Alireza Nazempoor, Hosein Poorbaygi, Seyed Mahmud Reza Aghamiri, Mohammad Reza Javanshir, Shahab Sheibani, Reza Naghdi, Somayeh Moradi, Ali Habibpanah
{"title":"Development of a Relative Dosimetric System for Calibration of <sup>32</sup>P Eye Applicators Using Radiochromic Films.","authors":"Alireza Nazempoor, Hosein Poorbaygi, Seyed Mahmud Reza Aghamiri, Mohammad Reza Javanshir, Shahab Sheibani, Reza Naghdi, Somayeh Moradi, Ali Habibpanah","doi":"10.4103/jmp.jmp_99_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Beta irradiation after bare scleral surgery of primary pterygium is an effective and safe treatment, which reduces the risk of local recurrence.</p><p><strong>Purpose: </strong>Obtaining the reference dose rate for a radioactive applicator consisting of a plate as a 32P absorber, a steel window and a steel capsule.</p><p><strong>Methods: </strong>Relative dosimetry and dose profile were measured using two types of radiochromic films, HD-810 and EBT1, for the <sup>32</sup>P applicator and were compared with Monte Carlo simulation data. Dose uniformity in the <sup>32</sup>P applicator was obtained with radiochromic HD-810 film.</p><p><strong>Results: </strong>The measurement depth dose distribution data at distances up to 3.8 mm were compared with calculation data, and the values were not found to differ statistically. Depth dose distribution with a large dose gradient was determined and the dose rate data obtained 0.0053 ± 9.9% in unit of Gy/s.mCi at a 0.1 mm depth distance. Practical results indicated that the dose nonuniformity and the maximum symmetrical for the <sup>32</sup>P applicator were 11.5% and 9.2%, respectively.</p><p><strong>Conclusions: </strong>Our experiments show that the use of the radiochromic film to perform the relative dosimetric checks is feasible and the activity value with acceptable error can be determined through this indirect method.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 1","pages":"49-55"},"PeriodicalIF":0.7000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11141744/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_99_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/30 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Beta irradiation after bare scleral surgery of primary pterygium is an effective and safe treatment, which reduces the risk of local recurrence.
Purpose: Obtaining the reference dose rate for a radioactive applicator consisting of a plate as a 32P absorber, a steel window and a steel capsule.
Methods: Relative dosimetry and dose profile were measured using two types of radiochromic films, HD-810 and EBT1, for the 32P applicator and were compared with Monte Carlo simulation data. Dose uniformity in the 32P applicator was obtained with radiochromic HD-810 film.
Results: The measurement depth dose distribution data at distances up to 3.8 mm were compared with calculation data, and the values were not found to differ statistically. Depth dose distribution with a large dose gradient was determined and the dose rate data obtained 0.0053 ± 9.9% in unit of Gy/s.mCi at a 0.1 mm depth distance. Practical results indicated that the dose nonuniformity and the maximum symmetrical for the 32P applicator were 11.5% and 9.2%, respectively.
Conclusions: Our experiments show that the use of the radiochromic film to perform the relative dosimetric checks is feasible and the activity value with acceptable error can be determined through this indirect method.
期刊介绍:
JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.