Analytical performances description of the immunoturbidimetric method for the determination of HbA1c using Selectra Pro M automated system at the Institut National dHygine (INH) of Lom

Katawa Gnatoulma, Adjete Kossi, Marthe Amessoudji Oukoe, Halatoko Wemboo, Edlom Tchadie Pélagie, Bara Fagdeba, Dossou Irène, Nguepou Tchopba Christèle, Simplice Karou Damintoti, Ameyapoh Yovi
{"title":"Analytical performances description of the immunoturbidimetric method for the determination of HbA1c using Selectra Pro M automated system at the Institut National dHygine (INH) of Lom","authors":"Katawa Gnatoulma, Adjete Kossi, Marthe Amessoudji Oukoe, Halatoko Wemboo, Edlom Tchadie Pélagie, Bara Fagdeba, Dossou Irène, Nguepou Tchopba Christèle, Simplice Karou Damintoti, Ameyapoh Yovi","doi":"10.5897/ajb2023.17604","DOIUrl":null,"url":null,"abstract":"HbA1c is the best indicator for monitoring glycemic control in both type 1 and type 2 diabetic patients. Its value allows us to predict the development of long-term degenerative complications of diabetes. Several techniques offered by different laboratories allow the determination of HbA1c based on different principles. The purpose of the present study was to perform the verification of the analytical performances of the Selectra Pro M automated system, using immunoturbidimetry as the HbA1c assay method. Risk assessment of the method was performed; the analytical performances of the assay process was evaluated and the immunoturbidimetric method on the Selectra Pro M was compared to the capillary method on the Capillarys 2 Flex Piercing. Reproducibility and intermediate precision were satisfactory with CV in the range of 1.78 to 1.96% for the low level of Internal Quality Control (IQC) (HbA1c = 5.4%) and 0.82 to 2.28% for the high level of IQC (HbA1c = 11.3%). The reagent was linear from 2.5 to 16% (4 to 151 mmol/mol). The accuracy was considered satisfactory. The linear regression calculation (least squares line) showed an excellent correlation (R 2 = 0.96) between the two techniques with an equation of the type [Capillarys] = 1.13 [Selectra Pro M] - 0.83. All documented and tested performances correspond to the performance required for HbA1c assay at INH of Lomé.","PeriodicalId":7414,"journal":{"name":"African Journal of Biotechnology","volume":"480 ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-01-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"African Journal of Biotechnology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5897/ajb2023.17604","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

HbA1c is the best indicator for monitoring glycemic control in both type 1 and type 2 diabetic patients. Its value allows us to predict the development of long-term degenerative complications of diabetes. Several techniques offered by different laboratories allow the determination of HbA1c based on different principles. The purpose of the present study was to perform the verification of the analytical performances of the Selectra Pro M automated system, using immunoturbidimetry as the HbA1c assay method. Risk assessment of the method was performed; the analytical performances of the assay process was evaluated and the immunoturbidimetric method on the Selectra Pro M was compared to the capillary method on the Capillarys 2 Flex Piercing. Reproducibility and intermediate precision were satisfactory with CV in the range of 1.78 to 1.96% for the low level of Internal Quality Control (IQC) (HbA1c = 5.4%) and 0.82 to 2.28% for the high level of IQC (HbA1c = 11.3%). The reagent was linear from 2.5 to 16% (4 to 151 mmol/mol). The accuracy was considered satisfactory. The linear regression calculation (least squares line) showed an excellent correlation (R 2 = 0.96) between the two techniques with an equation of the type [Capillarys] = 1.13 [Selectra Pro M] - 0.83. All documented and tested performances correspond to the performance required for HbA1c assay at INH of Lomé.
洛姆国立血液研究所(INH)使用Selectra Pro M自动系统测定HbA1c的免疫比浊法的分析性能说明
HbA1c 是监测 1 型和 2 型糖尿病患者血糖控制情况的最佳指标。它的数值可以让我们预测糖尿病长期退行性并发症的发展。不同实验室提供的多种技术可根据不同原理测定 HbA1c。本研究的目的是使用免疫比浊法作为 HbA1c 检测方法,对 Selectra Pro M 自动化系统的分析性能进行验证。对该方法进行了风险评估;评估了检测过程的分析性能,并将 Selectra Pro M 上的免疫比浊法与 Capillarys 2 Flex Piercing 上的毛细管法进行了比较。再现性和中间精度令人满意,低水平内部质量控制(IQC)(HbA1c = 5.4%)的 CV 为 1.78%至 1.96%,高水平内部质量控制(IQC)(HbA1c = 11.3%)的 CV 为 0.82%至 2.28%。该试剂的线性范围为 2.5%至 16%(4 至 151 mmol/mol)。准确度令人满意。线性回归计算(最小二乘法线)显示,两种技术之间具有极好的相关性(R 2 = 0.96),方程类型为 [Capillarys] = 1.13 [Selectra Pro M] - 0.83。所有记录和测试的性能都符合洛美 INH 的 HbA1c 检测要求。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
African Journal of Biotechnology
African Journal of Biotechnology 工程技术-生物工程与应用微生物
自引率
0.00%
发文量
15
审稿时长
4.7 months
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信