Validation of the Raycome model M2 ambulatory blood pressure monitor in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018).
{"title":"Validation of the Raycome model M2 ambulatory blood pressure monitor in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018).","authors":"Shijie Yang, Zhanyang Zhou, Huanhuan Miao, Hongye Zhang, Qiong Zhou, Mei Zhai, Yuqing Zhang","doi":"10.1097/MBP.0000000000000697","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The aim of this study was to evaluate the accuracy of the Raycome model M2 oscillometric upper-arm blood pressure (BP) monitor developed for ambulatory BP measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) at rest and during dynamic exercise.</p><p><strong>Method: </strong>Subjects were recruited to fulfill the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in the general population using the same arm sequential BP measurement method. Three cuffs of the test device were used for arm circumference 18-22 cm (small), 22-32 cm (medium) and 32-42 cm (large).</p><p><strong>Results: </strong>For the general validation study, 106 subjects were recruited and 85 were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 0.5 ± 6.2/-0.2 ± 5.1 mmHg (systolic/diastolic). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 5.23/4.50 mmHg (systolic/diastolic). In the ambulatory validation study ( N = 35), the mean difference was 0.4 ± 5.9/-1.1 ± 5.8 mmHg. The Raycome model M2 performed well against the standard in both the general and ambulatory validations and the Bland-Altman plots did not show any systematic variation in the error.</p><p><strong>Conclusion: </strong>These data show that the Raycome model M2 monitor meets the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) and in the ambulatory setting, indicating its suitability for measuring BP in the general population.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11045396/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000697","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/23 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: The aim of this study was to evaluate the accuracy of the Raycome model M2 oscillometric upper-arm blood pressure (BP) monitor developed for ambulatory BP measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) at rest and during dynamic exercise.
Method: Subjects were recruited to fulfill the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in the general population using the same arm sequential BP measurement method. Three cuffs of the test device were used for arm circumference 18-22 cm (small), 22-32 cm (medium) and 32-42 cm (large).
Results: For the general validation study, 106 subjects were recruited and 85 were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 0.5 ± 6.2/-0.2 ± 5.1 mmHg (systolic/diastolic). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 5.23/4.50 mmHg (systolic/diastolic). In the ambulatory validation study ( N = 35), the mean difference was 0.4 ± 5.9/-1.1 ± 5.8 mmHg. The Raycome model M2 performed well against the standard in both the general and ambulatory validations and the Bland-Altman plots did not show any systematic variation in the error.
Conclusion: These data show that the Raycome model M2 monitor meets the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) and in the ambulatory setting, indicating its suitability for measuring BP in the general population.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.