Electroacoustic evaluation of the bone conduction transducer B250 for vestibular and hearing diagnostics in comparison with Radioear B71 and B81.

IF 1.8 3区 医学 Q2 AUDIOLOGY & SPEECH-LANGUAGE PATHOLOGY
International Journal of Audiology Pub Date : 2025-04-01 Epub Date: 2024-05-14 DOI:10.1080/14992027.2024.2352054
Karl-Johan Fredén Jansson, Bo Håkansson, Ann-Charlotte Persson, Luca Verrecchia, Sabine Reinfeldt
{"title":"Electroacoustic evaluation of the bone conduction transducer B250 for vestibular and hearing diagnostics in comparison with Radioear B71 and B81.","authors":"Karl-Johan Fredén Jansson, Bo Håkansson, Ann-Charlotte Persson, Luca Verrecchia, Sabine Reinfeldt","doi":"10.1080/14992027.2024.2352054","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The objective is to evaluate the electroacoustic performance of the B250 transducer and to compare it with the two most widely used audiometric transducers B71 and B81.</p><p><strong>Design: </strong>The electroacoustic performance was evaluated in terms of sensitivity level, distortion, maximum hearing level and electrical impedance.</p><p><strong>Study sample: </strong>Six B250 prototype transducers were evaluated and compared with published data of B71 and B81 together with complementary measurements of maximum hearing level at 125 Hz and phase of electrical impedance. Differences in reference equivalent threshold vibratory force levels were estimated by comparing hearing threshold measurements of 60 healthy ears using B81 and B250.</p><p><strong>Results: </strong>B250 has approximately 27 dB higher sensitivity levels than both B71 and B81 at 250 Hz and can generate higher maximum hearing level at low frequencies: 11.8 to 35.8 dB (125-1000 Hz) higher than B71, and 1.4 to 18.6 dB (125-750 Hz) higher than B81. The maximum average difference in reference threshold force levels was 13.5 ± 8.7 dB higher for B250 at 250 Hz compared to B81.</p><p><strong>Conclusions: </strong>B250 can produce higher output force with less distortion than B71 and B81, especially at 125 and 250 Hz, which could possibly improve low frequency investigations of the audio-vestibular system.</p>","PeriodicalId":13759,"journal":{"name":"International Journal of Audiology","volume":" ","pages":"320-326"},"PeriodicalIF":1.8000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Audiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/14992027.2024.2352054","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/14 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"AUDIOLOGY & SPEECH-LANGUAGE PATHOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: The objective is to evaluate the electroacoustic performance of the B250 transducer and to compare it with the two most widely used audiometric transducers B71 and B81.

Design: The electroacoustic performance was evaluated in terms of sensitivity level, distortion, maximum hearing level and electrical impedance.

Study sample: Six B250 prototype transducers were evaluated and compared with published data of B71 and B81 together with complementary measurements of maximum hearing level at 125 Hz and phase of electrical impedance. Differences in reference equivalent threshold vibratory force levels were estimated by comparing hearing threshold measurements of 60 healthy ears using B81 and B250.

Results: B250 has approximately 27 dB higher sensitivity levels than both B71 and B81 at 250 Hz and can generate higher maximum hearing level at low frequencies: 11.8 to 35.8 dB (125-1000 Hz) higher than B71, and 1.4 to 18.6 dB (125-750 Hz) higher than B81. The maximum average difference in reference threshold force levels was 13.5 ± 8.7 dB higher for B250 at 250 Hz compared to B81.

Conclusions: B250 can produce higher output force with less distortion than B71 and B81, especially at 125 and 250 Hz, which could possibly improve low frequency investigations of the audio-vestibular system.

用于前庭和听力诊断的骨传导传感器 B250 与 Radioear B71 和 B81 的电声评估比较。
目标:目的是评估 B250 传感器的电声性能,并将其与两种最广泛使用的听力传感器 B71 和 B81 进行比较:设计:从灵敏度水平、失真、最大听力水平和电阻抗等方面对电声性能进行评估:研究样本:对 6 个 B250 原型传感器进行了评估,并将其与 B71 和 B81 的公开数据以及 125 Hz 最大听力水平和电阻抗相位的补充测量结果进行了比较。通过比较使用 B81 和 B250 对 60 只健康耳朵进行的听阈测量结果,估算出参考等效阈值振动力水平的差异:结果:在 250 赫兹时,B250 的灵敏度比 B71 和 B81 高约 27 分贝,在低频时能产生更高的最大听力水平:比 B71 高 11.8 至 35.8 分贝(125-1000 赫兹),比 B81 高 1.4 至 18.6 分贝(125-750 赫兹)。与 B81 相比,B250 在 250 Hz 时的参考阈值力水平的最大平均差异为 13.5 ± 8.7 dB:结论:与 B71 和 B81 相比,B250 可以产生更高的输出力,失真更小,尤其是在 125 和 250 Hz 时,这可能会改善对听觉前庭系统的低频调查。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
International Journal of Audiology
International Journal of Audiology 医学-耳鼻喉科学
CiteScore
4.90
自引率
14.80%
发文量
133
审稿时长
4-8 weeks
期刊介绍: International Journal of Audiology is committed to furthering development of a scientifically robust evidence base for audiology. The journal is published by the British Society of Audiology, the International Society of Audiology and the Nordic Audiological Society.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信