{"title":"Performance evaluation of the new Sysmex XR-Series haematology analyser","authors":"Kenichi Fujimaki , Kornelia Hummel , Immaculate Magonde , Katharina Dammert , Yoshiko Hamaguchi , Konstantinos Mintzas , Jarob Saker , Ondrej Valina , Klaus-Martin Otte","doi":"10.1016/j.plabm.2024.e00370","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>The new XR-Series haematology analyser from Sysmex provides increased throughput and automation, along with a new reagent in WDF channel for optimised WBC differential.</p></div><div><h3>Methods</h3><p>An analytical performance study for the XR analyser was conducted to evaluate the WDF channel parameters in comparison to the instrument specifications. Additionally, 7460 samples were measured on XR and XN analysers to compare selected parameters and flags, and 930 randomly selected samples were further evaluated with microscopy.</p></div><div><h3>Results</h3><p>All investigated aspects of the analytical performance study for the XR fell within the manufacturer specifications. The correlation coefficients between the two systems for the parameters tested were greater than 0.983 for the main CBC and DIFF parameters, greater than 0.909 for the Extended Inflammation Parameters, and greater than 0.932 for the parameters used in the workflow rulesets of the <em>Extended</em> IPU. Similarly high sensitivities for the detection of abnormal cells were observed for the ‘Blasts/Abn Lympho?’ flag (XN: 100%, XR: 99.0%) and WPC abnormal flags (‘Blasts?’ or ‘Abn Lympho?‘) (XN: 97.0%, XR: 96.0%). XN with WPC channel had a 26% reduction of false positive smears compared to XR with 22% reduction, a statistically non-significant difference.</p></div><div><h3>Conclusion</h3><p>The XR analyser had very good analytical performance, and highly comparable results to the predecessor XN analyser in all investigated parameters, flags and workflow aspects.</p></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":"39 ","pages":"Article e00370"},"PeriodicalIF":1.7000,"publicationDate":"2024-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2352551724000167/pdfft?md5=b03d47a7f3e5f233a84c776dd83c29fb&pid=1-s2.0-S2352551724000167-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352551724000167","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
The new XR-Series haematology analyser from Sysmex provides increased throughput and automation, along with a new reagent in WDF channel for optimised WBC differential.
Methods
An analytical performance study for the XR analyser was conducted to evaluate the WDF channel parameters in comparison to the instrument specifications. Additionally, 7460 samples were measured on XR and XN analysers to compare selected parameters and flags, and 930 randomly selected samples were further evaluated with microscopy.
Results
All investigated aspects of the analytical performance study for the XR fell within the manufacturer specifications. The correlation coefficients between the two systems for the parameters tested were greater than 0.983 for the main CBC and DIFF parameters, greater than 0.909 for the Extended Inflammation Parameters, and greater than 0.932 for the parameters used in the workflow rulesets of the Extended IPU. Similarly high sensitivities for the detection of abnormal cells were observed for the ‘Blasts/Abn Lympho?’ flag (XN: 100%, XR: 99.0%) and WPC abnormal flags (‘Blasts?’ or ‘Abn Lympho?‘) (XN: 97.0%, XR: 96.0%). XN with WPC channel had a 26% reduction of false positive smears compared to XR with 22% reduction, a statistically non-significant difference.
Conclusion
The XR analyser had very good analytical performance, and highly comparable results to the predecessor XN analyser in all investigated parameters, flags and workflow aspects.
期刊介绍:
Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.