Effect of high irradiance and short exposure times on the depth of cure of six bulk-fill resin composites

IF 1.8 4区 医学 Q2 DENTISTRY, ORAL SURGERY & MEDICINE
Daniela Bandeira dos Santos, Beatriz de Cássia Romano, Vanessa Gallego Arias Pecorari, Richard Bengt Price, Marcelo Giannini
{"title":"Effect of high irradiance and short exposure times on the depth of cure of six bulk-fill resin composites","authors":"Daniela Bandeira dos Santos,&nbsp;Beatriz de Cássia Romano,&nbsp;Vanessa Gallego Arias Pecorari,&nbsp;Richard Bengt Price,&nbsp;Marcelo Giannini","doi":"10.1111/eos.12990","DOIUrl":null,"url":null,"abstract":"<p>This study examined the effect of high irradiance and short exposure times on the depth of cure of six resin-based composites (RBCs). Bluephase PowerCure and the Valo X light-curing units (LCUs) were used to photocure bulk-fill RBCs for their recommended exposure times: Admira Fusion x-tra (AFX/20s), Aura Bulk Fill (ABF/20s), Filtek One Bulk Fill (FOB/20s), Opus Bulk Fill APS (OBF/30s), Tetric EvoCeram Bulk Fill (TEC/10s) and Tetric PowerFill (TPF/10s). In addition, all bulk-fill RBCs were tested for depth of cure with one short 3 s exposure time from the Bluephase PowerCure or the Valo X in the Xtra Power mode. The RBCs (<i>n</i> = 10 per RBC) were inserted into a 4 mm diameter metal mold and covered by a polyester strip before being photocured. After 24 h of storage, uncured RBC was scraped away to determine the depth of cure of the RBCs. None of the RBCs achieved a 4 mm depth of cure. The depth of cure of TEC and TPF was unaffected by the exposure times (recommended or short) when using the Valo X. The depth of cure of AFX/20s, AFX/Xtra Power, ABF/Xtra Power, FOB/Xtra Power, and OBF/30s RBCs was greater when using Valo X compared to the Bluephase PowerCure. It was concluded that short exposure times can reduce depth of cure and should only be used for some RBCs.</p>","PeriodicalId":11983,"journal":{"name":"European Journal of Oral Sciences","volume":"132 4","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2024-05-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Oral Sciences","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/eos.12990","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

This study examined the effect of high irradiance and short exposure times on the depth of cure of six resin-based composites (RBCs). Bluephase PowerCure and the Valo X light-curing units (LCUs) were used to photocure bulk-fill RBCs for their recommended exposure times: Admira Fusion x-tra (AFX/20s), Aura Bulk Fill (ABF/20s), Filtek One Bulk Fill (FOB/20s), Opus Bulk Fill APS (OBF/30s), Tetric EvoCeram Bulk Fill (TEC/10s) and Tetric PowerFill (TPF/10s). In addition, all bulk-fill RBCs were tested for depth of cure with one short 3 s exposure time from the Bluephase PowerCure or the Valo X in the Xtra Power mode. The RBCs (n = 10 per RBC) were inserted into a 4 mm diameter metal mold and covered by a polyester strip before being photocured. After 24 h of storage, uncured RBC was scraped away to determine the depth of cure of the RBCs. None of the RBCs achieved a 4 mm depth of cure. The depth of cure of TEC and TPF was unaffected by the exposure times (recommended or short) when using the Valo X. The depth of cure of AFX/20s, AFX/Xtra Power, ABF/Xtra Power, FOB/Xtra Power, and OBF/30s RBCs was greater when using Valo X compared to the Bluephase PowerCure. It was concluded that short exposure times can reduce depth of cure and should only be used for some RBCs.

高辐照度和短曝光时间对六种散装填充树脂复合材料固化深度的影响
本研究考察了高辐照度和短曝光时间对六种树脂基复合材料(RBC)固化深度的影响。研究人员使用 Bluephase PowerCure 和 Valo X 光固化装置 (LCU) 在其建议的曝光时间内对批量填充 RBC 进行光固化:Admira Fusion x-tra (AFX/20s)、Aura Bulk Fill (ABF/20s)、Filtek One Bulk Fill (FOB/20s)、Opus Bulk Fill APS (OBF/30s)、Tetric EvoCeram Bulk Fill (TEC/10s) 和 Tetric PowerFill (TPF/10s)。此外,在 Bluephase PowerCure 或 Valo X 的 Xtra Power 模式下,用 3 秒钟的短时间曝光对所有散装填充 RBC 进行了固化深度测试。光固化前,将 RBC(每个 RBC 10 个)放入直径为 4 毫米的金属模具中,并用聚酯条覆盖。存放 24 小时后,刮去未固化的 RBC,以确定 RBC 的固化深度。没有一种 RBC 的固化深度达到 4 毫米。与 Bluephase PowerCure 相比,使用 Valo X 时 AFX/20s、AFX/Xtra Power、ABF/Xtra Power、FOB/Xtra Power 和 OBF/30s RBC 的固化深度更大。结论是,短曝光时间会降低固化深度,因此只能用于某些 RBC。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
European Journal of Oral Sciences
European Journal of Oral Sciences 医学-牙科与口腔外科
CiteScore
3.50
自引率
5.30%
发文量
61
审稿时长
2 months
期刊介绍: The European Journal of Oral Sciences is an international journal which publishes original research papers within clinical dentistry, on all basic science aspects of structure, chemistry, developmental biology, physiology and pathology of relevant tissues, as well as on microbiology, biomaterials and the behavioural sciences as they relate to dentistry. In general, analytical studies are preferred to descriptive ones. Reviews, Short Communications and Letters to the Editor will also be considered for publication. The journal is published bimonthly.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信