Plasma Glucose Concentrations in Different Sampling Tubes Measured on Different Glucose Analysers.

Stefan Pleus, Alexandra Beil, Annette Baumstark, Cornelia Haug, Guido Freckmann
{"title":"Plasma Glucose Concentrations in Different Sampling Tubes Measured on Different Glucose Analysers.","authors":"Stefan Pleus, Alexandra Beil, Annette Baumstark, Cornelia Haug, Guido Freckmann","doi":"10.1055/a-2260-3715","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>The German Diabetes Association recommends using sampling tubes with citrate and fluoride additives to diagnose diabetes by oral glucose tolerance test to inhibit glycolysis. The effect of different tubes on measurement results was assessed.</p><p><strong>Materials and methods: </strong>In a first study, an oral glucose tolerance test was performed on 41 participants without anamnestically known diabetes. Venous blood was sampled in two different tubes with citrate/fluoride additives from different manufacturers and one with only lithium-heparin additive. A second study with 42 participants was performed to verify the initial results with an adapted design, in which a third tube with citrate buffer was used, and glucose measurements were performed on two additional devices of another analyser model. Samples were centrifuged either immediately (<5 min incubation time) or after 20 min or 4 h. All glucose measurements were performed in plasma. Glucose concentrations in lithium-heparin tubes with<5 min incubation time served as baseline concentrations.</p><p><strong>Results: </strong>In the first study, glucose concentrations in one of the citrate/fluoride tubes were similar to the baseline. In the other citrate/fluoride tube, markedly lower concentrations (approximately - 5 mg/dL (- 0.28 mmol/L)) were measured. This was reproduced in the verification study for the same analyser, but not with the other analyser model. Lithium-heparin tubes centrifuged after 20 and 240 min showed systematically lower glucose concentrations.</p><p><strong>Conclusions: </strong>The results confirm that glycolysis can be effectively inhibited in citrate/fluoride-containing sampling tubes. However, glucose measurement results of one analyser showed a relevant negative bias in tubes containing liquid citrate buffer.</p>","PeriodicalId":94001,"journal":{"name":"Experimental and clinical endocrinology & diabetes : official journal, German Society of Endocrinology [and] German Diabetes Association","volume":" ","pages":"260-266"},"PeriodicalIF":0.0000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Experimental and clinical endocrinology & diabetes : official journal, German Society of Endocrinology [and] German Diabetes Association","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1055/a-2260-3715","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/2 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: The German Diabetes Association recommends using sampling tubes with citrate and fluoride additives to diagnose diabetes by oral glucose tolerance test to inhibit glycolysis. The effect of different tubes on measurement results was assessed.

Materials and methods: In a first study, an oral glucose tolerance test was performed on 41 participants without anamnestically known diabetes. Venous blood was sampled in two different tubes with citrate/fluoride additives from different manufacturers and one with only lithium-heparin additive. A second study with 42 participants was performed to verify the initial results with an adapted design, in which a third tube with citrate buffer was used, and glucose measurements were performed on two additional devices of another analyser model. Samples were centrifuged either immediately (<5 min incubation time) or after 20 min or 4 h. All glucose measurements were performed in plasma. Glucose concentrations in lithium-heparin tubes with<5 min incubation time served as baseline concentrations.

Results: In the first study, glucose concentrations in one of the citrate/fluoride tubes were similar to the baseline. In the other citrate/fluoride tube, markedly lower concentrations (approximately - 5 mg/dL (- 0.28 mmol/L)) were measured. This was reproduced in the verification study for the same analyser, but not with the other analyser model. Lithium-heparin tubes centrifuged after 20 and 240 min showed systematically lower glucose concentrations.

Conclusions: The results confirm that glycolysis can be effectively inhibited in citrate/fluoride-containing sampling tubes. However, glucose measurement results of one analyser showed a relevant negative bias in tubes containing liquid citrate buffer.

用不同的葡萄糖分析仪测量不同采样管中的血浆葡萄糖浓度。
导言:德国糖尿病协会建议在通过口服葡萄糖耐量试验诊断糖尿病时使用带有柠檬酸盐和氟化物添加剂的采样管,以抑制糖酵解。本研究评估了不同试管对测量结果的影响。材料和方法 在第一项研究中,对 41 名无糖尿病史的参与者进行了口服葡萄糖耐量试验。静脉血采样在两种不同的试管中进行,一种试管使用不同厂家生产的柠檬酸盐/氟添加剂,另一种试管仅使用锂-肝素添加剂。为了验证初步结果,我们又对 42 名参与者进行了第二次研究,并对设计进行了调整,使用了第三支装有柠檬酸盐缓冲液的试管,并在另外两台不同型号的分析仪上进行了血糖测量。样本立即离心 (
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信