The impact of V-flow on preoperative diagnosis of thyroid tumors: individually and as part of multimodal sonographic imaging.

IF 1.3 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Moritz Konstantin Brandenstein, Liang Zhang, Gregor Scharf, Sylvia Thurn, Matthias Hornung, Karin Menhart, Stefanie Meiler, Christian Stroszczynski, Ernst Michael Jung
{"title":"The impact of V-flow on preoperative diagnosis of thyroid tumors: individually and as part of multimodal sonographic imaging.","authors":"Moritz Konstantin Brandenstein, Liang Zhang, Gregor Scharf, Sylvia Thurn, Matthias Hornung, Karin Menhart, Stefanie Meiler, Christian Stroszczynski, Ernst Michael Jung","doi":"10.1055/a-2350-0107","DOIUrl":null,"url":null,"abstract":"<p><p>V-flow is a dynamic ultrasound technique that visualizes perfusion patterns by displaying dynamic arrows that change in response to the flow of erythrocytes. Furthermore, it provides quantitative values for the maximum and mean velocity of blood flow as well as a percentage value for turbulence. The aim was to enhance the preoperative diagnostic accuracy of thyroid lesions by combining V-flow with established ultrasound modes.B-mode, CCDS, elastography, CEUS, and V-flow were performed on 101 patients. After the ultrasound examination, every nodule was confirmed as benign or malignant via histopathology. The Kruskal-Wallis test, ROC curve, and binary logistic regression were used for the statistical analysis.93 benign regressive thyroid nodules and 8 carcinomas were included in this study. The average mean velocity value for benign lesions was measured at 19.5 cm/s and at 10.7 cm/s for malignant lesions (p = 0.039). The average turbulence percentage was 26.1% for benign nodules and 46.7% for carcinomas (p = 0.016). Carcinomas exhibited a slower and more turbulent perfusion pattern compared to benign tumors. A V-flow-centered system achieves a sensitivity of 100.0% and a specificity of 84.9% in predicting malignancy. This system could have reduced the number of unnecessary thyroid surgeries for benign lesions in our patient group by 70%.The capillary perfusion of thyroid nodules represents a significant indicator of its status. By analyzing the velocity and turbulence level of microvascular blood flow, V-flow offers promising prospects for accurately distinguishing between benign and malignant thyroid lesions. When integrated into a comprehensive multimodal sonographic imaging approach, V-flow further enhances diagnostic accuracy. · V-flow allows for qualitative and quantitative analysis of microvascular perfusion. · Malignant tumors are associated with slower and more turbulent microvascular hemodynamics. · Combining V-flow with other ultrasound modes eases the diagnosis of thyroid carcinomas. · Brandenstein MK, Zhang L, Scharf G et al. The impact of V-flow on preoperative diagnosis of thyroid tumors: individually and as part of multimodal sonographic imaging. Fortschr Röntgenstr 2025; 197: 436-446.</p>","PeriodicalId":21490,"journal":{"name":"Rofo-fortschritte Auf Dem Gebiet Der Rontgenstrahlen Und Der Bildgebenden Verfahren","volume":" ","pages":"436-446"},"PeriodicalIF":1.3000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Rofo-fortschritte Auf Dem Gebiet Der Rontgenstrahlen Und Der Bildgebenden Verfahren","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2350-0107","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/22 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

V-flow is a dynamic ultrasound technique that visualizes perfusion patterns by displaying dynamic arrows that change in response to the flow of erythrocytes. Furthermore, it provides quantitative values for the maximum and mean velocity of blood flow as well as a percentage value for turbulence. The aim was to enhance the preoperative diagnostic accuracy of thyroid lesions by combining V-flow with established ultrasound modes.B-mode, CCDS, elastography, CEUS, and V-flow were performed on 101 patients. After the ultrasound examination, every nodule was confirmed as benign or malignant via histopathology. The Kruskal-Wallis test, ROC curve, and binary logistic regression were used for the statistical analysis.93 benign regressive thyroid nodules and 8 carcinomas were included in this study. The average mean velocity value for benign lesions was measured at 19.5 cm/s and at 10.7 cm/s for malignant lesions (p = 0.039). The average turbulence percentage was 26.1% for benign nodules and 46.7% for carcinomas (p = 0.016). Carcinomas exhibited a slower and more turbulent perfusion pattern compared to benign tumors. A V-flow-centered system achieves a sensitivity of 100.0% and a specificity of 84.9% in predicting malignancy. This system could have reduced the number of unnecessary thyroid surgeries for benign lesions in our patient group by 70%.The capillary perfusion of thyroid nodules represents a significant indicator of its status. By analyzing the velocity and turbulence level of microvascular blood flow, V-flow offers promising prospects for accurately distinguishing between benign and malignant thyroid lesions. When integrated into a comprehensive multimodal sonographic imaging approach, V-flow further enhances diagnostic accuracy. · V-flow allows for qualitative and quantitative analysis of microvascular perfusion. · Malignant tumors are associated with slower and more turbulent microvascular hemodynamics. · Combining V-flow with other ultrasound modes eases the diagnosis of thyroid carcinomas. · Brandenstein MK, Zhang L, Scharf G et al. The impact of V-flow on preoperative diagnosis of thyroid tumors: individually and as part of multimodal sonographic imaging. Fortschr Röntgenstr 2025; 197: 436-446.

