Yeabsira Mesfin BS , Alan Kong MD , Benjamin T. Backus PhD , Michael Deiner PhD , Yvonne Ou MD , Julius T. Oatts MD
{"title":"Pilot study comparing a new virtual reality–based visual field test to standard perimetry in children","authors":"Yeabsira Mesfin BS , Alan Kong MD , Benjamin T. Backus PhD , Michael Deiner PhD , Yvonne Ou MD , Julius T. Oatts MD","doi":"10.1016/j.jaapos.2024.103933","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>To assess the feasibility and performance of Vivid Vision Perimetry (VVP), a new virtual reality (VR)–based visual field platform.</p></div><div><h3>Methods</h3><p>Children 7-18 years of age with visual acuity of 20/80 or better undergoing Humphrey visual field (HVF) testing were recruited to perform VVP, a VR-based test that uses suprathreshold stimuli to test 54 field locations and calculates a fraction seen score. Pearson correlation coefficients were calculated to evaluate correlation between HVF mean sensitivity and VVP mean fraction seen scores. Participants were surveyed regarding their experience.</p></div><div><h3>Results</h3><p>A total of 37 eyes of 23 participants (average age, 12.9 ± 3.1 years; 48% female) were included. All participants successfully completed VVP testing. Diagnoses included glaucoma (12), glaucoma suspect (7), steroid-induced ocular hypertension (3), and craniopharyngioma (1). Sixteen participants had prior HVF experience, and none had prior VVP experience, although 7 had previously used VR. Of the 23 HVF tests performed, 9 (39%) were unreliable due to fixation losses, false positives, or false negatives. Similarly, 35% of VVP tests were unreliable (as defined by accuracy of blind spot detection). Excluding unreliable HVF tests, the correlation between HVF average mean sensitivity and VVP mean fraction seen score was 0.48 (<em>P</em> = 0.02; 95% CI, 0.09-0.74). When asked about preference for the VVP or HVF examination, all participants favored the VVP, and 70% were “very satisfied” with VVP.</p></div><div><h3>Conclusions</h3><p>In our cohort of 23 pediatric subjects, VVP proved to be a clinically feasible VR-based visual field testing, which was uniformly preferred over HVF.</p></div>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S1091853124002039/pdfft?md5=b3a481391874f35f579ee5a209a3984f&pid=1-s2.0-S1091853124002039-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1091853124002039","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose
To assess the feasibility and performance of Vivid Vision Perimetry (VVP), a new virtual reality (VR)–based visual field platform.
Methods
Children 7-18 years of age with visual acuity of 20/80 or better undergoing Humphrey visual field (HVF) testing were recruited to perform VVP, a VR-based test that uses suprathreshold stimuli to test 54 field locations and calculates a fraction seen score. Pearson correlation coefficients were calculated to evaluate correlation between HVF mean sensitivity and VVP mean fraction seen scores. Participants were surveyed regarding their experience.
Results
A total of 37 eyes of 23 participants (average age, 12.9 ± 3.1 years; 48% female) were included. All participants successfully completed VVP testing. Diagnoses included glaucoma (12), glaucoma suspect (7), steroid-induced ocular hypertension (3), and craniopharyngioma (1). Sixteen participants had prior HVF experience, and none had prior VVP experience, although 7 had previously used VR. Of the 23 HVF tests performed, 9 (39%) were unreliable due to fixation losses, false positives, or false negatives. Similarly, 35% of VVP tests were unreliable (as defined by accuracy of blind spot detection). Excluding unreliable HVF tests, the correlation between HVF average mean sensitivity and VVP mean fraction seen score was 0.48 (P = 0.02; 95% CI, 0.09-0.74). When asked about preference for the VVP or HVF examination, all participants favored the VVP, and 70% were “very satisfied” with VVP.
Conclusions
In our cohort of 23 pediatric subjects, VVP proved to be a clinically feasible VR-based visual field testing, which was uniformly preferred over HVF.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.