Enes Elvin Gul, Pierre Baudinaud, Victor Waldmann, Avi Sabbag, Yousef Jubeh, Nicholas Clementy, Arnaud Bisson, Pierre Ollitrault, Sergio Conti, Adrian Carabelli, Zeki Dogan
{"title":"Leadless pacemaker implantation following tricuspid interventions: multicenter collaboration of feasibility and safety.","authors":"Enes Elvin Gul, Pierre Baudinaud, Victor Waldmann, Avi Sabbag, Yousef Jubeh, Nicholas Clementy, Arnaud Bisson, Pierre Ollitrault, Sergio Conti, Adrian Carabelli, Zeki Dogan","doi":"10.1007/s10840-024-01796-w","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Permanent pacing is often required following valve intervention (either surgical or percutaneous); however, tricuspid interventions pose specific challenges to conventional pacing. Therefore, leadless pacemaker (LP) implantation may be the preferred strategy when permanent pacing is required after tricuspid valve intervention.</p><p><strong>Purpose: </strong>To report periprocedural outcomes and follow-up of patients undergoing implantation of a LP system following tricuspid valve interventions.</p><p><strong>Methods: </strong>Patients with previous tricuspid valve intervention at the time of attempted implantation of a LP (MicraTM, Medtronic, Minneapolis, MN, USA) were included.</p><p><strong>Results: </strong>Between 2019 and 2022, 40 patients underwent LP implantations following tricuspid interventions in 5 large tertiary centers. The mean age was 68.9 ± 13.7 years, and 48% patients were male. The indication for pacing was as following: AVB in 27 (68%) patients, AF with slow ventricular response in 10 (25%) patients, and refractory rapid atrial fibrillation (AF) referred to AV junction ablation in 3 (7%) patients. Most of the patients received Micra VR (78%). The procedure was successful in all patients. The mean procedural time is 58 ± 32 min, and the median fluoroscopy time is 7.5 min. Electrical parameters were within normal range (threshold: 1.35 ± 1.2 V@0.24 ms, impedance: 772 ± 245 Ohm, R-wave: 6.9 ± 5.4 mV). No acute complications were observed. During a mean follow-up of 10 months, electrical parameters remained stable, and 4 deaths were occurred (not related to the procedure).</p><p><strong>Conclusion: </strong>A LP is a safe and efficient option following tricuspid valve interventions.</p>","PeriodicalId":16202,"journal":{"name":"Journal of Interventional Cardiac Electrophysiology","volume":null,"pages":null},"PeriodicalIF":2.1000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Interventional Cardiac Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10840-024-01796-w","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/4/1 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Permanent pacing is often required following valve intervention (either surgical or percutaneous); however, tricuspid interventions pose specific challenges to conventional pacing. Therefore, leadless pacemaker (LP) implantation may be the preferred strategy when permanent pacing is required after tricuspid valve intervention.
Purpose: To report periprocedural outcomes and follow-up of patients undergoing implantation of a LP system following tricuspid valve interventions.
Methods: Patients with previous tricuspid valve intervention at the time of attempted implantation of a LP (MicraTM, Medtronic, Minneapolis, MN, USA) were included.
Results: Between 2019 and 2022, 40 patients underwent LP implantations following tricuspid interventions in 5 large tertiary centers. The mean age was 68.9 ± 13.7 years, and 48% patients were male. The indication for pacing was as following: AVB in 27 (68%) patients, AF with slow ventricular response in 10 (25%) patients, and refractory rapid atrial fibrillation (AF) referred to AV junction ablation in 3 (7%) patients. Most of the patients received Micra VR (78%). The procedure was successful in all patients. The mean procedural time is 58 ± 32 min, and the median fluoroscopy time is 7.5 min. Electrical parameters were within normal range (threshold: 1.35 ± 1.2 V@0.24 ms, impedance: 772 ± 245 Ohm, R-wave: 6.9 ± 5.4 mV). No acute complications were observed. During a mean follow-up of 10 months, electrical parameters remained stable, and 4 deaths were occurred (not related to the procedure).
Conclusion: A LP is a safe and efficient option following tricuspid valve interventions.
期刊介绍:
The Journal of Interventional Cardiac Electrophysiology is an international publication devoted to fostering research in and development of interventional techniques and therapies for the management of cardiac arrhythmias. It is designed primarily to present original research studies and scholarly scientific reviews of basic and applied science and clinical research in this field. The Journal will adopt a multidisciplinary approach to link physical, experimental, and clinical sciences as applied to the development of and practice in interventional electrophysiology. The Journal will examine techniques ranging from molecular, chemical and pharmacologic therapies to device and ablation technology. Accordingly, original research in clinical, epidemiologic and basic science arenas will be considered for publication. Applied engineering or physical science studies pertaining to interventional electrophysiology will be encouraged. The Journal is committed to providing comprehensive and detailed treatment of major interventional therapies and innovative techniques in a structured and clinically relevant manner. It is directed at clinical practitioners and investigators in the rapidly growing field of interventional electrophysiology. The editorial staff and board reflect this bias and include noted international experts in this area with a wealth of expertise in basic and clinical investigation. Peer review of all submissions, conflict of interest guidelines and periodic editorial board review of all Journal policies have been established.