A High-Resolution Portable Gamma-Camera for Estimation of Absorbed Dose in Molecular Radiotherapy

IF 4.6 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
T. Bossis;M.-A. Verdier;C. Trigila;L. Pinot;F. Bouvet;A. Blot;H. Ramarijaona;T. Beaumont;D. Broggio;O. Caselles;S. Zerdoud;L. Ménard
{"title":"A High-Resolution Portable Gamma-Camera for Estimation of Absorbed Dose in Molecular Radiotherapy","authors":"T. Bossis;M.-A. Verdier;C. Trigila;L. Pinot;F. Bouvet;A. Blot;H. Ramarijaona;T. Beaumont;D. Broggio;O. Caselles;S. Zerdoud;L. Ménard","doi":"10.1109/TRPMS.2024.3376826","DOIUrl":null,"url":null,"abstract":"Molecular radiotherapy is a treatment modality that requires personalized dosimetry for efficient treatment and reduced toxicity. Current clinical imaging systems and miniaturized gamma-cameras lack the necessary features for this task. In this article, we present the design and optimization of a mobile gamma-camera with a \n<inline-formula> <tex-math>$10\\times 10$ </tex-math></inline-formula>\n cm2 field of view tailored for quantitative imaging during \n<inline-formula> <tex-math>$^{131}\\text{I}$ </tex-math></inline-formula>\n therapy of thyroid diseases. The camera uses a monolithic \n<inline-formula> <tex-math>$10\\times 10\\times 1$ </tex-math></inline-formula>\n cm3 CeBr3 scintillator coupled to a \n<inline-formula> <tex-math>$16\\times 16$ </tex-math></inline-formula>\n SiPMs array and commercial electronics. It exhibits high imaging performance with an intrinsic spatial resolution (SR) of 1.15-mm FWHM, an energy resolution of 8% FWHM at 356 keV and negligible deadtime up to 150 kcps. Images are reconstructed in real time using a convolutional neural network. The manufacturing method of tungsten collimators and shielding was optimized using laser 3-D printing to achieve an effective density of 97% that of bulk tungsten. Their geometry was adjusted with Monte-Carlo simulations in order to reduce septal penetration and scattering and optimize the signal-to-noise ratio at short times after treatment administration. Two high-energy parallel-hole collimators with high sensitivity or very high SR were designed for treatment planning and post-treatment control. The fully operational gamma-camera will soon be clinically assessed.","PeriodicalId":46807,"journal":{"name":"IEEE Transactions on Radiation and Plasma Medical Sciences","volume":"8 5","pages":"556-570"},"PeriodicalIF":4.6000,"publicationDate":"2024-03-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=10472319","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Radiation and Plasma Medical Sciences","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/10472319/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Molecular radiotherapy is a treatment modality that requires personalized dosimetry for efficient treatment and reduced toxicity. Current clinical imaging systems and miniaturized gamma-cameras lack the necessary features for this task. In this article, we present the design and optimization of a mobile gamma-camera with a $10\times 10$ cm2 field of view tailored for quantitative imaging during $^{131}\text{I}$ therapy of thyroid diseases. The camera uses a monolithic $10\times 10\times 1$ cm3 CeBr3 scintillator coupled to a $16\times 16$ SiPMs array and commercial electronics. It exhibits high imaging performance with an intrinsic spatial resolution (SR) of 1.15-mm FWHM, an energy resolution of 8% FWHM at 356 keV and negligible deadtime up to 150 kcps. Images are reconstructed in real time using a convolutional neural network. The manufacturing method of tungsten collimators and shielding was optimized using laser 3-D printing to achieve an effective density of 97% that of bulk tungsten. Their geometry was adjusted with Monte-Carlo simulations in order to reduce septal penetration and scattering and optimize the signal-to-noise ratio at short times after treatment administration. Two high-energy parallel-hole collimators with high sensitivity or very high SR were designed for treatment planning and post-treatment control. The fully operational gamma-camera will soon be clinically assessed.
用于估算分子放射治疗吸收剂量的高分辨率便携式伽马相机
分子放射治疗是一种需要个性化剂量测定的治疗方式,以实现高效治疗和减少毒性。目前的临床成像系统和微型伽马相机缺乏完成这项任务的必要功能。在这篇文章中,我们介绍了一种移动式伽马相机的设计和优化,它的视场为10美元/10平方厘米,专门用于甲状腺疾病的^{131}\text{I}$治疗过程中的定量成像。该相机使用了一个10美元/次 10美元/次 1立方厘米的单片CeBr3闪烁体,该闪烁体与一个16美元/次 16美元的SiPMs阵列和商用电子设备相连。它具有很高的成像性能,固有空间分辨率(SR)为 1.15 mm FWHM,在 356 keV 时的能量分辨率为 8% FWHM,死区时间可忽略不计,最高可达 150 kcps。使用卷积神经网络实时重建图像。钨准直器和屏蔽的制造方法通过激光三维打印进行了优化,使其有效密度达到块状钨的 97%。通过蒙特卡洛模拟调整了它们的几何形状,以减少隔膜穿透和散射,优化治疗后短时间内的信噪比。设计了两个高能量平行孔准直器,具有高灵敏度或极高的 SR,用于治疗计划和治疗后控制。全面运行的伽马相机不久将进行临床评估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
IEEE Transactions on Radiation and Plasma Medical Sciences
IEEE Transactions on Radiation and Plasma Medical Sciences RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
8.00
自引率
18.20%
发文量
109
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信