Addressing challenges in diagnostic X-ray dosimetry: uncertainties and corrections for Al2O3:C-based optically stimulated luminescent dosimeters.

IF 4.6 Q2 MATERIALS SCIENCE, BIOMATERIALS
ACS Applied Bio Materials Pub Date : 2024-09-01 Epub Date: 2024-03-25 DOI:10.1007/s13246-024-01407-y
Jeannie Hsiu Ding Wong, Wan Hazlinda Ismail
{"title":"Addressing challenges in diagnostic X-ray dosimetry: uncertainties and corrections for Al<sub>2</sub>O<sub>3</sub>:C-based optically stimulated luminescent dosimeters.","authors":"Jeannie Hsiu Ding Wong, Wan Hazlinda Ismail","doi":"10.1007/s13246-024-01407-y","DOIUrl":null,"url":null,"abstract":"<p><p>The use of Al<sub>2</sub>O<sub>3</sub>:C-based optically stimulated luminescent dosimeters (OSLDs) in diagnostic X-ray is a challenge because of their energy dependence (ED) and variability of element sensitivity factors (ESFs). This study aims to develop a method to determine ED and ESFs of Landauer nanoDot™ OSLDs for clinical X-ray and investigate the uncertainties associated with ESF and ED correction factors. An area of 2 × 2 cm<sup>2</sup> at the central axis of the X-ray field was used to establish the ESFs. A total of 80 OSLDs were categorized into \"controlled\" (n = 40) and \"less-controlled\" groups (n = 40). The ESFs of the OSLDs were determined using an 80 kVp X-ray beam quality in free-air geometry. The OSLDs were cross-calibrated with an ion chamber to establish the average calibration coefficient and ESFs. The OSLDs were then irradiated at tube potentials ranging from 50 to 150 kVp to determine their ED. The uniformity of the X-ray field was ± 1.5% at 100 cm source-to-surface distance. The batch homogeneities of user-defined ESFs were 2.4% and 8.7% for controlled and less-controlled OSLDs, respectively. The ED of OSLDs ranged from 1.125 to 0.812 as tube potential increased from 50 kVp to 150 kVp. The total uncertainty of OSLDs, without ED correction, could be as high as 16%. After applying ESF and ED correction, the total uncertainties were reduced to 6.3% in controlled OLSDs and 11.6% in less-controlled ones. OSLDs corrected with user-defined ESF and ED can reduce the uncertainty of dose measurements in diagnostic X-rays, particularly in managing less-controlled OSLDs.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s13246-024-01407-y","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/25 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0

Abstract

The use of Al2O3:C-based optically stimulated luminescent dosimeters (OSLDs) in diagnostic X-ray is a challenge because of their energy dependence (ED) and variability of element sensitivity factors (ESFs). This study aims to develop a method to determine ED and ESFs of Landauer nanoDot™ OSLDs for clinical X-ray and investigate the uncertainties associated with ESF and ED correction factors. An area of 2 × 2 cm2 at the central axis of the X-ray field was used to establish the ESFs. A total of 80 OSLDs were categorized into "controlled" (n = 40) and "less-controlled" groups (n = 40). The ESFs of the OSLDs were determined using an 80 kVp X-ray beam quality in free-air geometry. The OSLDs were cross-calibrated with an ion chamber to establish the average calibration coefficient and ESFs. The OSLDs were then irradiated at tube potentials ranging from 50 to 150 kVp to determine their ED. The uniformity of the X-ray field was ± 1.5% at 100 cm source-to-surface distance. The batch homogeneities of user-defined ESFs were 2.4% and 8.7% for controlled and less-controlled OSLDs, respectively. The ED of OSLDs ranged from 1.125 to 0.812 as tube potential increased from 50 kVp to 150 kVp. The total uncertainty of OSLDs, without ED correction, could be as high as 16%. After applying ESF and ED correction, the total uncertainties were reduced to 6.3% in controlled OLSDs and 11.6% in less-controlled ones. OSLDs corrected with user-defined ESF and ED can reduce the uncertainty of dose measurements in diagnostic X-rays, particularly in managing less-controlled OSLDs.

Abstract Image

应对诊断 X 射线剂量测定方面的挑战:基于 Al2O3:C 的光激发发光剂量计的不确定性和修正。
由于 Al2O3:C 基光激发发光剂量计(OSLD)的能量依赖性(ED)和元素灵敏度系数(ESF)的可变性,将其用于诊断 X 射线是一项挑战。本研究旨在开发一种方法来确定用于临床 X 射线的 Landauer nanoDot™ OSLD 的 ED 和 ESF,并研究与 ESF 和 ED 校正因子相关的不确定性。X 射线场中心轴上 2 × 2 cm2 的区域用于确定 ESF。共有80例OSLD被分为 "控制 "组(40例)和 "非控制 "组(40例)。OSLD 的 ESF 是在自由空气几何条件下使用 80 kVp 的 X 射线束质量测定的。用离子室对 OSLD 进行交叉校准,以确定平均校准系数和 ESF。然后在 50 至 150 kVp 的管电位下对 OSLD 进行辐照,以确定其 ED。在 100 厘米的源到表面距离上,X 射线场的均匀性为 ± 1.5%。用户定义的 ESF 的批次均匀度分别为 2.4% 和 8.7%。当管电位从 50 kVp 上升到 150 kVp 时,OSLD 的 ED 值从 1.125 到 0.812 不等。在没有进行 ED 校正的情况下,OSLD 的总不确定性可高达 16%。应用 ESF 和 ED 校正后,受控 OLSD 的总不确定性降低到 6.3%,而受控程度较低的 OLSD 的总不确定性降低到 11.6%。用用户定义的 ESF 和 ED 修正 OSLD 可以减少诊断 X 射线剂量测量的不确定性,尤其是在管理控制较差的 OSLD 时。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
ACS Applied Bio Materials
ACS Applied Bio Materials Chemistry-Chemistry (all)
CiteScore
9.40
自引率
2.10%
发文量
464
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信