Francisco Pastor-Pascual, Paz Orts-Vila, Pedro Tañá-Sanz, Santiago Tañá-Sanz, Ramón Ruiz-Mesa, Pedro Tañá-Rivero
{"title":"Non-diffractive, toric, extended depth-of-focus intraocular lenses in eyes with low corneal astigmatism.","authors":"Francisco Pastor-Pascual, Paz Orts-Vila, Pedro Tañá-Sanz, Santiago Tañá-Sanz, Ramón Ruiz-Mesa, Pedro Tañá-Rivero","doi":"10.1186/s40662-024-00380-7","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To assess clinical outcomes after implanting toric, extended-depth-of-focus intraocular lenses (IOLs) to correct low corneal astigmatism in eyes with cataracts.</p><p><strong>Methods: </strong>47 eyes were implanted with the AcrySof IQ Vivity Toric DFT215 IOL. Main outcome measures were refractive error, monocular uncorrected and corrected distance (UDVA/CDVA), uncorrected and distance-corrected intermediate (UIVA/DCIVA), and uncorrected near and distance-corrected near (UNVA/DCNVA) visual acuities, monocular defocus curve, rotational stability, and IOLSAT and QUVID questionnaires. Patients were assessed at 3 months postsurgery.</p><p><strong>Results: </strong>All eyes had a postoperative spherical equivalent (SE) within ± 0.50 D and 97.87% (n = 46) had a refractive cylinder ≤ 0.50 D. The mean SE and refractive cylinder were - 0.10 ± 0.17 D and - 0.16 ± 0.24 D, respectively. The CDVA was ≥ 20/25 and ≥ 20/32 in 95.74% (n = 45) and 97.87% (n = 46) of eyes, respectively. The DCIVA was ≥ 20/32 in 85.11% (n = 40) of eyes and the DCNVA was ≥ 20/40 in 74.47% (n = 35). The mean values of CDVA, DCIVA, and DCNVA were - 0.02 ± 0.08, 0.14 ± 0.09, and 0.23 ± 0.12 logMAR, respectively. The defocus curve revealed good visual acuity at far and intermediate distances with a depth-of-focus of about 1.75 D. IOL rotation was 0.74 ± 1.13 degrees and all eyes had a rotation of less than 5 degrees. Patients reported either good or very good postoperative vision without eyeglasses under bright-light-conditions at distance (87.80%, 36/41) and intermediate distance (92.68%, 38/41). Between about 63.83%-72.34% (30-34) of patients reported no starburst, halos, or glare, or if experienced, were not bothersome.</p><p><strong>Conclusions: </strong>The Vivity toric IOL implanted in eyes with low-astigmatism provides accurate refractive outcomes, good visual acuity at different distances and excellent rotational stability. Trial Registration The study was registered with the German Clinical Trials Register (DRKS00030579).</p>","PeriodicalId":12194,"journal":{"name":"Eye and Vision","volume":"11 1","pages":"14"},"PeriodicalIF":4.1000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10983649/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Eye and Vision","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40662-024-00380-7","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: To assess clinical outcomes after implanting toric, extended-depth-of-focus intraocular lenses (IOLs) to correct low corneal astigmatism in eyes with cataracts.
Methods: 47 eyes were implanted with the AcrySof IQ Vivity Toric DFT215 IOL. Main outcome measures were refractive error, monocular uncorrected and corrected distance (UDVA/CDVA), uncorrected and distance-corrected intermediate (UIVA/DCIVA), and uncorrected near and distance-corrected near (UNVA/DCNVA) visual acuities, monocular defocus curve, rotational stability, and IOLSAT and QUVID questionnaires. Patients were assessed at 3 months postsurgery.
Results: All eyes had a postoperative spherical equivalent (SE) within ± 0.50 D and 97.87% (n = 46) had a refractive cylinder ≤ 0.50 D. The mean SE and refractive cylinder were - 0.10 ± 0.17 D and - 0.16 ± 0.24 D, respectively. The CDVA was ≥ 20/25 and ≥ 20/32 in 95.74% (n = 45) and 97.87% (n = 46) of eyes, respectively. The DCIVA was ≥ 20/32 in 85.11% (n = 40) of eyes and the DCNVA was ≥ 20/40 in 74.47% (n = 35). The mean values of CDVA, DCIVA, and DCNVA were - 0.02 ± 0.08, 0.14 ± 0.09, and 0.23 ± 0.12 logMAR, respectively. The defocus curve revealed good visual acuity at far and intermediate distances with a depth-of-focus of about 1.75 D. IOL rotation was 0.74 ± 1.13 degrees and all eyes had a rotation of less than 5 degrees. Patients reported either good or very good postoperative vision without eyeglasses under bright-light-conditions at distance (87.80%, 36/41) and intermediate distance (92.68%, 38/41). Between about 63.83%-72.34% (30-34) of patients reported no starburst, halos, or glare, or if experienced, were not bothersome.
Conclusions: The Vivity toric IOL implanted in eyes with low-astigmatism provides accurate refractive outcomes, good visual acuity at different distances and excellent rotational stability. Trial Registration The study was registered with the German Clinical Trials Register (DRKS00030579).
期刊介绍:
Eye and Vision is an open access, peer-reviewed journal for ophthalmologists and visual science specialists. It welcomes research articles, reviews, methodologies, commentaries, case reports, perspectives and short reports encompassing all aspects of eye and vision. Topics of interest include but are not limited to: current developments of theoretical, experimental and clinical investigations in ophthalmology, optometry and vision science which focus on novel and high-impact findings on central issues pertaining to biology, pathophysiology and etiology of eye diseases as well as advances in diagnostic techniques, surgical treatment, instrument updates, the latest drug findings, results of clinical trials and research findings. It aims to provide ophthalmologists and visual science specialists with the latest developments in theoretical, experimental and clinical investigations in eye and vision.