{"title":"Communication Under Sharply Degraded Auditory Input and the \"2-Sentence\" Problem.","authors":"Mario A Svirsky, Jonathan D Neukam, Nicole Hope Capach, Nicole M Amichetti, Annette Lavender, Arthur Wingfield","doi":"10.1097/AUD.0000000000001500","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Despite performing well in standard clinical assessments of speech perception, many cochlear implant (CI) users report experiencing significant difficulties when listening in real-world environments. We hypothesize that this disconnect may be related, in part, to the limited ecological validity of tests that are currently used clinically and in research laboratories. The challenges that arise from degraded auditory information provided by a CI, combined with the listener's finite cognitive resources, may lead to difficulties when processing speech material that is more demanding than the single words or single sentences that are used in clinical tests.</p><p><strong>Design: </strong>Here, we investigate whether speech identification performance and processing effort (indexed by pupil dilation measures) are affected when CI users or normal-hearing control subjects are asked to repeat two sentences presented sequentially instead of just one sentence.</p><p><strong>Results: </strong>Response accuracy was minimally affected in normal-hearing listeners, but CI users showed a wide range of outcomes, from no change to decrements of up to 45 percentage points. The amount of decrement was not predictable from the CI users' performance in standard clinical tests. Pupillometry measures tracked closely with task difficulty in both the CI group and the normal-hearing group, even though the latter had speech perception scores near ceiling levels for all conditions.</p><p><strong>Conclusions: </strong>Speech identification performance is significantly degraded in many (but not all) CI users in response to input that is only slightly more challenging than standard clinical tests; specifically, when two sentences are presented sequentially before requesting a response, instead of presenting just a single sentence at a time. This potential \"2-sentence problem\" represents one of the simplest possible scenarios that go beyond presentation of the single words or sentences used in most clinical tests of speech perception, and it raises the possibility that even good performers in single-sentence tests may be seriously impaired by other ecologically relevant manipulations. The present findings also raise the possibility that a clinical version of a 2-sentence test may provide actionable information for counseling and rehabilitating CI users, and for people who interact with them closely.</p>","PeriodicalId":55172,"journal":{"name":"Ear and Hearing","volume":null,"pages":null},"PeriodicalIF":2.6000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ear and Hearing","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/AUD.0000000000001500","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/25 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"AUDIOLOGY & SPEECH-LANGUAGE PATHOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: Despite performing well in standard clinical assessments of speech perception, many cochlear implant (CI) users report experiencing significant difficulties when listening in real-world environments. We hypothesize that this disconnect may be related, in part, to the limited ecological validity of tests that are currently used clinically and in research laboratories. The challenges that arise from degraded auditory information provided by a CI, combined with the listener's finite cognitive resources, may lead to difficulties when processing speech material that is more demanding than the single words or single sentences that are used in clinical tests.
Design: Here, we investigate whether speech identification performance and processing effort (indexed by pupil dilation measures) are affected when CI users or normal-hearing control subjects are asked to repeat two sentences presented sequentially instead of just one sentence.
Results: Response accuracy was minimally affected in normal-hearing listeners, but CI users showed a wide range of outcomes, from no change to decrements of up to 45 percentage points. The amount of decrement was not predictable from the CI users' performance in standard clinical tests. Pupillometry measures tracked closely with task difficulty in both the CI group and the normal-hearing group, even though the latter had speech perception scores near ceiling levels for all conditions.
Conclusions: Speech identification performance is significantly degraded in many (but not all) CI users in response to input that is only slightly more challenging than standard clinical tests; specifically, when two sentences are presented sequentially before requesting a response, instead of presenting just a single sentence at a time. This potential "2-sentence problem" represents one of the simplest possible scenarios that go beyond presentation of the single words or sentences used in most clinical tests of speech perception, and it raises the possibility that even good performers in single-sentence tests may be seriously impaired by other ecologically relevant manipulations. The present findings also raise the possibility that a clinical version of a 2-sentence test may provide actionable information for counseling and rehabilitating CI users, and for people who interact with them closely.
目的:尽管在语音感知的标准临床评估中表现良好,但许多人工耳蜗 (CI) 用户表示在实际环境中聆听时遇到了很大的困难。我们推测,这种脱节可能部分与目前临床和研究实验室使用的测试的生态有效性有限有关。人工耳蜗提供的听觉信息退化所带来的挑战,再加上听者有限的认知资源,可能会导致他们在处理比临床测试中使用的单词或单句要求更高的语音材料时遇到困难:在此,我们研究了当要求 CI 使用者或听力正常的对照组受试者重复连续出现的两个句子而非一个句子时,语音识别能力和处理努力(以瞳孔放大测量为指标)是否会受到影响:结果:正常听力听者的反应准确性受到的影响很小,但 CI 使用者的影响范围很广,从没有变化到下降高达 45 个百分点。从 CI 使用者在标准临床测试中的表现无法预测下降的幅度。瞳孔测量与 CI 组和正常听力组的任务难度密切相关,尽管后者在所有条件下的言语感知得分都接近上限水平:许多(但不是所有)人工耳蜗用户在应对难度仅略高于标准临床测试的输入时,语音识别能力会明显下降;具体来说,就是在要求回答之前连续呈现两个句子,而不是一次只呈现一个句子。这种潜在的 "两句话问题 "是最简单的可能情况之一,它超越了大多数临床言语感知测试中使用的单词或句子的呈现方式,它提出了一种可能性,即即使是在单句测试中表现出色的人,也可能会受到其他生态相关操作的严重影响。本研究结果还提出了一种可能性,即临床版本的双句子测试可以为 CI 使用者的咨询和康复以及与他们密切互动的人提供可操作的信息。
期刊介绍:
From the basic science of hearing and balance disorders to auditory electrophysiology to amplification and the psychological factors of hearing loss, Ear and Hearing covers all aspects of auditory and vestibular disorders. This multidisciplinary journal consolidates the various factors that contribute to identification, remediation, and audiologic and vestibular rehabilitation. It is the one journal that serves the diverse interest of all members of this professional community -- otologists, audiologists, educators, and to those involved in the design, manufacture, and distribution of amplification systems. The original articles published in the journal focus on assessment, diagnosis, and management of auditory and vestibular disorders.