Jhon E Goez-Mora, Natalia Arbeláez-Córdoba, Norman Balcazar-Morales, Pablo S Rivadeneira
{"title":"A concept for human use of real-time and remote monitoring of diabetic subjects using intermittent scanned continuous glucose measurement.","authors":"Jhon E Goez-Mora, Natalia Arbeláez-Córdoba, Norman Balcazar-Morales, Pablo S Rivadeneira","doi":"10.1186/s12938-024-01217-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Flash glucose monitoring systems like the FreeStyle Libre (FSL) sensor have gained popularity for monitoring glucose levels in people with diabetes mellitus. This sensor can be paired with an off-label converted real-time continuous glucose monitor (c-rtCGM) plus an ad hoc computer/smartphone interface for remote real-time monitoring of diabetic subjects, allowing for trend analysis and alarm generation.</p><p><strong>Objectives: </strong>This work evaluates the accuracy and agreement between the FSL sensor and the developed c-rtCGM system. As real-time monitoring is the main feature, the system's connectivity was assessed at 5-min intervals during the trials.</p><p><strong>Methods: </strong>One week of glucose data were collected from 16 type 1 diabetic rats using the FSL sensor and the c-rtCGM. Baseline blood samples were taken the first day before inducing type 1 diabetes with streptozotocin. Once confirmed diabetic rats, FSL and c-rtCGM, were implanted, and to improve data matching between the two monitoring devices, the c-rtCGM was calibrated to the FSL glucometer readings. A factorial design 2 × 3^3 and a second-order regression was used to find the base values of the linear model transformation of the raw data obtained from the sensor. Accuracy, agreement, and connectivity were assessed by median absolute relative difference (Median ARD), range averaging times, Parkes consensus error grid analysis (EGA), and Bland-Altman analysis with a non-parametric approach.</p><p><strong>Results: </strong>Compared to the FSL sensor, the c-rtCGM had an overall Median ARD of 6.58%, with 93.06% of results in zone A when calibration was not carried out. When calibration frequency changed from every 50 h to 1 h, the overall Median ARD improved from 6.68% to 2.41%, respectively. The connectivity evaluation showed that 95% of data was successfully received every 5 min by the computer interface.</p><p><strong>Conclusions and clinical importance: </strong>The results demonstrate the feasibility and reliability of real-time and remote subjects with diabetes monitoring using the developed c-rtCGM system. Performing calibrations relative to the FSL readings increases the accuracy of the data displayed at the interface.</p>","PeriodicalId":8927,"journal":{"name":"BioMedical Engineering OnLine","volume":"23 1","pages":"26"},"PeriodicalIF":2.9000,"publicationDate":"2024-02-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10903066/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"BioMedical Engineering OnLine","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1186/s12938-024-01217-z","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Flash glucose monitoring systems like the FreeStyle Libre (FSL) sensor have gained popularity for monitoring glucose levels in people with diabetes mellitus. This sensor can be paired with an off-label converted real-time continuous glucose monitor (c-rtCGM) plus an ad hoc computer/smartphone interface for remote real-time monitoring of diabetic subjects, allowing for trend analysis and alarm generation.
Objectives: This work evaluates the accuracy and agreement between the FSL sensor and the developed c-rtCGM system. As real-time monitoring is the main feature, the system's connectivity was assessed at 5-min intervals during the trials.
Methods: One week of glucose data were collected from 16 type 1 diabetic rats using the FSL sensor and the c-rtCGM. Baseline blood samples were taken the first day before inducing type 1 diabetes with streptozotocin. Once confirmed diabetic rats, FSL and c-rtCGM, were implanted, and to improve data matching between the two monitoring devices, the c-rtCGM was calibrated to the FSL glucometer readings. A factorial design 2 × 3^3 and a second-order regression was used to find the base values of the linear model transformation of the raw data obtained from the sensor. Accuracy, agreement, and connectivity were assessed by median absolute relative difference (Median ARD), range averaging times, Parkes consensus error grid analysis (EGA), and Bland-Altman analysis with a non-parametric approach.
Results: Compared to the FSL sensor, the c-rtCGM had an overall Median ARD of 6.58%, with 93.06% of results in zone A when calibration was not carried out. When calibration frequency changed from every 50 h to 1 h, the overall Median ARD improved from 6.68% to 2.41%, respectively. The connectivity evaluation showed that 95% of data was successfully received every 5 min by the computer interface.
Conclusions and clinical importance: The results demonstrate the feasibility and reliability of real-time and remote subjects with diabetes monitoring using the developed c-rtCGM system. Performing calibrations relative to the FSL readings increases the accuracy of the data displayed at the interface.
期刊介绍:
BioMedical Engineering OnLine is an open access, peer-reviewed journal that is dedicated to publishing research in all areas of biomedical engineering.
BioMedical Engineering OnLine is aimed at readers and authors throughout the world, with an interest in using tools of the physical and data sciences and techniques in engineering to understand and solve problems in the biological and medical sciences. Topical areas include, but are not limited to:
Bioinformatics-
Bioinstrumentation-
Biomechanics-
Biomedical Devices & Instrumentation-
Biomedical Signal Processing-
Healthcare Information Systems-
Human Dynamics-
Neural Engineering-
Rehabilitation Engineering-
Biomaterials-
Biomedical Imaging & Image Processing-
BioMEMS and On-Chip Devices-
Bio-Micro/Nano Technologies-
Biomolecular Engineering-
Biosensors-
Cardiovascular Systems Engineering-
Cellular Engineering-
Clinical Engineering-
Computational Biology-
Drug Delivery Technologies-
Modeling Methodologies-
Nanomaterials and Nanotechnology in Biomedicine-
Respiratory Systems Engineering-
Robotics in Medicine-
Systems and Synthetic Biology-
Systems Biology-
Telemedicine/Smartphone Applications in Medicine-
Therapeutic Systems, Devices and Technologies-
Tissue Engineering