Longevity of different abutments placed on narrow diameter implants: Assessment of structural damage and loosening

IF 4.6 1区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
{"title":"Longevity of different abutments placed on narrow diameter implants: Assessment of structural damage and loosening","authors":"","doi":"10.1016/j.dental.2024.06.003","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><p>To evaluate structural damage and loosening of abutments placed on narrow diameter implants after cyclic fatigue.</p></div><div><h3>Methods</h3><p><span>Sixty Morse taper narrow diameter implants (Neodent, Brazil) received two types of abutments (1PA- one-piece abutment or 2PA- two-piece abutment), which were randomly divided into 3 fatigue experiments (n = 10). The implants were placed into a customized supporting holder and a software-assisted digital torque wrench secured the manufacturer recommended torque for each abutment. Cone beam computed tomography (CBCT) scans were acquired, before and after fatigue, and post-processed (software e-Vol DX) to assess damage and abutment displacement. The boundary fatigue method was adapted to use 2 × 10</span><sup>6</sup> cycles, 2 Hz of frequency and constant peak load of 80 N (first experiment) that varied according to the failure rate of previous specimens (second and third experiments). Failure was evaluated using CBCT scans and removal torque values. Data were used to estimate long-term torque degradation, probability of failure and Weibull modulus (software ALTA PRO9).</p></div><div><h3>Results</h3><p>All 2PA specimens became loosen independently of the applied fatigue load, and structural bending was observed in 14 abutments. Eight 1PA got loosen during the fatigue experiment. The Weibull analysis showed a lower modulus (<em>m</em> = 1.0; 0.7, 1.4) for 1PA than for 2PA (<em>m</em> = 2.6; 2, 3.4) resulting in longer predicted lifetimes and slower torque degradation for 1PA than for 2PA specimens.</p></div><div><h3>Significance</h3><p>1PA showed greater long-term survival probability than 2PA. Predicting the lifetime and mechanical behavior of implant-abutment systems are useful information to clinicians during the decision-making process of oral rehabilitations.</p></div>","PeriodicalId":298,"journal":{"name":"Dental Materials","volume":"40 9","pages":"Pages 1332-1340"},"PeriodicalIF":4.6000,"publicationDate":"2024-06-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dental Materials","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0109564124001337","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives

To evaluate structural damage and loosening of abutments placed on narrow diameter implants after cyclic fatigue.

Methods

Sixty Morse taper narrow diameter implants (Neodent, Brazil) received two types of abutments (1PA- one-piece abutment or 2PA- two-piece abutment), which were randomly divided into 3 fatigue experiments (n = 10). The implants were placed into a customized supporting holder and a software-assisted digital torque wrench secured the manufacturer recommended torque for each abutment. Cone beam computed tomography (CBCT) scans were acquired, before and after fatigue, and post-processed (software e-Vol DX) to assess damage and abutment displacement. The boundary fatigue method was adapted to use 2 × 106 cycles, 2 Hz of frequency and constant peak load of 80 N (first experiment) that varied according to the failure rate of previous specimens (second and third experiments). Failure was evaluated using CBCT scans and removal torque values. Data were used to estimate long-term torque degradation, probability of failure and Weibull modulus (software ALTA PRO9).

Results

All 2PA specimens became loosen independently of the applied fatigue load, and structural bending was observed in 14 abutments. Eight 1PA got loosen during the fatigue experiment. The Weibull analysis showed a lower modulus (m = 1.0; 0.7, 1.4) for 1PA than for 2PA (m = 2.6; 2, 3.4) resulting in longer predicted lifetimes and slower torque degradation for 1PA than for 2PA specimens.

Significance

1PA showed greater long-term survival probability than 2PA. Predicting the lifetime and mechanical behavior of implant-abutment systems are useful information to clinicians during the decision-making process of oral rehabilitations.

窄直径种植体上不同基台的使用寿命:评估结构损伤和松动。
目的评估安装在窄直径种植体上的基台在循环疲劳后的结构损伤和松动情况:60颗莫尔斯锥度窄直径种植体(Neodent,巴西)接受了两种类型的基台(1PA--一片式基台或2PA--两片式基台),随机分为3次疲劳实验(n = 10)。将种植体放入定制的支撑架中,并使用软件辅助的数字扭矩扳手确保每个基台都能达到制造商推荐的扭矩。在疲劳前后采集锥形束计算机断层扫描(CBCT),并进行后处理(软件 e-Vol DX),以评估损伤和基台位移。边界疲劳法采用 2 × 106 次循环、2 Hz 频率和 80 N 恒定峰值载荷(第一次实验),并根据之前试样的失效率而变化(第二次和第三次实验)。使用 CBCT 扫描和移除扭矩值对失效进行评估。数据用于估算长期扭矩衰减、失效概率和威布尔模量(软件 ALTA PRO9):结果:所有 2PA 试件都出现了松动,与所施加的疲劳载荷无关,14 个基台出现了结构弯曲。8 个 1PA 试样在疲劳实验过程中出现松动。Weibull 分析表明,1PA 的模量(m = 1.0; 0.7, 1.4)低于 2PA(m = 2.6; 2, 3.4),因此 1PA 的预测寿命比 2PA 长,扭矩退化速度比 2PA 慢。预测种植体-基台系统的寿命和机械性能对临床医生在口腔修复决策过程中非常有用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Dental Materials
Dental Materials 工程技术-材料科学:生物材料
CiteScore
9.80
自引率
10.00%
发文量
290
审稿时长
67 days
期刊介绍: Dental Materials publishes original research, review articles, and short communications. Academy of Dental Materials members click here to register for free access to Dental Materials online. The principal aim of Dental Materials is to promote rapid communication of scientific information between academia, industry, and the dental practitioner. Original Manuscripts on clinical and laboratory research of basic and applied character which focus on the properties or performance of dental materials or the reaction of host tissues to materials are given priority publication. Other acceptable topics include application technology in clinical dentistry and dental laboratory technology. Comprehensive reviews and editorial commentaries on pertinent subjects will be considered.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信