Accuracy of intraocular lens power formulas in patients with average keratometry greater than 46 diopters

H. Rodríguez Ortiz , L. Flores Ortiz , R. Devereux García , C. Palacio Pastrana , M. Hilario Pérez , G. Villanueva Pérez , J.M. Piña Valdez , G.Y. Montalvo Domínguez , L.P. Gutierrez Díaz
{"title":"Accuracy of intraocular lens power formulas in patients with average keratometry greater than 46 diopters","authors":"H. Rodríguez Ortiz ,&nbsp;L. Flores Ortiz ,&nbsp;R. Devereux García ,&nbsp;C. Palacio Pastrana ,&nbsp;M. Hilario Pérez ,&nbsp;G. Villanueva Pérez ,&nbsp;J.M. Piña Valdez ,&nbsp;G.Y. Montalvo Domínguez ,&nbsp;L.P. Gutierrez Díaz","doi":"10.1016/j.oftale.2024.07.006","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>To compare the accuracy of Kane, Barrett Universal II, Haigis, and SRK-T formulas in eyes with average keratometry greater than 46 diopters (D).</div></div><div><h3>Methods</h3><div>A retrospective analysis was conducted on 101 eyes of 101 patients with average keratometry greater than 46 D. The absolute prediction error (EA) was obtained for each patient one month after surgery. The mean absolute prediction error (MEA), median absolute prediction error (MedEA) and the percentage of patients with absolute refractive error less than 0.25 D, 0.50 D, and 1.00 D were calculated for each formula analyzed.</div></div><div><h3>Results</h3><div>The Kane formula achieved the lowest MEA (0.53 ± 0.43) and the lowest MedEA (0.41), followed by Barrett Universal II (MEA: 0.56 ± 0.42, MedEA: 0.49), SRK-T (MEA: 0.59 ± 0.44, MedEA: 0.54), and Haigis (MEA: 0.77 ± 0.47, MedEA: 0.69), showing a significant difference in the results. It was also observed that the Kane formula was the most accurate, with the highest percentage of patients, with EA less than 0.25 D, 0.50 D, and 1.00 D (30.7%, 54.4%, and 86.1%, respectively), while the Haigis formula was the least accurate (12.9%, 33.7%, and 69.3%, respectively).</div></div><div><h3>Conclusion</h3><div>In eyes with corneas having average keratometry greater than 46 D, the Kane formula proves to be a useful tool in intraocular lens (IOL) power calculation and demonstrates higher precision compared to the Barrett Universal II, SRK-T, and Haigis formulas.</div></div>","PeriodicalId":93886,"journal":{"name":"Archivos de la Sociedad Espanola de Oftalmologia","volume":"99 10","pages":"Pages 413-419"},"PeriodicalIF":0.0000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archivos de la Sociedad Espanola de Oftalmologia","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2173579424001361","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objective

To compare the accuracy of Kane, Barrett Universal II, Haigis, and SRK-T formulas in eyes with average keratometry greater than 46 diopters (D).

Methods

A retrospective analysis was conducted on 101 eyes of 101 patients with average keratometry greater than 46 D. The absolute prediction error (EA) was obtained for each patient one month after surgery. The mean absolute prediction error (MEA), median absolute prediction error (MedEA) and the percentage of patients with absolute refractive error less than 0.25 D, 0.50 D, and 1.00 D were calculated for each formula analyzed.

Results

The Kane formula achieved the lowest MEA (0.53 ± 0.43) and the lowest MedEA (0.41), followed by Barrett Universal II (MEA: 0.56 ± 0.42, MedEA: 0.49), SRK-T (MEA: 0.59 ± 0.44, MedEA: 0.54), and Haigis (MEA: 0.77 ± 0.47, MedEA: 0.69), showing a significant difference in the results. It was also observed that the Kane formula was the most accurate, with the highest percentage of patients, with EA less than 0.25 D, 0.50 D, and 1.00 D (30.7%, 54.4%, and 86.1%, respectively), while the Haigis formula was the least accurate (12.9%, 33.7%, and 69.3%, respectively).

Conclusion

In eyes with corneas having average keratometry greater than 46 D, the Kane formula proves to be a useful tool in intraocular lens (IOL) power calculation and demonstrates higher precision compared to the Barrett Universal II, SRK-T, and Haigis formulas.
平均角膜度数大于 46 度的患者眼内晶状体功率公式的准确性。
目的:比较凯恩公式、巴雷特通用 II 公式、海吉斯公式和 SRK-T 公式对平均角膜度数大于 46 度 (D) 眼睛的准确性:比较 Kane、Barrett Universal II、Haigis 和 SRK-T 公式在平均角膜度数大于 46 度 (D) 的眼睛中的准确性:对平均角膜度数大于 46 D 的 101 名患者的 101 只眼睛进行了回顾性分析。计算了每个分析公式的平均绝对预测误差(MEA)、中位数绝对预测误差(MedEA)以及绝对屈光误差小于 0.25 D、0.50 D 和 1.00 D 的患者比例:Kane 公式的 MEA(0.53 ± 0.43)和 MedEA(0.41)最低,其次是 Barrett Universal II(MEA:0.56 ± 0.42,MedEA:0.49)、SRK-T(MEA:0.59 ± 0.44,MedEA:0.54)和 Haigis(MEA:0.77 ± 0.47,MedEA:0.69),结果显示差异显著。同时还观察到,凯恩公式的准确度最高,EA小于0.25 D、0.50 D和1.00 D的患者比例最高(分别为30.7%、54.4%和86.1%),而海吉斯公式的准确度最低(分别为12.9%、33.7%和69.3%):结论:对于平均角膜屈光度大于 46 D 的眼睛,Kane 公式被证明是计算人工晶体(IOL)功率的有用工具,与 Barrett Universal II、SRK-T 和 Haigis 公式相比,Kane 公式的精确度更高。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信