Mateusz Ulman, Krzysztof Boczar, Katarzyna Holcman, Magdalena Ziąbka, Maciej Dębski, Jacek Lelakowski, Andrzej Ząbek
{"title":"Endocardial lead insulation wear in a scanning and optical microscope.","authors":"Mateusz Ulman, Krzysztof Boczar, Katarzyna Holcman, Magdalena Ziąbka, Maciej Dębski, Jacek Lelakowski, Andrzej Ząbek","doi":"10.17219/acem/186864","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The path and interaction of leads within the cardiovascular system are influenced by various factors, including the implantation technique. Furthermore, the multifaceted composition of these leads, often comprising multiple materials, can contribute to their potential degradation and wear over time.</p><p><strong>Objectives: </strong>Our aim was to investigate the wear of lead insulation following the removal of transvenous leads and pinpoint the regions of the lead most vulnerable to damage.</p><p><strong>Material and methods: </strong>We undertook a prospective analysis of patients from a single tertiary center who underwent transvenous lead explantation (TLE) between October 1, 2013, and July 31, 2015. Specifically, our examination focused on endocardial leads removed using simple screw-out and gentle traction techniques. Subsequent lead evaluations were conducted utilizing scanning electron and optical microscopes.</p><p><strong>Results: </strong>Among the 86 patients who underwent the TLE procedure, 26 patients (30%) required the removal of 39 leads through simple traction. Inspection using scanning electron microscopy consistently indicated insulation damage across all leads. A total of 347 damaged sites were identified: 261 without lead unsealing and 86 exhibiting unsealing. Notably, the sections of the leads located within the intra-pocket area demonstrated the highest vulnerability to damage (odds ratio (OR): = 9.112, 95% confidence interval (95% CI): 3.326-24.960), whereas the intravenous regions displayed the lowest susceptibility (OR: 0.323, 95% CI: 0.151-0.694).</p><p><strong>Conclusions: </strong>Our study reveals that all evaluated leads exhibited insulation damage, with the intra-pocket segments manifesting a notably higher prevalence of damage than the intravenous segments.</p>","PeriodicalId":7306,"journal":{"name":"Advances in Clinical and Experimental Medicine","volume":" ","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2024-06-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Advances in Clinical and Experimental Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.17219/acem/186864","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The path and interaction of leads within the cardiovascular system are influenced by various factors, including the implantation technique. Furthermore, the multifaceted composition of these leads, often comprising multiple materials, can contribute to their potential degradation and wear over time.
Objectives: Our aim was to investigate the wear of lead insulation following the removal of transvenous leads and pinpoint the regions of the lead most vulnerable to damage.
Material and methods: We undertook a prospective analysis of patients from a single tertiary center who underwent transvenous lead explantation (TLE) between October 1, 2013, and July 31, 2015. Specifically, our examination focused on endocardial leads removed using simple screw-out and gentle traction techniques. Subsequent lead evaluations were conducted utilizing scanning electron and optical microscopes.
Results: Among the 86 patients who underwent the TLE procedure, 26 patients (30%) required the removal of 39 leads through simple traction. Inspection using scanning electron microscopy consistently indicated insulation damage across all leads. A total of 347 damaged sites were identified: 261 without lead unsealing and 86 exhibiting unsealing. Notably, the sections of the leads located within the intra-pocket area demonstrated the highest vulnerability to damage (odds ratio (OR): = 9.112, 95% confidence interval (95% CI): 3.326-24.960), whereas the intravenous regions displayed the lowest susceptibility (OR: 0.323, 95% CI: 0.151-0.694).
Conclusions: Our study reveals that all evaluated leads exhibited insulation damage, with the intra-pocket segments manifesting a notably higher prevalence of damage than the intravenous segments.
期刊介绍:
Advances in Clinical and Experimental Medicine has been published by the Wroclaw Medical University since 1992. Establishing the medical journal was the idea of Prof. Bogumił Halawa, Chair of the Department of Cardiology, and was fully supported by the Rector of Wroclaw Medical University, Prof. Zbigniew Knapik. Prof. Halawa was also the first editor-in-chief, between 1992-1997. The journal, then entitled "Postępy Medycyny Klinicznej i Doświadczalnej", appeared quarterly.
Prof. Leszek Paradowski was editor-in-chief from 1997-1999. In 1998 he initiated alterations in the profile and cover design of the journal which were accepted by the Editorial Board. The title was changed to Advances in Clinical and Experimental Medicine. Articles in English were welcomed. A number of outstanding representatives of medical science from Poland and abroad were invited to participate in the newly established International Editorial Staff.
Prof. Antonina Harłozińska-Szmyrka was editor-in-chief in years 2000-2005, in years 2006-2007 once again prof. Leszek Paradowski and prof. Maria Podolak-Dawidziak was editor-in-chief in years 2008-2016. Since 2017 the editor-in chief is prof. Maciej Bagłaj.
Since July 2005, original papers have been published only in English. Case reports are no longer accepted. The manuscripts are reviewed by two independent reviewers and a statistical reviewer, and English texts are proofread by a native speaker.
The journal has been indexed in several databases: Scopus, Ulrich’sTM International Periodicals Directory, Index Copernicus and since 2007 in Thomson Reuters databases: Science Citation Index Expanded i Journal Citation Reports/Science Edition.
In 2010 the journal obtained Impact Factor which is now 1.179 pts. Articles published in the journal are worth 15 points among Polish journals according to the Polish Committee for Scientific Research and 169.43 points according to the Index Copernicus.
Since November 7, 2012, Advances in Clinical and Experimental Medicine has been indexed and included in National Library of Medicine’s MEDLINE database. English abstracts printed in the journal are included and searchable using PubMed http://www.ncbi.nlm.nih.gov/pubmed.