{"title":"Mobile infrared slit-light scanner for rapid eye disease screening.","authors":"Neelam Kaushik, Parmanand Sharma, Noriko Himori, Takuro Matsumoto, Takehiro Miya, Toru Nakazawa","doi":"10.1117/1.JMI.11.2.026003","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Timely detection and treatment of visual impairments and age-related eye diseases are essential for maintaining a longer, healthier life. However, the shortage of appropriate medical equipment often impedes early detection. We have developed a portable self-imaging slit-light device utilizing NIR light and a scanning mirror. The objective of our study is to assess the accuracy and compare the performance of our device with conventional nonportable slit-lamp microscopes and anterior segment optical coherence tomography (AS-OCT) for screening and remotely diagnosing eye diseases, such as cataracts and glaucoma, outside of an eye clinic.</p><p><strong>Approach: </strong>The NIR light provides an advantage as measurements are nonmydriatic and less traumatic for patients. A cross-sectional study involving Japanese adults was conducted. Cataract evaluation was performed using photographs captured by the device. Van-Herick grading was assessed by the ratio of peripheral anterior chamber depth to peripheral corneal thickness, in addition to the iridocorneal angle using Image J software.</p><p><strong>Results: </strong>The correlation coefficient between values obtained by AS-OCT, and our fabricated portable scanning slit-light device was notably high. The results indicate that our portable device is equally reliable as the conventional nonportable slit-lamp microscope and AS-OCT for screening and evaluating eye diseases.</p><p><strong>Conclusions: </strong>Our fabricated device matches the functionality of the traditional slit lamp, offering a cost-effective and portable solution. Ideal for remote locations, healthcare facilities, or areas affected by disasters, our scanning slit-light device can provide easy access to initial eye examinations and supports digital eye healthcare initiatives.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2024-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11003872/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1117/1.JMI.11.2.026003","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/4/10 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Timely detection and treatment of visual impairments and age-related eye diseases are essential for maintaining a longer, healthier life. However, the shortage of appropriate medical equipment often impedes early detection. We have developed a portable self-imaging slit-light device utilizing NIR light and a scanning mirror. The objective of our study is to assess the accuracy and compare the performance of our device with conventional nonportable slit-lamp microscopes and anterior segment optical coherence tomography (AS-OCT) for screening and remotely diagnosing eye diseases, such as cataracts and glaucoma, outside of an eye clinic.
Approach: The NIR light provides an advantage as measurements are nonmydriatic and less traumatic for patients. A cross-sectional study involving Japanese adults was conducted. Cataract evaluation was performed using photographs captured by the device. Van-Herick grading was assessed by the ratio of peripheral anterior chamber depth to peripheral corneal thickness, in addition to the iridocorneal angle using Image J software.
Results: The correlation coefficient between values obtained by AS-OCT, and our fabricated portable scanning slit-light device was notably high. The results indicate that our portable device is equally reliable as the conventional nonportable slit-lamp microscope and AS-OCT for screening and evaluating eye diseases.
Conclusions: Our fabricated device matches the functionality of the traditional slit lamp, offering a cost-effective and portable solution. Ideal for remote locations, healthcare facilities, or areas affected by disasters, our scanning slit-light device can provide easy access to initial eye examinations and supports digital eye healthcare initiatives.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.