In Vivo Computed Tomography Sizing for Redo-Transcatheter Aortic Valve Replacement in Evolut Valves: Impact on Sizing, Feasibility, and Prosthesis-Patient Mismatch.

IF 6.1 1区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS
Atsushi Okada, Miho Fukui, Syed Zaid, Kiahltone R Thao, Evan Walser-Kuntz, Larissa I Stanberry, Marcus R Burns, Hideki Koike, Cheng Wang, Asa Phichaphop, John R Lesser, João L Cavalcante, Paul Sorajja, Vinayak N Bapat
{"title":"In Vivo Computed Tomography Sizing for Redo-Transcatheter Aortic Valve Replacement in Evolut Valves: Impact on Sizing, Feasibility, and Prosthesis-Patient Mismatch.","authors":"Atsushi Okada, Miho Fukui, Syed Zaid, Kiahltone R Thao, Evan Walser-Kuntz, Larissa I Stanberry, Marcus R Burns, Hideki Koike, Cheng Wang, Asa Phichaphop, John R Lesser, João L Cavalcante, Paul Sorajja, Vinayak N Bapat","doi":"10.1161/CIRCINTERVENTIONS.123.013903","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>SAPIEN3 (S3) is a ubiquitous redo-transcatheter aortic valve (TAV) replacement alternative for degenerated Evolut valves, but S3 sizing for S3-in-Evolut remains unclear. We sought to compare the impact of in vivo computed tomography (CT)-sizing on redo-TAV feasibility for S3-in-Evolut with traditional bench-sizing.</p><p><strong>Methods: </strong>CT scans of 290 patients treated using Evolut R/PRO/PRO+ between July 2015 and December 2021 were analyzed. S3-in-Evolut was simulated using S3 outflow/neoskirt plane (NSP) at node-6, -5, and -4. CT-sizing for S3 was determined by averaging 4 areas of the Evolut stent frame at NSP level and 3 nodes below. Redo-TAV was deemed feasible if the NSP was below the coronaries, or the narrowest valve (virtual S3)-to-aorta distance was >4 mm. Risk of prosthesis-patient mismatch was estimated using predicted indexed-effective orifice area.</p><p><strong>Results: </strong>Compared with bench-sizing, CT-sizing yielded smaller S3 size in 82% at node-6, 81% at node-5, and 84% at node-4. Factors associated with CT-sizing less than bench-sizing were larger index Evolut size, underexpansion of index Evolut, and shallower implant depth (all <i>P</i><0.05). CT-sizing increased redo-TAV feasibility by +8% at node-6, +10% at node-5, and +4% at node-4. Redo-TAV feasibility increased with annulus size, sinotubular junction dimensions, coronary heights, index Evolut size, deeper Evolut implant depth, and lower NSP levels (all <i>P</i><0.05). CT-sizing had a slightly higher estimated risk of severe prosthesis-patient mismatch (9% at node-6, 7% at node-5, and 6% at node-4), which could be mitigated by changing the NSP.</p><p><strong>Conclusions: </strong>CT-sizing for S3-in-Evolut is associated with higher feasibility of redo-TAV compared with bench-sizing, potentially reducing the risk of excessive oversizing and S3 underexpansion. Further validation using real-world clinical data is necessary.</p>","PeriodicalId":10330,"journal":{"name":"Circulation: Cardiovascular Interventions","volume":" ","pages":"e013903"},"PeriodicalIF":6.1000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Circulation: Cardiovascular Interventions","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1161/CIRCINTERVENTIONS.123.013903","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/27 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: SAPIEN3 (S3) is a ubiquitous redo-transcatheter aortic valve (TAV) replacement alternative for degenerated Evolut valves, but S3 sizing for S3-in-Evolut remains unclear. We sought to compare the impact of in vivo computed tomography (CT)-sizing on redo-TAV feasibility for S3-in-Evolut with traditional bench-sizing.

Methods: CT scans of 290 patients treated using Evolut R/PRO/PRO+ between July 2015 and December 2021 were analyzed. S3-in-Evolut was simulated using S3 outflow/neoskirt plane (NSP) at node-6, -5, and -4. CT-sizing for S3 was determined by averaging 4 areas of the Evolut stent frame at NSP level and 3 nodes below. Redo-TAV was deemed feasible if the NSP was below the coronaries, or the narrowest valve (virtual S3)-to-aorta distance was >4 mm. Risk of prosthesis-patient mismatch was estimated using predicted indexed-effective orifice area.

Results: Compared with bench-sizing, CT-sizing yielded smaller S3 size in 82% at node-6, 81% at node-5, and 84% at node-4. Factors associated with CT-sizing less than bench-sizing were larger index Evolut size, underexpansion of index Evolut, and shallower implant depth (all P<0.05). CT-sizing increased redo-TAV feasibility by +8% at node-6, +10% at node-5, and +4% at node-4. Redo-TAV feasibility increased with annulus size, sinotubular junction dimensions, coronary heights, index Evolut size, deeper Evolut implant depth, and lower NSP levels (all P<0.05). CT-sizing had a slightly higher estimated risk of severe prosthesis-patient mismatch (9% at node-6, 7% at node-5, and 6% at node-4), which could be mitigated by changing the NSP.

Conclusions: CT-sizing for S3-in-Evolut is associated with higher feasibility of redo-TAV compared with bench-sizing, potentially reducing the risk of excessive oversizing and S3 underexpansion. Further validation using real-world clinical data is necessary.

使用 Evolut 瓣膜进行再经导管主动脉瓣置换术的体内计算机断层扫描选型:对尺寸、可行性和假体与患者不匹配的影响。
背景:SAPIEN3(S3)是一种无处不在的重做经导管主动脉瓣(TAV)替代退化的Evolut瓣膜的方法,但S3-in-Evolut的S3尺寸仍不清楚。我们试图比较活体计算机断层扫描(CT)与传统的工作台尺寸对S3-in-Evolut重做TAV可行性的影响:分析了2015年7月至2021年12月期间使用Evolut R/PRO/PRO+治疗的290名患者的CT扫描结果。在结节-6、-5和-4处使用S3流出/新裙平面(NSP)模拟S3-in-Evolut。S3 的 CT 大小由 Evolut 支架框架在 NSP 水平的 4 个区域和下面 3 个节点的平均值决定。如果 NSP 位于冠状动脉下方,或最窄瓣膜(虚拟 S3)到主动脉的距离大于 4 毫米,则认为可以重新进行 TAV。使用预测的指数化有效瓣口面积估算假体与患者不匹配的风险:结果:与台式测量相比,CT测量的S3尺寸较小的患者比例分别为:节点6的82%、节点5的81%和节点4的84%。CT尺寸小于工作台尺寸的相关因素是指数Evolut尺寸较大、指数Evolut扩张不足和植入深度较浅(均为PPConclusions):S3-in-Evolut的CT尺寸与台式尺寸相比,重做TAV的可行性更高,有可能降低尺寸过大和S3扩张不足的风险。有必要使用真实世界的临床数据进行进一步验证。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Circulation: Cardiovascular Interventions
Circulation: Cardiovascular Interventions CARDIAC & CARDIOVASCULAR SYSTEMS-
CiteScore
10.30
自引率
1.80%
发文量
221
审稿时长
6-12 weeks
期刊介绍: Circulation: Cardiovascular Interventions, an American Heart Association journal, focuses on interventional techniques pertaining to coronary artery disease, structural heart disease, and vascular disease, with priority placed on original research and on randomized trials and large registry studies. In addition, pharmacological, diagnostic, and pathophysiological aspects of interventional cardiology are given special attention in this online-only journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信