V型血流对甲状腺肿瘤术前诊断的影响:单独和作为多模态超声成像的一部分。
V 型血流是一种动态超声技术,它通过显示随红细胞流动而变化的动态箭头来显示灌注模式。此外,它还能提供血流最大速度和平均速度的定量值以及湍流的百分比值。我们对 101 名患者进行了 B 型、CCDS、弹性成像、CEUS 和 V 型血流检查。超声检查后,每个结节均通过组织病理学证实为良性或恶性。本研究采用 Kruskal-Wallis 检验、ROC 曲线和二元逻辑回归进行统计分析,其中包括 93 个良性退行性甲状腺结节和 8 个癌。良性病变的平均速度值为 19.5 厘米/秒,恶性病变的平均速度值为 10.7 厘米/秒(P = 0.039)。良性结节的平均湍流率为 26.1%,癌细胞为 46.7%(p = 0.016)。与良性肿瘤相比,癌肿的灌注速度更慢,湍流更多。以 V 型血流为中心的系统在预测恶性肿瘤方面的灵敏度为 100.0%,特异性为 84.9%。在我们的患者群体中,该系统可以将甲状腺良性病变的不必要手术数量减少 70%。通过分析微血管血流的速度和湍流程度,V-flow有望准确区分甲状腺良性和恶性病变。如果将 V 型血流纳入全面的多模态声学成像方法,还能进一步提高诊断的准确性。- V 型血流可对微血管灌注进行定性和定量分析。- 恶性肿瘤的微血管血流动力学速度更慢、更湍急。- 将V-flow与其他超声模式相结合可简化甲状腺癌的诊断。- Brandenstein MK、Zhang L、Scharf G 等:V-flow 对甲状腺肿瘤术前诊断的影响:单独使用和作为多模式超声成像的一部分。Fortschr Röntgenstr 2024; DOI 10.1055/a-2350-0107.
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
1.20
自引率
5.60%
发文量
340
期刊介绍: Die RöFo veröffentlicht Originalarbeiten, Übersichtsartikel und Fallberichte aus dem Bereich der Radiologie und den weiteren bildgebenden Verfahren in der Medizin. Es dürfen nur Arbeiten eingereicht werden, die noch nicht veröffentlicht sind und die auch nicht gleichzeitig einer anderen Zeitschrift zur Veröffentlichung angeboten wurden. Alle eingereichten Beiträge unterliegen einer sorgfältigen fachlichen Begutachtung. Gegründet 1896 – nur knapp 1 Jahr nach der Entdeckung der Röntgenstrahlen durch C.W. Röntgen – blickt die RöFo auf über 100 Jahre Erfahrung als wichtigstes Publikationsmedium in der deutschsprachigen Radiologie zurück. Sie ist damit die älteste radiologische Fachzeitschrift und schafft es erfolgreich, lange Kontinuität mit dem Anspruch an wissenschaftliches Publizieren auf internationalem Niveau zu verbinden. Durch ihren zentralen Platz im Verlagsprogramm stellte die RöFo die Basis für das heute umfassende und erfolgreiche Radiologie-Medienangebot im Georg Thieme Verlag. Besonders eng verbunden ist die RöFo mit der Geschichte der Röntgengesellschaften in Deutschland und Österreich. Sie ist offizielles Organ von DRG und ÖRG und die Mitglieder der Fachgesellschaften erhalten die Zeitschrift im Rahmen ihrer Mitgliedschaft. Mit ihrem wissenschaftlichen Kernteil und dem eigenen Mitteilungsteil der Fachgesellschaften bietet die RöFo Monat für Monat ein Forum für den Austausch von Inhalten und Botschaften der radiologischen Community im deutschsprachigen Raum.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信