Mechanical Testing of Sliding on Pivot-Locking Clamp (SOP-LC) Fracture Repair System in Four-Point Bending and Torsion.

IF 1 2区 农林科学 Q3 VETERINARY SCIENCES
Angel Miller
{"title":"Mechanical Testing of Sliding on Pivot-Locking Clamp (SOP-LC) Fracture Repair System in Four-Point Bending and Torsion.","authors":"Angel Miller","doi":"10.1055/s-0043-1778682","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong> The first objective of this study was to compare the mechanical performance of a straight and contoured rod for the sliding on pivot-locking clamp (SOP-LC) system. The second objective was to compare single-cycle bending and torsion of the SOP-LC rod-bone constructs to a bone gap model between two clamp configurations.</p><p><strong>Study design: </strong> First experiment, 5 mm diameter rods from the 3.5 mm SOP-LC system, 4 straight and 4 bent rods were mechanically compared in four-point bending. The second experiment, 8 constructs with clamps on a single side of the rod and 8 constructs with clamps on alternating sides were compared in four-point bending and torsion. A torque limiting screwdriver at 2.5N was used for all constructs.</p><p><strong>Results: </strong> There were not significant differences between the noncontoured versus contoured rods. Constructs with clamps on a single side had a significantly higher yield load, yield displacement and bending strength, but the constructs with alternating clamps had a significantly higher initial torsional stiffness. There were no other significant differences. Mild clamp slippage and mild screw bending were noted in all the torsion tested constructs graphically with multiple stiffness slopes and visually on the post-tested constructs.</p><p><strong>Conclusion: </strong> Torque applied to initial screw placement may play a role in reducing clamp slippage. Although bending yield load, bending yield displacement, bending strength and torsional stiffness were improved by different clamp placement, clinically this will depend on fracture type. Further studies are needed for clinical significance.</p>","PeriodicalId":51204,"journal":{"name":"Veterinary and Comparative Orthopaedics and Traumatology","volume":" ","pages":"163-172"},"PeriodicalIF":1.0000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Veterinary and Comparative Orthopaedics and Traumatology","FirstCategoryId":"97","ListUrlMain":"https://doi.org/10.1055/s-0043-1778682","RegionNum":2,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/2 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives:  The first objective of this study was to compare the mechanical performance of a straight and contoured rod for the sliding on pivot-locking clamp (SOP-LC) system. The second objective was to compare single-cycle bending and torsion of the SOP-LC rod-bone constructs to a bone gap model between two clamp configurations.

Study design:  First experiment, 5 mm diameter rods from the 3.5 mm SOP-LC system, 4 straight and 4 bent rods were mechanically compared in four-point bending. The second experiment, 8 constructs with clamps on a single side of the rod and 8 constructs with clamps on alternating sides were compared in four-point bending and torsion. A torque limiting screwdriver at 2.5N was used for all constructs.

Results:  There were not significant differences between the noncontoured versus contoured rods. Constructs with clamps on a single side had a significantly higher yield load, yield displacement and bending strength, but the constructs with alternating clamps had a significantly higher initial torsional stiffness. There were no other significant differences. Mild clamp slippage and mild screw bending were noted in all the torsion tested constructs graphically with multiple stiffness slopes and visually on the post-tested constructs.

Conclusion:  Torque applied to initial screw placement may play a role in reducing clamp slippage. Although bending yield load, bending yield displacement, bending strength and torsional stiffness were improved by different clamp placement, clinically this will depend on fracture type. Further studies are needed for clinical significance.

四点弯曲和扭转滑动枢轴锁定夹(SOP-LC)断裂修复系统的机械测试。
研究目的本研究的第一个目标是比较用于滑动枢轴锁定夹钳(SOP-LC)系统的直杆和轮廓杆的机械性能。第二个目标是比较 SOP-LC 杆骨结构的单周期弯曲和扭转,以及两种夹钳配置之间的骨间隙模型:第一项实验:对来自 3.5 毫米 SOP-LC 系统的 5 毫米直径杆、4 根直杆和 4 根弯曲杆进行四点弯曲机械比较。第二项实验:在四点弯曲和扭转中对 8 个在杆的单侧夹紧的结构和 8 个在交替两侧夹紧的结构进行比较。所有结构都使用了扭矩限制为 2.5N 的螺丝刀:结果:无轮廓杆与有轮廓杆之间没有明显差异。单侧夹具结构的屈服载荷、屈服位移和弯曲强度明显更高,但交替夹具结构的初始扭转刚度明显更高。其他方面没有明显差异。所有经过扭转测试的结构都出现了轻微的夹钳滑动和轻微的螺钉弯曲:结论:最初置入螺钉时施加的扭矩可能会在减少夹钳滑动方面发挥作用。虽然不同的夹钳置放位置可改善弯曲屈服载荷、弯曲屈服位移、弯曲强度和扭转刚度,但在临床上这将取决于骨折类型。临床意义还需进一步研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.00
自引率
15.40%
发文量
49
审稿时长
18-36 weeks
期刊介绍: Veterinary and Comparative Orthopaedics and Traumatology (VCOT) is the most important single source for clinically relevant information in orthopaedics and neurosurgery available anywhere in the world today. It is unique in that it is truly comparative and there is an unrivalled mix of review articles and basic science amid the information that is immediately clinically relevant in veterinary surgery today.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信