Rodrigo Brant, Luis Filipe Nakayama, Talita Virgínia Fernandes de Oliveira, Juliana Angelica Estevão de Oliveira, Lucas Zago Ribeiro, Gabriela Dalmedico Richter, Rafael Rodacki, Fernando Marcondes Penha
{"title":"Image quality comparison of AirDoc portable retina camera versus eyer in a diabetic retinopathy screening program.","authors":"Rodrigo Brant, Luis Filipe Nakayama, Talita Virgínia Fernandes de Oliveira, Juliana Angelica Estevão de Oliveira, Lucas Zago Ribeiro, Gabriela Dalmedico Richter, Rafael Rodacki, Fernando Marcondes Penha","doi":"10.1186/s40942-024-00559-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Diabetic retinopathy (DR) stands as the foremost cause of preventable blindness in adults. Despite efforts to expand DR screening coverage in the Brazilian public healthcare system, challenges persist due to various factors including social, medical, and financial constraints. Our objective was to evaluate the quality of images obtained with the AirDoc, a novel device, compared to Eyer portable camera which has already been clinically validated.</p><p><strong>Methods: </strong>Images were captured by two portable retinal devices: AirDoc and Eyer. The included patients had their fundus images obtained in a screening program conducted in Blumenau, Santa Catarina. Two retina specialists independently assessed image's quality. A comparison was performed between both devices regarding image quality and the presence of artifacts.</p><p><strong>Results: </strong>The analysis included 129 patients (mean age of 61 years), with 29 (43.28%) male and an average disease duration of 11.1 ± 8 years. In Ardoc, 21 (16.28%) images were classified as poor quality, with 88 (68%) presenting artifacts; in Eyer, 4 (3.1%) images were classified as poor quality, with 94 (72.87%) presenting artifacts.</p><p><strong>Conclusions: </strong>Although both Eyer and AirDoc devices show potential as screening tools, the AirDoc images displayed higher rates of ungradable and low-quality images, that may directly affect the DR and DME grading. We must acknowledge the limitations of our study, including the relatively small sample size. Therefore, the interpretations of our analyses should be approached with caution, and further investigations with larger patient cohorts are warranted to validate our findings.</p>","PeriodicalId":14289,"journal":{"name":"International Journal of Retina and Vitreous","volume":null,"pages":null},"PeriodicalIF":1.9000,"publicationDate":"2024-06-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11177418/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Retina and Vitreous","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s40942-024-00559-z","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Diabetic retinopathy (DR) stands as the foremost cause of preventable blindness in adults. Despite efforts to expand DR screening coverage in the Brazilian public healthcare system, challenges persist due to various factors including social, medical, and financial constraints. Our objective was to evaluate the quality of images obtained with the AirDoc, a novel device, compared to Eyer portable camera which has already been clinically validated.
Methods: Images were captured by two portable retinal devices: AirDoc and Eyer. The included patients had their fundus images obtained in a screening program conducted in Blumenau, Santa Catarina. Two retina specialists independently assessed image's quality. A comparison was performed between both devices regarding image quality and the presence of artifacts.
Results: The analysis included 129 patients (mean age of 61 years), with 29 (43.28%) male and an average disease duration of 11.1 ± 8 years. In Ardoc, 21 (16.28%) images were classified as poor quality, with 88 (68%) presenting artifacts; in Eyer, 4 (3.1%) images were classified as poor quality, with 94 (72.87%) presenting artifacts.
Conclusions: Although both Eyer and AirDoc devices show potential as screening tools, the AirDoc images displayed higher rates of ungradable and low-quality images, that may directly affect the DR and DME grading. We must acknowledge the limitations of our study, including the relatively small sample size. Therefore, the interpretations of our analyses should be approached with caution, and further investigations with larger patient cohorts are warranted to validate our findings.
期刊介绍:
International Journal of Retina and Vitreous focuses on the ophthalmic subspecialty of vitreoretinal disorders. The journal presents original articles on new approaches to diagnosis, outcomes of clinical trials, innovations in pharmacological therapy and surgical techniques, as well as basic science advances that impact clinical practice. Topical areas include, but are not limited to: -Imaging of the retina, choroid and vitreous -Innovations in optical coherence tomography (OCT) -Small-gauge vitrectomy, retinal detachment, chromovitrectomy -Electroretinography (ERG), microperimetry, other functional tests -Intraocular tumors -Retinal pharmacotherapy & drug delivery -Diabetic retinopathy & other vascular diseases -Age-related macular degeneration (AMD) & other macular entities