Dam-Bé L. Douti, Chérif Ouro-Wassara, Essohana Mabafei
{"title":"FriiTriga, a triggered controller with an optical beam shutter using a hard disk drive voice-coil actuator","authors":"Dam-Bé L. Douti, Chérif Ouro-Wassara, Essohana Mabafei","doi":"10.1016/j.ohx.2024.e00542","DOIUrl":null,"url":null,"abstract":"<div><p>We present in this article the complete setup to build a triggered controller with a mechanical optical shutter. The system is a low cost, Do-It-Yourself, and easy to implement setup with three functionalities: Manual mode, direct mode with TTL signal command and Triggered mode with a TTL signal command. Our setup is primarily intended to be integrated in optical setup where one needs to control the opening time of a light path, but can be used also for any other setup where one wants to send a TTL signal to command another subsystem (in our case the shutter is that subsystem). The shutter used here is hard disk drive voice-coil actuator, which was already demonstrated to have interesting potentialities to be a mechanical shutter.</p><p>In the Manual mode, this setup achieves an opening and closing time of 3 ms. In Direct mode with TTL signal command, the setup has a delay response time of 19 ms and a minimum open pulse time of 23 ms. This low cost device which can be made with less than 50€, have similar characteristics with commercial ones which can be twenty times more expensive.</p></div>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-06-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2468067224000361/pdfft?md5=2caef0e9157e28a65c774adc62ab45fe&pid=1-s2.0-S2468067224000361-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2468067224000361","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0
Abstract
We present in this article the complete setup to build a triggered controller with a mechanical optical shutter. The system is a low cost, Do-It-Yourself, and easy to implement setup with three functionalities: Manual mode, direct mode with TTL signal command and Triggered mode with a TTL signal command. Our setup is primarily intended to be integrated in optical setup where one needs to control the opening time of a light path, but can be used also for any other setup where one wants to send a TTL signal to command another subsystem (in our case the shutter is that subsystem). The shutter used here is hard disk drive voice-coil actuator, which was already demonstrated to have interesting potentialities to be a mechanical shutter.
In the Manual mode, this setup achieves an opening and closing time of 3 ms. In Direct mode with TTL signal command, the setup has a delay response time of 19 ms and a minimum open pulse time of 23 ms. This low cost device which can be made with less than 50€, have similar characteristics with commercial ones which can be twenty times more expensive.