Validation of the RBP-9801 oscillometric blood pressure monitor in the general population according to the association for the Advancement of Medical Instrumentation/European Society of Hypertension/ International Organization for Standardization Universal Standard

IF 2.7 3区 医学 Q2 PERIPHERAL VASCULAR DISEASE
Shijie Yang MM, Zhanyang Zhou MM, Huanhuan Miao MM, Qiong Zhou MD, PhD, Mei Zhai MD, PhD, Yuqing Zhang MD, PhD
{"title":"Validation of the RBP-9801 oscillometric blood pressure monitor in the general population according to the association for the Advancement of Medical Instrumentation/European Society of Hypertension/ International Organization for Standardization Universal Standard","authors":"Shijie Yang MM,&nbsp;Zhanyang Zhou MM,&nbsp;Huanhuan Miao MM,&nbsp;Qiong Zhou MD, PhD,&nbsp;Mei Zhai MD, PhD,&nbsp;Yuqing Zhang MD, PhD","doi":"10.1111/jch.14812","DOIUrl":null,"url":null,"abstract":"<p>The aim of this study was to evaluate the accuracy of the single upper-arm cuff oscillometric blood pressure (BP) monitor RBP-9801 developed for office and home BP measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060–2:2018). Subjects were recruited to fulfil the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method. A total of 105 subjects were recruited and 85 were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 2.3 ± 6.4/3.1 ± 5.8 mmHg (systolic/diastolic). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 5.24/5.03 mmHg (systolic/diastolic). The conclusion is that the RBP-9801 oscillometric device for office and home BP measurement fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060–2:2018) in the general population and can be recommended for clinic and self-use at home.</p>","PeriodicalId":50237,"journal":{"name":"Journal of Clinical Hypertension","volume":null,"pages":null},"PeriodicalIF":2.7000,"publicationDate":"2024-04-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/jch.14812","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Hypertension","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/jch.14812","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

The aim of this study was to evaluate the accuracy of the single upper-arm cuff oscillometric blood pressure (BP) monitor RBP-9801 developed for office and home BP measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060–2:2018). Subjects were recruited to fulfil the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method. A total of 105 subjects were recruited and 85 were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 2.3 ± 6.4/3.1 ± 5.8 mmHg (systolic/diastolic). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 5.24/5.03 mmHg (systolic/diastolic). The conclusion is that the RBP-9801 oscillometric device for office and home BP measurement fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060–2:2018) in the general population and can be recommended for clinic and self-use at home.

Abstract Image

根据美国医学仪器促进协会/欧洲高血压学会/国际标准化组织通用标准,在普通人群中验证 RBP-9801 示波血压计。
本研究的目的是根据医疗仪器促进协会/欧洲高血压学会/国际标准化组织(AAMI/ESH/ISO)通用标准(ISO 81060-2:2018),评估开发用于普通人群办公室和家庭血压测量的单上臂袖带示波血压(BP)监测仪 RBP-9801 的准确性。受试者的招募符合 AAMI/ESH/ISO 通用标准的年龄、性别、血压和袖带分布标准,在普通人群中采用同臂顺序血压测量方法。共招募了 105 名受试者,对其中 85 名进行了分析。对于验证标准 1,测试设备和参考血压读数之间差异的平均值(± SD)为 2.3 ± 6.4/3.1 ± 5.8 mmHg(收缩压/舒张压)。就标准 2 而言,每个受试者的测试设备与参考血压之间的平均血压差异的 SD 值为 5.24/5.03 mmHg(收缩压/舒张压)。结论是用于办公室和家庭血压测量的 RBP-9801 血压示波仪符合 AAMI/ESH/ISO 通用标准(ISO 81060-2:2018)对普通人群的所有要求,可推荐用于诊所和家庭自用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Clinical Hypertension
Journal of Clinical Hypertension PERIPHERAL VASCULAR DISEASE-
CiteScore
5.80
自引率
7.10%
发文量
191
审稿时长
4-8 weeks
期刊介绍: The Journal of Clinical Hypertension is a peer-reviewed, monthly publication that serves internists, cardiologists, nephrologists, endocrinologists, hypertension specialists, primary care practitioners, pharmacists and all professionals interested in hypertension by providing objective, up-to-date information and practical recommendations on the full range of clinical aspects of hypertension. Commentaries and columns by experts in the field provide further insights into our original research articles as well as on major articles published elsewhere. Major guidelines for the management of hypertension are also an important feature of the Journal. Through its partnership with the World Hypertension League, JCH will include a new focus on hypertension and public health, including major policy issues, that features research and reviews related to disease characteristics and management at the population level.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信