An Integrated Approach to Control the Penetration Depth of 3D-Printed Hollow Microneedles

IF 1.8
Kendall Marie Defelippi, Allyson Yuuka Saumei Kwong, Julia Rose Appleget, Rana Altay, Maya Bree Matheny, Mary Margaret Dubus, Lily Marie Eribes, M. Mobed-Miremadi
{"title":"An Integrated Approach to Control the Penetration Depth of 3D-Printed Hollow Microneedles","authors":"Kendall Marie Defelippi, Allyson Yuuka Saumei Kwong, Julia Rose Appleget, Rana Altay, Maya Bree Matheny, Mary Margaret Dubus, Lily Marie Eribes, M. Mobed-Miremadi","doi":"10.3390/applmech5020015","DOIUrl":null,"url":null,"abstract":"A variety of hollow microneedle (HMN) designs has emerged for minimally invasive therapies and monitoring systems. In this study, a design change limiting the indentation depth of the (3D) printed custom microneedle assembly (circular array of five conical frusta with and without a stopper, aspect ratio = 1.875) fabricated using stereolithography has been experimentally validated and modeled in silico. The micro-indentation profiles generated in confined compression on 1 mm ± 0.073 mm alginate films enabled the generation of a Prony series, where displacement ranged from 100 to 250 µm. These constants were used as intrinsic properties simulating experimental ramp/release profiles. Puncture occurred on two distinct hydrogel formulations at the design depth of 150 µm and indentation rate of 0.1 mm/s characterized by a peak force of 3.5 N (H = 31 kPa) and 8.3 N (H = 36.5 kPa), respectively. Experimental and theoretical alignments for peak force trends were obtained when the printing resolution was simulated. Higher puncture force and uniformity inferred by the stopper was confirmed via microscopy and profilometry. Meanwhile, poroviscoelasticity characterization is required to distinguish mass loss vs. redistribution post-indentation through pycnometry. Results from this paper highlight the feasibility of insertion-depth control within the epidermis thickness for the first time in solid HMN literature.","PeriodicalId":503399,"journal":{"name":"Applied Mechanics","volume":" 1","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2024-03-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Applied Mechanics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/applmech5020015","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

A variety of hollow microneedle (HMN) designs has emerged for minimally invasive therapies and monitoring systems. In this study, a design change limiting the indentation depth of the (3D) printed custom microneedle assembly (circular array of five conical frusta with and without a stopper, aspect ratio = 1.875) fabricated using stereolithography has been experimentally validated and modeled in silico. The micro-indentation profiles generated in confined compression on 1 mm ± 0.073 mm alginate films enabled the generation of a Prony series, where displacement ranged from 100 to 250 µm. These constants were used as intrinsic properties simulating experimental ramp/release profiles. Puncture occurred on two distinct hydrogel formulations at the design depth of 150 µm and indentation rate of 0.1 mm/s characterized by a peak force of 3.5 N (H = 31 kPa) and 8.3 N (H = 36.5 kPa), respectively. Experimental and theoretical alignments for peak force trends were obtained when the printing resolution was simulated. Higher puncture force and uniformity inferred by the stopper was confirmed via microscopy and profilometry. Meanwhile, poroviscoelasticity characterization is required to distinguish mass loss vs. redistribution post-indentation through pycnometry. Results from this paper highlight the feasibility of insertion-depth control within the epidermis thickness for the first time in solid HMN literature.
控制三维打印空心微针穿透深度的综合方法
微创疗法和监测系统中出现了多种空心微针(HMN)设计。在这项研究中,对使用立体光刻技术制造的(三维)打印定制微针组件(由五个带或不带塞子的圆锥形栅栏组成的圆形阵列,长宽比 = 1.875)的压痕深度进行限制的设计变更已在实验中得到验证,并在硅学中进行了建模。在 1 毫米 ± 0.073 毫米的藻酸盐薄膜上,通过限制压缩产生的微压痕轮廓可以生成普罗尼系列,其中位移范围为 100 到 250 微米。这些常数被用作模拟实验斜坡/释放剖面的内在属性。在设计深度为 150 微米、压痕速率为 0.1 毫米/秒的条件下,两种不同的水凝胶配方发生了穿刺,其峰值力分别为 3.5 牛顿(H = 31 千帕)和 8.3 牛顿(H = 36.5 千帕)。在模拟印刷分辨率时,峰值力趋势的实验结果与理论结果一致。通过显微镜和轮廓测量法证实了塞子所推断的更高穿刺力和均匀性。同时,还需要进行孔隙弹性表征,以通过比重计区分压痕后的质量损失与再分布。本文的研究结果首次在固体 HMN 文献中强调了在表皮厚度内控制插入深度的可行性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.30
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书