The Performance of Spot Photoscreener in 6 to 10 Weeks Infants in China: A Cross-Sectional Study

IF 1.8 4区 医学 Q3 OPHTHALMOLOGY
Yaoling Li, Jing Li, Huiyu Wang, Mingyang Du, Lirong Wei, Teng Su, Gang Ding, Xuehan Qian, Ning Hua
{"title":"The Performance of Spot Photoscreener in 6 to 10 Weeks Infants in China: A Cross-Sectional Study","authors":"Yaoling Li, Jing Li, Huiyu Wang, Mingyang Du, Lirong Wei, Teng Su, Gang Ding, Xuehan Qian, Ning Hua","doi":"10.1155/2024/8817530","DOIUrl":null,"url":null,"abstract":"<i>Purpose</i>. To compare the refractive errors measured by the Spot photoscreener (with or without cycloplegia) to cycloplegic retinoscopy in 6- to 10-week-old infants. <i>Materials and Methods</i>. 101 right eyes from 101 healthy infants aged 6 to 10 weeks were recruited for this cross-sectional observational study. Refractive errors were measured using Spot photoscreener before and after cycloplegia, as well as cycloplegic retinoscopy. Comparisons between the refractive measurements were performed using one-way ANOVA with the post hoc Tukey HSD test or Kruskal–Wallis test with the Steel–Dwass test according to the data normality. Pearson’s correlation test and 95% confidence intervals were calculated. The agreement was evaluated using a Bland–Altman plot with 95% limits of agreement of the differences. <i>Results</i>. Spot photoscreener was found to underestimate the spherical equivalent by 2.33 Diopters (D) in these infants. Following the induction of cycloplegia, the spherical equivalent measured by Spot photoscreener was in excellent agreement with cycloplegic retinoscopy with the mean difference of 0.01 D. Spot photoscreener overestimated cylindrical parameter by 0.2 D with poor agreement with cycloplegic retinoscopy no matter whether cycloplegia was induced. It had good agreement with cycloplegic retinoscopy in the <svg height=\"11.927pt\" style=\"vertical-align:-3.291101pt\" version=\"1.1\" viewbox=\"-0.0498162 -8.6359 10.0959 11.927\" width=\"10.0959pt\" xmlns=\"http://www.w3.org/2000/svg\" xmlns:xlink=\"http://www.w3.org/1999/xlink\"><g transform=\"matrix(.013,0,0,-0.013,0,0)\"></path></g><g transform=\"matrix(.0091,0,0,-0.0091,5.031,3.132)\"></path></g></svg> vector than the <svg height=\"11.927pt\" style=\"vertical-align:-3.291101pt\" version=\"1.1\" viewbox=\"-0.0498162 -8.6359 14.5442 11.927\" width=\"14.5442pt\" xmlns=\"http://www.w3.org/2000/svg\" xmlns:xlink=\"http://www.w3.org/1999/xlink\"><g transform=\"matrix(.013,0,0,-0.013,0,0)\"><use xlink:href=\"#g113-75\"></use></g><g transform=\"matrix(.0091,0,0,-0.0091,5.031,3.132)\"></path></g><g transform=\"matrix(.0091,0,0,-0.0091,9.463,3.132)\"></path></g></svg> vector measurement. <i>Conclusions</i>. With the induction of cycloplegia, Spot photoscreener can accurately evaluate spherical equivalent in hyperopic infants with mild-to-moderate astigmatism. While it may provide valuable measurements of astigmatism, discrepancies in cylinder and axis should be taken into account.","PeriodicalId":16674,"journal":{"name":"Journal of Ophthalmology","volume":"21 1","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2024-05-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1155/2024/8817530","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose. To compare the refractive errors measured by the Spot photoscreener (with or without cycloplegia) to cycloplegic retinoscopy in 6- to 10-week-old infants. Materials and Methods. 101 right eyes from 101 healthy infants aged 6 to 10 weeks were recruited for this cross-sectional observational study. Refractive errors were measured using Spot photoscreener before and after cycloplegia, as well as cycloplegic retinoscopy. Comparisons between the refractive measurements were performed using one-way ANOVA with the post hoc Tukey HSD test or Kruskal–Wallis test with the Steel–Dwass test according to the data normality. Pearson’s correlation test and 95% confidence intervals were calculated. The agreement was evaluated using a Bland–Altman plot with 95% limits of agreement of the differences. Results. Spot photoscreener was found to underestimate the spherical equivalent by 2.33 Diopters (D) in these infants. Following the induction of cycloplegia, the spherical equivalent measured by Spot photoscreener was in excellent agreement with cycloplegic retinoscopy with the mean difference of 0.01 D. Spot photoscreener overestimated cylindrical parameter by 0.2 D with poor agreement with cycloplegic retinoscopy no matter whether cycloplegia was induced. It had good agreement with cycloplegic retinoscopy in the vector than the vector measurement. Conclusions. With the induction of cycloplegia, Spot photoscreener can accurately evaluate spherical equivalent in hyperopic infants with mild-to-moderate astigmatism. While it may provide valuable measurements of astigmatism, discrepancies in cylinder and axis should be taken into account.
中国 6 至 10 周婴儿的光斑照度计性能:横断面研究
目的在 6 到 10 周大的婴儿中,比较用 Spot 光筛仪(带或不带环形屈光镜)和环形屈光镜视网膜镜测得的屈光不正。材料和方法。这项横断面观察研究招募了 101 名 6 至 10 周大健康婴儿的 101 只右眼。在环形屈光手术前后使用点光筛查仪测量屈光不正,并进行环形屈光视网膜镜检查。根据数据的正态性,采用单因素方差分析和事后Tukey HSD检验或Kruskal-Wallis检验和Steel-Dwass检验对屈光测量结果进行比较。计算了皮尔逊相关性检验和 95% 的置信区间。使用布兰德-阿尔特曼图对差异的 95% 一致度进行评估。结果。发现点光筛查仪将这些婴儿的球面等效度数低估了 2.33 屈光度 (D)。在诱导环形角膜塑形术后,点阵光度计测量的球面等效度数与环形角膜塑形术视网膜镜测得的球面等效度数非常一致,平均相差 0.01 D。无论是否诱导了屈光不正,点阵光度计都会高估圆柱参数 0.2 D,与屈光视网膜镜的一致性较差。在矢量测量中,它与环形视网膜镜的一致性要好于矢量测量。结论。在诱导屈光不正的情况下,点光筛可以准确评估轻度至中度散光的远视婴儿的球面等值。虽然它可以提供有价值的散光测量值,但应考虑到柱面和轴线的差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Ophthalmology
Journal of Ophthalmology MEDICINE, RESEARCH & EXPERIMENTAL-OPHTHALMOLOGY
CiteScore
4.30
自引率
5.30%
发文量
194
审稿时长
6-12 weeks
期刊介绍: Journal of Ophthalmology is a peer-reviewed, Open Access journal that publishes original research articles, review articles, and clinical studies related to the anatomy, physiology and diseases of the eye. Submissions should focus on new diagnostic and surgical techniques, instrument and therapy updates, as well as clinical trials and research findings.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信