Day-to-day reliability of basal heart rate and short-term and ultra short-term heart rate variability assessment by the Equivital eq02+ LifeMonitor in US Army soldiers.
Christopher L Chapman, E A Schafer, A W Potter, E M Lavoie, B M Roberts, J W Castellani, K E Friedl, D P Looney
{"title":"Day-to-day reliability of basal heart rate and short-term and ultra short-term heart rate variability assessment by the Equivital eq02+ LifeMonitor in US Army soldiers.","authors":"Christopher L Chapman, E A Schafer, A W Potter, E M Lavoie, B M Roberts, J W Castellani, K E Friedl, D P Looney","doi":"10.1136/military-2024-002687","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>The present study determined the (1) day-to-day reliability of basal heart rate (HR) and HR variability (HRV) measured by the Equivital eq02+ LifeMonitor and (2) agreement of ultra short-term HRV compared with short-term HRV.</p><p><strong>Methods: </strong>Twenty-three active-duty US Army Soldiers (5 females, 18 males) completed two experimental visits separated by >48 hours with restrictions consistent with basal monitoring (eg, exercise, dietary), with measurements after supine rest at minutes 20-21 (ultra short-term) and minutes 20-25 (short-term). HRV was assessed as the SD of R-R intervals (SDNN) and the square root of the mean squared differences between consecutive R-R intervals (RMSSD).</p><p><strong>Results: </strong>The day-to-day reliability (intraclass correlation coefficient (ICC)) using linear-mixed model approach was good for HR (0.849, 95% CI: 0.689 to 0.933) and RMSSD (ICC: 0.823, 95% CI: 0.623 to 0.920). SDNN had moderate day-to-day reliability with greater variation (ICC: 0.689, 95% CI: 0.428 to 0.858). The reliability of RMSSD was slightly improved when considering the effect of respiration (ICC: 0.821, 95% CI: 0.672 to 0.944). There was no bias for HR measured for 1 min versus 5 min (p=0.511). For 1 min measurements versus 5 min, there was a very modest mean bias of -4 ms for SDNN and -1 ms for RMSSD (p≤0.023).</p><p><strong>Conclusion: </strong>When preceded by a 20 min stabilisation period using restrictions consistent with basal monitoring and measuring respiration, military personnel can rely on the eq02+ for basal HR and RMSSD monitoring but should be more cautious using SDNN. These data also support using ultra short-term measurements when following these procedures.</p>","PeriodicalId":48485,"journal":{"name":"Bmj Military Health","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2024-07-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bmj Military Health","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1136/military-2024-002687","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: The present study determined the (1) day-to-day reliability of basal heart rate (HR) and HR variability (HRV) measured by the Equivital eq02+ LifeMonitor and (2) agreement of ultra short-term HRV compared with short-term HRV.
Methods: Twenty-three active-duty US Army Soldiers (5 females, 18 males) completed two experimental visits separated by >48 hours with restrictions consistent with basal monitoring (eg, exercise, dietary), with measurements after supine rest at minutes 20-21 (ultra short-term) and minutes 20-25 (short-term). HRV was assessed as the SD of R-R intervals (SDNN) and the square root of the mean squared differences between consecutive R-R intervals (RMSSD).
Results: The day-to-day reliability (intraclass correlation coefficient (ICC)) using linear-mixed model approach was good for HR (0.849, 95% CI: 0.689 to 0.933) and RMSSD (ICC: 0.823, 95% CI: 0.623 to 0.920). SDNN had moderate day-to-day reliability with greater variation (ICC: 0.689, 95% CI: 0.428 to 0.858). The reliability of RMSSD was slightly improved when considering the effect of respiration (ICC: 0.821, 95% CI: 0.672 to 0.944). There was no bias for HR measured for 1 min versus 5 min (p=0.511). For 1 min measurements versus 5 min, there was a very modest mean bias of -4 ms for SDNN and -1 ms for RMSSD (p≤0.023).
Conclusion: When preceded by a 20 min stabilisation period using restrictions consistent with basal monitoring and measuring respiration, military personnel can rely on the eq02+ for basal HR and RMSSD monitoring but should be more cautious using SDNN. These data also support using ultra short-term measurements when following these procedures.