Yue Pan , Yu Qiang , Wenjie Liang , Wenyue Huang , Ningyuan Wang , Xingying Wang , Zhiqiang Zhang , Weibao Qiu , Hairong Zheng
{"title":"A transcranial multiple waves suppression method for plane wave imaging based on Radon transform","authors":"Yue Pan , Yu Qiang , Wenjie Liang , Wenyue Huang , Ningyuan Wang , Xingying Wang , Zhiqiang Zhang , Weibao Qiu , Hairong Zheng","doi":"10.1016/j.ultras.2024.107405","DOIUrl":null,"url":null,"abstract":"<div><p>Transcranial ultrasound imaging presents a significant challenge due to the intricate interplay between ultrasound waves and the heterogeneous human skull. The skull’s presence induces distortion, refraction, multiple scattering, and reflection of ultrasound signals, thereby complicating the acquisition of high-quality images. Extracting reflections from the entire waveform is crucial yet exceedingly challenging, as intracranial reflections are often obscured by strong amplitude direct waves and multiple scattering. In this paper, a multiple wave suppression method for ultrasound plane wave imaging is proposed to mitigate the impact of skull interference. Drawing upon prior research, we developed an enhanced high-resolution linear Radon transform using the maximum entropy principle and Bayesian method, facilitating wavefield separation. We detailed the process of wave field separation in the Radon domain through simulation of a model with a high velocity layer. When plane waves emitted at any steering angles, both multiple waves and first arrival waves manifested as distinct energy points. In the brain simulation, we contrasted the characteristic differences between skull reflection and brain-internal signal in Radon domain, and demonstrated that multiples suppression method reduces side and grating lobe levels by approximately 30 dB. Finally, we executed <em>in vitro</em> experiments using a monkey skull to separate weak intracranial reflection signals from strong skull reflections, enhancing the contrast-to-noise ratio by 85 % compared to conventional method using full waveform. This study deeply explores the effect of multiples on effective signal separation, addresses the complexity of wavefield separation, and verifies its efficacy through imaging, thereby significantly advancing ultrasound transcranial imaging techniques.</p></div>","PeriodicalId":23522,"journal":{"name":"Ultrasonics","volume":"143 ","pages":"Article 107405"},"PeriodicalIF":3.8000,"publicationDate":"2024-07-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ultrasonics","FirstCategoryId":"101","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0041624X24001689","RegionNum":2,"RegionCategory":"物理与天体物理","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ACOUSTICS","Score":null,"Total":0}
引用次数: 0
Abstract
Transcranial ultrasound imaging presents a significant challenge due to the intricate interplay between ultrasound waves and the heterogeneous human skull. The skull’s presence induces distortion, refraction, multiple scattering, and reflection of ultrasound signals, thereby complicating the acquisition of high-quality images. Extracting reflections from the entire waveform is crucial yet exceedingly challenging, as intracranial reflections are often obscured by strong amplitude direct waves and multiple scattering. In this paper, a multiple wave suppression method for ultrasound plane wave imaging is proposed to mitigate the impact of skull interference. Drawing upon prior research, we developed an enhanced high-resolution linear Radon transform using the maximum entropy principle and Bayesian method, facilitating wavefield separation. We detailed the process of wave field separation in the Radon domain through simulation of a model with a high velocity layer. When plane waves emitted at any steering angles, both multiple waves and first arrival waves manifested as distinct energy points. In the brain simulation, we contrasted the characteristic differences between skull reflection and brain-internal signal in Radon domain, and demonstrated that multiples suppression method reduces side and grating lobe levels by approximately 30 dB. Finally, we executed in vitro experiments using a monkey skull to separate weak intracranial reflection signals from strong skull reflections, enhancing the contrast-to-noise ratio by 85 % compared to conventional method using full waveform. This study deeply explores the effect of multiples on effective signal separation, addresses the complexity of wavefield separation, and verifies its efficacy through imaging, thereby significantly advancing ultrasound transcranial imaging techniques.
期刊介绍:
Ultrasonics is the only internationally established journal which covers the entire field of ultrasound research and technology and all its many applications. Ultrasonics contains a variety of sections to keep readers fully informed and up-to-date on the whole spectrum of research and development throughout the world. Ultrasonics publishes papers of exceptional quality and of relevance to both academia and industry. Manuscripts in which ultrasonics is a central issue and not simply an incidental tool or minor issue, are welcomed.
As well as top quality original research papers and review articles by world renowned experts, Ultrasonics also regularly features short communications, a calendar of forthcoming events and special issues dedicated to topical subjects.