Reliability of a point of care testing blood gas analyzer for measurement of lactate levels in cerebrospinal fluid.

IF 1.3 4区 医学 Q4 MEDICINE, RESEARCH & EXPERIMENTAL
Alberto Fernández Reina, María Dolores López Abellán, Samir Attaibi Hadri, Ángela Puche Candel, María Isabel Díaz López, Eva Pérez Fernández, Manuel Tomás Orgaz Morales, Luis García de Guadiana-Romualdo
{"title":"Reliability of a point of care testing blood gas analyzer for measurement of lactate levels in cerebrospinal fluid.","authors":"Alberto Fernández Reina, María Dolores López Abellán, Samir Attaibi Hadri, Ángela Puche Candel, María Isabel Díaz López, Eva Pérez Fernández, Manuel Tomás Orgaz Morales, Luis García de Guadiana-Romualdo","doi":"10.1080/00365513.2024.2318619","DOIUrl":null,"url":null,"abstract":"<p><p>Analysis of cerebrospinal fluid (CSF), including lactate, is key for diagnosis of acute meningitis. Since blood gas analyzers (BGA) enable rapid and safe blood-lactate measurements, we evaluated the reliability of RAPIDPoint 500 BGA to provide a fast and accurate measure of CSF lactate. In this study, CSF lactate levels were measured by a reference assay and on RAPIDPoint 500 BGA. Comparability was evaluated through difference analysis, using Bland Altman test, and linear regression analysis, using the Passing Bablok test. Agreement rate according to CSF lactate (≥3.5 and <3.5 mmol/L) was calculated using kappa (κ) statistic. Population study included 98 CSF samples. Concerning difference analysis, according to Bland-Altman test, bias was 0.13 mmol/L (CI 95%: -0.26 to 0.52 mmol/L. In regression analysis, according to Passing-Bablok equation a systematic difference between both assays was found. In concordance analysis, the interrate realibility was very high (κ: 0.964). According to our resuls, although a systematic difference was detected when lactate levels were measured on RAPIDPoint 500 BGA, the results from Bland-Altman test and the high agreement rate support that this POCT analyzer could be useful for a early and safe detection of patients with high probability of increased CSF lactate level.</p>","PeriodicalId":21474,"journal":{"name":"Scandinavian Journal of Clinical & Laboratory Investigation","volume":" ","pages":"76-78"},"PeriodicalIF":1.3000,"publicationDate":"2024-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Scandinavian Journal of Clinical & Laboratory Investigation","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/00365513.2024.2318619","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/20 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0

Abstract

Analysis of cerebrospinal fluid (CSF), including lactate, is key for diagnosis of acute meningitis. Since blood gas analyzers (BGA) enable rapid and safe blood-lactate measurements, we evaluated the reliability of RAPIDPoint 500 BGA to provide a fast and accurate measure of CSF lactate. In this study, CSF lactate levels were measured by a reference assay and on RAPIDPoint 500 BGA. Comparability was evaluated through difference analysis, using Bland Altman test, and linear regression analysis, using the Passing Bablok test. Agreement rate according to CSF lactate (≥3.5 and <3.5 mmol/L) was calculated using kappa (κ) statistic. Population study included 98 CSF samples. Concerning difference analysis, according to Bland-Altman test, bias was 0.13 mmol/L (CI 95%: -0.26 to 0.52 mmol/L. In regression analysis, according to Passing-Bablok equation a systematic difference between both assays was found. In concordance analysis, the interrate realibility was very high (κ: 0.964). According to our resuls, although a systematic difference was detected when lactate levels were measured on RAPIDPoint 500 BGA, the results from Bland-Altman test and the high agreement rate support that this POCT analyzer could be useful for a early and safe detection of patients with high probability of increased CSF lactate level.

用于测量脑脊液中乳酸水平的护理点检测血气分析仪的可靠性。
包括乳酸在内的脑脊液(CSF)分析是诊断急性脑膜炎的关键。由于血气分析仪(BGA)能快速、安全地测量血乳酸,我们对 RAPIDPoint 500 BGA 快速、准确测量 CSF 乳酸的可靠性进行了评估。在这项研究中,CSF 乳酸盐水平是通过参考测定法和 RAPIDPoint 500 BGA 进行测量的。通过差异分析(使用 Bland Altman 检验)和线性回归分析(使用 Passing Bablok 检验)对可比性进行了评估。根据 CSF 乳酸盐(≥3.5 和
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.50
自引率
4.80%
发文量
85
审稿时长
4-8 weeks
期刊介绍: The Scandinavian Journal of Clinical and Laboratory Investigation is an international scientific journal covering clinically oriented biochemical and physiological research. Since the launch of the journal in 1949, it has been a forum for international laboratory medicine, closely related to, and edited by, The Scandinavian Society for Clinical Chemistry. The journal contains peer-reviewed articles, editorials, invited reviews, and short technical notes, as well as several supplements each year. Supplements consist of monographs, and symposium and congress reports covering subjects within clinical chemistry and clinical physiology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信