Annie E Moroco, Zhikui Wei, Israel Byrd, Andrea Rasmussen, Eugene G Chio, Ryan J Soose, Phillip Huyett, Armin Steffen, Clemens Heiser, Maurits S Boon, Colin T Huntley
{"title":"Device-related outcomes following hypoglossal nerve stimulator implantation.","authors":"Annie E Moroco, Zhikui Wei, Israel Byrd, Andrea Rasmussen, Eugene G Chio, Ryan J Soose, Phillip Huyett, Armin Steffen, Clemens Heiser, Maurits S Boon, Colin T Huntley","doi":"10.5664/jcsm.11176","DOIUrl":null,"url":null,"abstract":"<p><strong>Study objectives: </strong>Hypoglossal nerve stimulation (HGNS) has been widely used to treat obstructive sleep apnea in selected patients. Here we evaluate rates of revision and explant related to HGNS implantation and assess types of adverse events contributing to revision and explant.</p><p><strong>Methods: </strong>Postmarket surveillance data for HGNS implanted between January 1, 2018 and March 31, 2022, were collected. Event rates and risk were calculated using the postmarket surveillance event counts and sales volume over the same period. Indications were categorized for analysis. Descriptive statistics were reported and freedom from explant or revision curves were grouped by year of implantation.</p><p><strong>Results: </strong>Of the 20,881 HGNS implants assessed, rates of explant and revision within the first year were 0.723% and 1.542%, respectively. The most common indication for explant was infection (0.378%) and for revision was surgical correction (0.680%). Of the 5,820 devices with 3-year postimplantation data, the rate of explant was 2.680% and of revision was 3.557%. During this same interval, elective removal (1.478%) was the most common indication, and for revisions, surgical correction (1.134%).</p><p><strong>Conclusions: </strong>The efficacy of HGNS is comparable in the real world setting to published clinical trial data. Rates of explant and revision are low, supporting a satisfactory safety profile for this technology.</p><p><strong>Citation: </strong>Moroco AE, Wei Z, Byrd I, et al. Device-related outcomes following hypoglossal nerve stimulator implantation. <i>J Clin Sleep Med</i>. 2024;20(9):1497-1503.</p>","PeriodicalId":50233,"journal":{"name":"Journal of Clinical Sleep Medicine","volume":null,"pages":null},"PeriodicalIF":3.5000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11367714/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Sleep Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5664/jcsm.11176","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Study objectives: Hypoglossal nerve stimulation (HGNS) has been widely used to treat obstructive sleep apnea in selected patients. Here we evaluate rates of revision and explant related to HGNS implantation and assess types of adverse events contributing to revision and explant.
Methods: Postmarket surveillance data for HGNS implanted between January 1, 2018 and March 31, 2022, were collected. Event rates and risk were calculated using the postmarket surveillance event counts and sales volume over the same period. Indications were categorized for analysis. Descriptive statistics were reported and freedom from explant or revision curves were grouped by year of implantation.
Results: Of the 20,881 HGNS implants assessed, rates of explant and revision within the first year were 0.723% and 1.542%, respectively. The most common indication for explant was infection (0.378%) and for revision was surgical correction (0.680%). Of the 5,820 devices with 3-year postimplantation data, the rate of explant was 2.680% and of revision was 3.557%. During this same interval, elective removal (1.478%) was the most common indication, and for revisions, surgical correction (1.134%).
Conclusions: The efficacy of HGNS is comparable in the real world setting to published clinical trial data. Rates of explant and revision are low, supporting a satisfactory safety profile for this technology.
Citation: Moroco AE, Wei Z, Byrd I, et al. Device-related outcomes following hypoglossal nerve stimulator implantation. J Clin Sleep Med. 2024;20(9):1497-1503.
期刊介绍:
Journal of Clinical Sleep Medicine focuses on clinical sleep medicine. Its emphasis is publication of papers with direct applicability and/or relevance to the clinical practice of sleep medicine. This includes clinical trials, clinical reviews, clinical commentary and debate, medical economic/practice perspectives, case series and novel/interesting case reports. In addition, the journal will publish proceedings from conferences, workshops and symposia sponsored by the American Academy of Sleep Medicine or other organizations related to improving the practice of sleep medicine.