Riza M Cetik, Charles H Crawford, Steven D Glassman, John R Dimar, Jeffrey L Gum, Mladen Djurasovic, Leah Y Carreon
{"title":"Accuracy of Phantomless Calibration of Routine Computed Tomography Scans for Opportunistic Osteoporosis Screening in the Spine Clinic.","authors":"Riza M Cetik, Charles H Crawford, Steven D Glassman, John R Dimar, Jeffrey L Gum, Mladen Djurasovic, Leah Y Carreon","doi":"10.1097/BRS.0000000000005080","DOIUrl":null,"url":null,"abstract":"<p><strong>Study design: </strong>Diagnostic accuracy study.</p><p><strong>Objective: </strong>To establish a simple method of phantomless bone mineral density (BMD) measurement by using preoperative lumbar computed tomography (CT) scans, and compare the accuracy of reference tissue combinations to diagnose low BMD against uncalibrated Hounsfield units (HUs).</p><p><strong>Summary of background data: </strong>HUs are used as a measure of BMD; however, associations between HU and T-scores vary widely. Quantitative CT (qCT) scans are more accurate, but they require density calibration with an object of known density (phantom), which limits feasibility. As an emerging technique, phantomless (internal) calibration of routine CT scans may provide a good opportunity for screening.</p><p><strong>Materials and methods: </strong>Patients who were scheduled to undergo lumbar surgery, with a preoperative CT scan, and a dual-energy x-ray absorptiometry (DXA) scan within six months were included. Four tissues were selected for calibration: subcutaneous adipose (A), erector spinae (ES), psoas (P), and aortic blood (AB). The HUs of these tissues were used in linear regression against ground-truth values. Calibrations were performed by using two different internal tissues at a time to maintain simplicity and in-office applicability.Volumetric bone mineral densities (vBMD) derived from internally calibrated CT scans were analyzed for new threshold values for low bone density. Areas under the curve (AUC) were calculated with 95% CI.</p><p><strong>Results: </strong>Forty-five patients were included (M/F=10/35, mean age: 63.3). Calibrated vBMDs had stronger correlations with DXA T -scores when compared with HUs, with L2 exhibiting the highest coefficients. Calibration by using A and ES with the threshold of 162 mg/cm 3 had a sensitivity of 90% in detecting low BMD (AUC=0.671).</p><p><strong>Conclusions: </strong>This novel method allows simple, in-office calibration of routine preoperative CT scans without the use of a phantom. Calibration using adipose and erector spinae with a threshold of 162 mg/cm 3 is proposed for low bone density screening with high sensitivity (90%).</p><p><strong>Level of evidence: </strong>Level III.</p>","PeriodicalId":22193,"journal":{"name":"Spine","volume":" ","pages":"E142-E150"},"PeriodicalIF":2.6000,"publicationDate":"2025-04-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Spine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/BRS.0000000000005080","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/26 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Study design: Diagnostic accuracy study.
Objective: To establish a simple method of phantomless bone mineral density (BMD) measurement by using preoperative lumbar computed tomography (CT) scans, and compare the accuracy of reference tissue combinations to diagnose low BMD against uncalibrated Hounsfield units (HUs).
Summary of background data: HUs are used as a measure of BMD; however, associations between HU and T-scores vary widely. Quantitative CT (qCT) scans are more accurate, but they require density calibration with an object of known density (phantom), which limits feasibility. As an emerging technique, phantomless (internal) calibration of routine CT scans may provide a good opportunity for screening.
Materials and methods: Patients who were scheduled to undergo lumbar surgery, with a preoperative CT scan, and a dual-energy x-ray absorptiometry (DXA) scan within six months were included. Four tissues were selected for calibration: subcutaneous adipose (A), erector spinae (ES), psoas (P), and aortic blood (AB). The HUs of these tissues were used in linear regression against ground-truth values. Calibrations were performed by using two different internal tissues at a time to maintain simplicity and in-office applicability.Volumetric bone mineral densities (vBMD) derived from internally calibrated CT scans were analyzed for new threshold values for low bone density. Areas under the curve (AUC) were calculated with 95% CI.
Results: Forty-five patients were included (M/F=10/35, mean age: 63.3). Calibrated vBMDs had stronger correlations with DXA T -scores when compared with HUs, with L2 exhibiting the highest coefficients. Calibration by using A and ES with the threshold of 162 mg/cm 3 had a sensitivity of 90% in detecting low BMD (AUC=0.671).
Conclusions: This novel method allows simple, in-office calibration of routine preoperative CT scans without the use of a phantom. Calibration using adipose and erector spinae with a threshold of 162 mg/cm 3 is proposed for low bone density screening with high sensitivity (90%).
期刊介绍:
Lippincott Williams & Wilkins is a leading international publisher of professional health information for physicians, nurses, specialized clinicians and students. For a complete listing of titles currently published by Lippincott Williams & Wilkins and detailed information about print, online, and other offerings, please visit the LWW Online Store.
Recognized internationally as the leading journal in its field, Spine is an international, peer-reviewed, bi-weekly periodical that considers for publication original articles in the field of Spine. It is the leading subspecialty journal for the treatment of spinal disorders. Only original papers are considered for publication with the understanding that they are contributed solely to Spine. The Journal does not publish articles reporting material that has been reported at length elsewhere.