Primary Stability and Fracture Resistance of Zirconia and Titanium Implants: A Paired Comparative In Vitro Study.

Olívia Francescato, Ismael Noan Souza Rodrigues, Dhelfeson Willya Douglas de Oliveira, Allyson Nogueira Moreira, Cláudia Silami Magalhães, Frederico Santos Lages
{"title":"Primary Stability and Fracture Resistance of Zirconia and Titanium Implants: A Paired Comparative In Vitro Study.","authors":"Olívia Francescato, Ismael Noan Souza Rodrigues, Dhelfeson Willya Douglas de Oliveira, Allyson Nogueira Moreira, Cláudia Silami Magalhães, Frederico Santos Lages","doi":"10.11607/10974","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To compare the insertion torque (IT) and removal torque (RT) of Neodent Zi zirconia (Zr) implants (Straumann) and Neodent Alvim titanium (Ti) implants (Straumann) while measuring both the maximum torque (MT) supported by the Zr implant and the MT sustained by the assembler of the implants until fracture.</p><p><strong>Materials and methods: </strong>In this in vitro study, 24 implants with the same microgeometry were split into two groups-12 Zr implants and 12 Ti implants-and compared. Implant bed preparations were completed in bovine ribs following a standardized drilling protocol. Then the IT, RT, MT to fracture, and MT to fracture of the assembler were completed using a calibrated torque meter. Data was presented using descriptive statistics including means, SDs, medians, and quartiles. The Shapiro-Wilk test was used to verify data normality and the Wilcoxon test was used to evaluate differences between groups. Statistical significance was established as P < .05.</p><p><strong>Results: </strong>Zr implants showed an IT of 89.33 ± 31.18 Ncm and an RT of 84.89 ± 32.92 Ncm. Ti implants showed an IT of 77.58 ± 28.96 Ncm and an RT of 76.75 ± 31.29 Ncm without significant differences (P > .05). In relation to fracture under rotational force, the Zr implants. fractured at 106.17 ± 22.54 Ncm, and the implant assembly fractured at 84.00 ± 13.14 Ncm.</p><p><strong>Conclusions: </strong>Neodent Zi implants showed stability but a lower fracture torque than the Alvim Ti implants. As the fracture values of the assembler were significantly lower than the fracture values of the Zr implants, it can be stated that, as recommended by the manufacturer, they act as a safety measure during placement.</p>","PeriodicalId":94230,"journal":{"name":"The International journal of oral & maxillofacial implants","volume":"0 0","pages":"321-328"},"PeriodicalIF":0.0000,"publicationDate":"2025-05-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The International journal of oral & maxillofacial implants","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.11607/10974","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To compare the insertion torque (IT) and removal torque (RT) of Neodent Zi zirconia (Zr) implants (Straumann) and Neodent Alvim titanium (Ti) implants (Straumann) while measuring both the maximum torque (MT) supported by the Zr implant and the MT sustained by the assembler of the implants until fracture.

Materials and methods: In this in vitro study, 24 implants with the same microgeometry were split into two groups-12 Zr implants and 12 Ti implants-and compared. Implant bed preparations were completed in bovine ribs following a standardized drilling protocol. Then the IT, RT, MT to fracture, and MT to fracture of the assembler were completed using a calibrated torque meter. Data was presented using descriptive statistics including means, SDs, medians, and quartiles. The Shapiro-Wilk test was used to verify data normality and the Wilcoxon test was used to evaluate differences between groups. Statistical significance was established as P < .05.

Results: Zr implants showed an IT of 89.33 ± 31.18 Ncm and an RT of 84.89 ± 32.92 Ncm. Ti implants showed an IT of 77.58 ± 28.96 Ncm and an RT of 76.75 ± 31.29 Ncm without significant differences (P > .05). In relation to fracture under rotational force, the Zr implants. fractured at 106.17 ± 22.54 Ncm, and the implant assembly fractured at 84.00 ± 13.14 Ncm.

Conclusions: Neodent Zi implants showed stability but a lower fracture torque than the Alvim Ti implants. As the fracture values of the assembler were significantly lower than the fracture values of the Zr implants, it can be stated that, as recommended by the manufacturer, they act as a safety measure during placement.

氧化锆和钛植入体的初级稳定性和抗断裂性:体外配对比较研究
目的:比较 Neodent Zi® 氧化锆种植体与 Neodent Alvim® 钛种植体的插入扭矩(IT)和拔出扭矩(RT)。测量氧化锆种植体在断裂前所承受的最大扭矩(MT)和该种植体装配者在断裂前所承受的最大扭矩(MTA):这项体外研究使用了二十四个种植体。两组具有相同宏观几何形状且来自同一制造商的种植体进行了比较,分别是氧化锆(12 个)和钛(12 个)。按照标准化钻孔方案在牛肋骨上完成种植床准备。然后,使用校准的扭矩仪完成插入扭矩(IT)、取出扭矩(RT)、最大折断扭矩(MT)和装配器最大折断扭矩(MTA)的测量。数据采用描述性统计,包括平均值、标准差 (SD)、中位数和四分位数。Shapiro-Wilk 检验用于验证数据的正态性,Wilcoxon 检验用于评估组间差异。统计学意义以 p < 0.05 为准:结果:氧化锆种植体显示IT 89.33 ± 31.18 Ncm,RT 84.89 ± 32.92 Ncm。钛种植体显示IT 77.58 ± 28.96 Ncm,RT 76.75 ± 31.29 Ncm,无明显差异(p>0.05)。在旋转力的作用下,氧化锆种植体在106.17 ± 22.54 Ncm时断裂,种植体组件在84.00 ± 13.14 Ncm时断裂:结论:与 Alvim® 钛相比,Neodent Zi® 种植体具有稳定性,但断裂扭矩较低。由于装配体的断裂值明显低于氧化锆种植体的断裂值,因此可以说,按照制造商的建议,装配体在安装过程中起到了安全措施的作用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信