Towards monitoring of critical illness via the detection of histones with extended gate field-effect transistor sensors

IF 10.61 Q3 Biochemistry, Genetics and Molecular Biology
Hayley Richardson , Jeffrey Barahona , Greg Medwig , Angela Johns , Lina M. Acosta Pérez , Koji Sode , Michael Daniele , Francis J. Miller , Edgar Lobaton , Spyridon Pavlidis
{"title":"Towards monitoring of critical illness via the detection of histones with extended gate field-effect transistor sensors","authors":"Hayley Richardson ,&nbsp;Jeffrey Barahona ,&nbsp;Greg Medwig ,&nbsp;Angela Johns ,&nbsp;Lina M. Acosta Pérez ,&nbsp;Koji Sode ,&nbsp;Michael Daniele ,&nbsp;Francis J. Miller ,&nbsp;Edgar Lobaton ,&nbsp;Spyridon Pavlidis","doi":"10.1016/j.biosx.2024.100501","DOIUrl":null,"url":null,"abstract":"<div><p>Extracellular histone proteins in the blood indicate a heightened risk of morbidity after trauma or in major illnesses such as sepsis. We present the development of an aptasensor for histone detection with an extended gate field-effect transistor (EGFET) configuration, which benefits from low power consumption, rapid response, and compatibility with miniaturized gold electrodes. Histones have a high isoelectric point and charge density, which cause them to physically adsorb to non-specific elements of the sensor that have available electrostatic charges. To combat this, the sensing surface is formed with a thiol-modified, high-affinity and histone-specific RNA aptamer sequence and by co-immobilizing with poly(ethylene glycol) methyl ether thiol (PEG) as a blocking agent. Surface plasmon resonance (SPR) is used to analyze aptamer and PEG immobilization strategies, confirm histone binding, and calculate kinetic binding constants. Through comparison of different blocking agents and time-dependent preparation, the ideal equilibrium dissociation constant (K<sub>D</sub>) is estimated to be below 200 pM, which is the upper range of extracellular histone concentrations in critically ill patients with high mortality. The EGFET sensitivity of the optimized aptasensor is 6.65 mV/decade concentration change for histone H4 with a physiologically relevant 5 pM limit of detection. Selectivity tests with 100 nM bovine serum albumin (BSA) demonstrate a signal response that is 13-fold smaller than for histones. This EGFET aptasensor platform is suitable for future point-of-care monitoring of histone levels in critically ill patients, thus permitting the early detection of increased risk and the need for more aggressive interventional measures to prevent mortality.</p></div>","PeriodicalId":260,"journal":{"name":"Biosensors and Bioelectronics: X","volume":"19 ","pages":"Article 100501"},"PeriodicalIF":10.6100,"publicationDate":"2024-05-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2590137024000657/pdfft?md5=5abf3262c26bbda68acae06aba70ee43&pid=1-s2.0-S2590137024000657-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biosensors and Bioelectronics: X","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2590137024000657","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Biochemistry, Genetics and Molecular Biology","Score":null,"Total":0}
引用次数: 0

Abstract

Extracellular histone proteins in the blood indicate a heightened risk of morbidity after trauma or in major illnesses such as sepsis. We present the development of an aptasensor for histone detection with an extended gate field-effect transistor (EGFET) configuration, which benefits from low power consumption, rapid response, and compatibility with miniaturized gold electrodes. Histones have a high isoelectric point and charge density, which cause them to physically adsorb to non-specific elements of the sensor that have available electrostatic charges. To combat this, the sensing surface is formed with a thiol-modified, high-affinity and histone-specific RNA aptamer sequence and by co-immobilizing with poly(ethylene glycol) methyl ether thiol (PEG) as a blocking agent. Surface plasmon resonance (SPR) is used to analyze aptamer and PEG immobilization strategies, confirm histone binding, and calculate kinetic binding constants. Through comparison of different blocking agents and time-dependent preparation, the ideal equilibrium dissociation constant (KD) is estimated to be below 200 pM, which is the upper range of extracellular histone concentrations in critically ill patients with high mortality. The EGFET sensitivity of the optimized aptasensor is 6.65 mV/decade concentration change for histone H4 with a physiologically relevant 5 pM limit of detection. Selectivity tests with 100 nM bovine serum albumin (BSA) demonstrate a signal response that is 13-fold smaller than for histones. This EGFET aptasensor platform is suitable for future point-of-care monitoring of histone levels in critically ill patients, thus permitting the early detection of increased risk and the need for more aggressive interventional measures to prevent mortality.

利用扩展栅场效应晶体管传感器检测组蛋白,实现危重病监测
血液中的细胞外组蛋白预示着创伤后或败血症等重大疾病的发病风险增高。我们介绍了利用扩展栅场效应晶体管(EGFET)配置开发的组蛋白检测灵敏传感器,该传感器具有功耗低、响应快以及与微型金电极兼容等优点。组蛋白具有较高的等电点和电荷密度,因此会吸附在传感器上具有可用静电荷的非特异性元件上。为解决这一问题,传感表面由硫醇修饰的高亲和性组蛋白特异性 RNA 合酶序列和作为阻断剂的聚乙二醇甲醚硫醇(PEG)共同固定而成。表面等离子体共振(SPR)用于分析适配体和 PEG 的固定策略、确认组蛋白结合以及计算动力学结合常数。通过比较不同的阻断剂和随时间变化的制备方法,理想的平衡解离常数(KD)估计低于 200 pM,而这正是死亡率较高的重症患者细胞外组蛋白浓度的上限范围。对组蛋白 H4 来说,优化后的适配传感器的 EGFET 灵敏度为 6.65 mV/十进制浓度变化,生理检测限为 5 pM。用 100 nM 牛血清白蛋白(BSA)进行的选择性测试表明,信号响应比组蛋白小 13 倍。这种 EGFET aptasensor 平台适用于未来对危重病人组蛋白水平的护理点监测,从而能够及早发现风险增加的情况,并需要采取更积极的干预措施来预防死亡。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Biosensors and Bioelectronics: X
Biosensors and Bioelectronics: X Biochemistry, Genetics and Molecular Biology-Biophysics
CiteScore
4.60
自引率
0.00%
发文量
166
审稿时长
54 days
期刊介绍: Biosensors and Bioelectronics: X, an open-access companion journal of Biosensors and Bioelectronics, boasts a 2020 Impact Factor of 10.61 (Journal Citation Reports, Clarivate Analytics 2021). Offering authors the opportunity to share their innovative work freely and globally, Biosensors and Bioelectronics: X aims to be a timely and permanent source of information. The journal publishes original research papers, review articles, communications, editorial highlights, perspectives, opinions, and commentaries at the intersection of technological advancements and high-impact applications. Manuscripts submitted to Biosensors and Bioelectronics: X are assessed based on originality and innovation in technology development or applications, aligning with the journal's goal to cater to a broad audience interested in this dynamic field.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信