{"title":"Evaluation of the Agreement Between a New Pyramid Wavefront Sensor Aberrometer and Scheiner-Smirnov Aberrometers","authors":"Yizhou Yang, MD, Rui Ning, MD, Shuoyu Xu, MD, Jinxuan Xiahou, MD, Jiawei Li, MD, Giacomo Savini, MD, PhD, Domenico Schiano-Lomoriello, MD, Xingtao Zhou, MD, PhD, Jinhai Huang, MD, PhD","doi":"10.3928/1081597x-20240311-02","DOIUrl":null,"url":null,"abstract":"<section><h3>Purpose:</h3><p>To assess agreement between a new aberrometer (Osiris-T; CSO) employing pyramid wavefront sensor technique and Scheiner-Smirnov aberrometer (OPD-Scan III; Nidek) on measuring ocular, corneal, and internal aberrations in healthy participants.</p></section><section><h3>Methods:</h3><p>The measurements were conducted three times consecutively by an experienced examiner. The total root mean square (RMS) aberrations, higher order aberration RMS, coma Z<sub>3</sub><sup>±1</sup>, trefoil Z<sub>3</sub><sup>±3</sup>, spherical aberration Z<sub>4</sub><sup>0</sup><sub>,</sub> and astigmatism II Z<sub>4</sub><sup>±2</sup> up to 7th order were exported in both 4-and 6-mm pupil zones. The parameters between the two devices were statistically compared using the paired <i>t</i>-test, and the differences assessed with Bland-Altman plots and 95% limits of agreement.</p></section><section><h3>Results:</h3><p>This prospective study included 70 right eyes of 70 healthy participants with an average age of 25.94 ± 6.59 years (range: 18 to 47 years). The mean difference in the two devices ranged from 0.01 µm for astigmatism II Z<sub>4</sub><sup>±2</sup> to 0.63 µm for total RMS in 4 mm and from 0.01 to 1.41 µm in 6-mm pupil size. The Bland-Altman analysis of ocular, corneal, and internal aberrations indicated high agreement between the two devices and the maximum absolute values for 95% limits of agreement ranged from 0.03 to 1.06 µm for 4-mm pupil diameters and 0.12 to 1.13 µm for 6-mm pupil diameters.</p></section><section><h3>Conclusions:</h3><p>The newly developed pyramid wavefront sensor technique aberrometer demonstrated a high agreement with a Scheiner-Smirnov aberrometer when measuring ocular, corneal, and internal aberrations in healthy participants. Thus, the two aberrometers may be considered interchangeable for clinical applications.</p><p><strong>[<i>J Refract Surg</i>. 2024;40(4):e218–e228.]</strong></p></section>","PeriodicalId":16951,"journal":{"name":"Journal of refractive surgery","volume":"27 1","pages":""},"PeriodicalIF":2.9000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3928/1081597x-20240311-02","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose:
To assess agreement between a new aberrometer (Osiris-T; CSO) employing pyramid wavefront sensor technique and Scheiner-Smirnov aberrometer (OPD-Scan III; Nidek) on measuring ocular, corneal, and internal aberrations in healthy participants.
Methods:
The measurements were conducted three times consecutively by an experienced examiner. The total root mean square (RMS) aberrations, higher order aberration RMS, coma Z3±1, trefoil Z3±3, spherical aberration Z40, and astigmatism II Z4±2 up to 7th order were exported in both 4-and 6-mm pupil zones. The parameters between the two devices were statistically compared using the paired t-test, and the differences assessed with Bland-Altman plots and 95% limits of agreement.
Results:
This prospective study included 70 right eyes of 70 healthy participants with an average age of 25.94 ± 6.59 years (range: 18 to 47 years). The mean difference in the two devices ranged from 0.01 µm for astigmatism II Z4±2 to 0.63 µm for total RMS in 4 mm and from 0.01 to 1.41 µm in 6-mm pupil size. The Bland-Altman analysis of ocular, corneal, and internal aberrations indicated high agreement between the two devices and the maximum absolute values for 95% limits of agreement ranged from 0.03 to 1.06 µm for 4-mm pupil diameters and 0.12 to 1.13 µm for 6-mm pupil diameters.
Conclusions:
The newly developed pyramid wavefront sensor technique aberrometer demonstrated a high agreement with a Scheiner-Smirnov aberrometer when measuring ocular, corneal, and internal aberrations in healthy participants. Thus, the two aberrometers may be considered interchangeable for clinical applications.
期刊介绍:
The Journal of Refractive Surgery, the official journal of the International Society of Refractive Surgery, a partner of the American Academy of Ophthalmology, has been a monthly peer-reviewed forum for original research, review, and evaluation of refractive and lens-based surgical procedures for more than 30 years. Practical, clinically valuable articles provide readers with the most up-to-date information regarding advances in the field of refractive surgery. Begin to explore the Journal and all of its great benefits such as:
• Columns including “Translational Science,” “Surgical Techniques,” and “Biomechanics”
• Supplemental videos and materials available for many articles
• Access to current articles, as well as several years of archived content
• Articles posted online just 2 months after acceptance.