{"title":"[Evaluation of a new portable analyzer, HemoCue<sup>®</sup> Hb801, for hemoglobin point-of-care testing].","authors":"Bryan Zamorano, Yolla Sakr, Frédérique Galopin-Dubois","doi":"10.1684/abc.2024.1868","DOIUrl":null,"url":null,"abstract":"<p><p>The use of portable hemoglobin measuring devices is widespread. In this context, the company HemoCue® has put on the market a new device, the Hb801. It uses a whole blood absorbance measurement method and not the azidmethemoglobin measurement method used by HemoCue's older devices. We evaluated this new equipment on EDTA venous blood. Hb801 is lightweight, compact, requires a volume of 10 μL of blood and renders its result in less than a second. The repeatability and intermediate precision are close to the values expected according to Ricos, with coefficients of variation respectively for a low level of hemoglobin: 2.1% and 1.9%, for an average level: 0.8% and 1.5% and for a high level: 1.5% and 1.3%. Comparison to our laboratory reference method (XN-10 Sysmex®) and HemoCue® Hb201+ was performed on 96 samples. Bias (SD) found were: XN-10: +0.42 g/dL (0.17), HemoCue® Hb201+: +0.17 g/dL (0.41). Clinically acceptable performance (within ± 1 g/dL of reference hemoglobin) was high: 93.8%. In the end, this device seems to us to be suitable for hemoglobin point-of-care testing.</p>","PeriodicalId":93870,"journal":{"name":"Annales de biologie clinique","volume":"82 1","pages":"103-111"},"PeriodicalIF":0.0000,"publicationDate":"2024-04-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annales de biologie clinique","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1684/abc.2024.1868","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
The use of portable hemoglobin measuring devices is widespread. In this context, the company HemoCue® has put on the market a new device, the Hb801. It uses a whole blood absorbance measurement method and not the azidmethemoglobin measurement method used by HemoCue's older devices. We evaluated this new equipment on EDTA venous blood. Hb801 is lightweight, compact, requires a volume of 10 μL of blood and renders its result in less than a second. The repeatability and intermediate precision are close to the values expected according to Ricos, with coefficients of variation respectively for a low level of hemoglobin: 2.1% and 1.9%, for an average level: 0.8% and 1.5% and for a high level: 1.5% and 1.3%. Comparison to our laboratory reference method (XN-10 Sysmex®) and HemoCue® Hb201+ was performed on 96 samples. Bias (SD) found were: XN-10: +0.42 g/dL (0.17), HemoCue® Hb201+: +0.17 g/dL (0.41). Clinically acceptable performance (within ± 1 g/dL of reference hemoglobin) was high: 93.8%. In the end, this device seems to us to be suitable for hemoglobin point-of-care testing.