) was used for newborn screening, the performance of the NeoBase 2 kit should be validated before its implementation in clinical diagnostic laboratories.
Methods: Leftover dried blood spot samples, quality control materials in the NeoBase 2 kit, and proficiency testing materials received from the NSQAP were used. Precision, accuracy, LOD, LLOQ, recovery, and stability were carried out to verify the performance of the Waters ACQUITY TQD MS/MS system with the NeoBase 2 kit for newborn screening. Cutoffs were determined and analytes requiring different cutoffs in preterm neonates were investigated.
Results: Within-run and between-run precisions ranged from 3.95% to 14.41%. The accuracy and stability were within 15%. All analytes demonstrated acceptable LOD, LLOQ, and recoveries. Cutoffs for term and preterm neonates were established.
Conclusions: The performance of the NeoBase 2 kit is acceptable and can be implemented in clinical diagnostic laboratories.
期刊介绍:
Fetal and Pediatric Pathology is an established bimonthly international journal that publishes data on diseases of the developing embryo, newborns, children, and adolescents. The journal publishes original and review articles and reportable case reports.
The expanded scope of the journal encompasses molecular basis of genetic disorders; molecular basis of diseases that lead to implantation failures; molecular basis of abnormal placentation; placentology and molecular basis of habitual abortion; intrauterine development and molecular basis of embryonic death; pathogenisis and etiologic factors involved in sudden infant death syndrome; the underlying molecular basis, and pathogenesis of diseases that lead to morbidity and mortality in newborns; prenatal, perinatal, and pediatric diseases and molecular basis of diseases of childhood including solid tumors and tumors of the hematopoietic system; and experimental and molecular pathology.