Dynamics of manual impaction instruments during total hip arthroplasty

IF 4.7 2区 医学 Q2 CELL & TISSUE ENGINEERING
Adam Reynolds, R. Doyle, Oliver R. Boughton, Justin P Cobb, S. Muirhead-Allwood, Jonathan Jeffers
{"title":"Dynamics of manual impaction instruments during total hip arthroplasty","authors":"Adam Reynolds, R. Doyle, Oliver R. Boughton, Justin P Cobb, S. Muirhead-Allwood, Jonathan Jeffers","doi":"10.1302/2046-3758.134.BJR-2023-0224.R1","DOIUrl":null,"url":null,"abstract":"Aims Manual impaction, with a mallet and introducer, remains the standard method of installing cementless acetabular cups during total hip arthroplasty (THA). This study aims to quantify the accuracy and precision of manual impaction strikes during the seating of an acetabular component. This understanding aims to help improve impaction surgical techniques and inform the development of future technologies. Methods Posterior approach THAs were carried out on three cadavers by an expert orthopaedic surgeon. An instrumented mallet and introducer were used to insert cementless acetabular cups. The motion of the mallet, relative to the introducer, was analyzed for a total of 110 strikes split into low-, medium-, and high-effort strikes. Three parameters were extracted from these data: strike vector, strike offset, and mallet face alignment. Results The force vector of the mallet strike, relative to the introducer axis, was misaligned by an average of 18.1°, resulting in an average wasted strike energy of 6.1%. Furthermore, the mean strike offset was 19.8 mm from the centre of the introducer axis and the mallet face, relative to the introducer strike face, was misaligned by a mean angle of 15.2° from the introducer strike face. Conclusion The direction of the impact vector in manual impaction lacks both accuracy and precision. There is an opportunity to improve this through more advanced impaction instruments or surgical training. Cite this article: Bone Joint Res 2024;13(4):193–200.","PeriodicalId":9074,"journal":{"name":"Bone & Joint Research","volume":null,"pages":null},"PeriodicalIF":4.7000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bone & Joint Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1302/2046-3758.134.BJR-2023-0224.R1","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CELL & TISSUE ENGINEERING","Score":null,"Total":0}
引用次数: 0

Abstract

Aims Manual impaction, with a mallet and introducer, remains the standard method of installing cementless acetabular cups during total hip arthroplasty (THA). This study aims to quantify the accuracy and precision of manual impaction strikes during the seating of an acetabular component. This understanding aims to help improve impaction surgical techniques and inform the development of future technologies. Methods Posterior approach THAs were carried out on three cadavers by an expert orthopaedic surgeon. An instrumented mallet and introducer were used to insert cementless acetabular cups. The motion of the mallet, relative to the introducer, was analyzed for a total of 110 strikes split into low-, medium-, and high-effort strikes. Three parameters were extracted from these data: strike vector, strike offset, and mallet face alignment. Results The force vector of the mallet strike, relative to the introducer axis, was misaligned by an average of 18.1°, resulting in an average wasted strike energy of 6.1%. Furthermore, the mean strike offset was 19.8 mm from the centre of the introducer axis and the mallet face, relative to the introducer strike face, was misaligned by a mean angle of 15.2° from the introducer strike face. Conclusion The direction of the impact vector in manual impaction lacks both accuracy and precision. There is an opportunity to improve this through more advanced impaction instruments or surgical training. Cite this article: Bone Joint Res 2024;13(4):193–200.
全髋关节置换术中手动撞击器械的动态变化
目的 在全髋关节置换术(THA)中,使用锤子和导入器手动撞击仍是安装无骨水泥髋臼杯的标准方法。本研究旨在量化髋臼组件就位过程中手动撞击的准确性和精确度。了解这些信息有助于改进撞击手术技术,并为未来技术的开发提供依据。方法 由一名骨科专家在三具尸体上进行后方入路 THAs。使用带器械的臼槌和导引器插入无骨水泥髋臼杯。对臼槌相对于导引器的运动进行了分析,共进行了 110 次撞击,分为低、中、高三种强度的撞击。从这些数据中提取了三个参数:撞击矢量、撞击偏移和臼槌面对齐。结果 相对于导引轴,击槌的力矢量平均错位 18.1°,导致击槌能量平均浪费 6.1%。此外,平均撞击偏移距导引器轴线中心 19.8 毫米,相对于导引器撞击面,槌面与导引器撞击面平均错位 15.2°。结论 人工撞击的撞击矢量方向缺乏准确性和精确性。有机会通过更先进的冲击器械或手术培训来改善这一问题。引用本文:Bone Joint Res 2024;13(4):193-200.
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Bone & Joint Research
Bone & Joint Research CELL & TISSUE ENGINEERING-ORTHOPEDICS
CiteScore
7.40
自引率
23.90%
发文量
156
审稿时长
12 weeks
期刊介绍: The gold open access journal for the musculoskeletal sciences. Included in PubMed and available in PubMed Central.
文献相关原料
公司名称 产品信息 采购帮参考价格
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信