Evaluation and experience from routine use of chemiluminescence assays for serological screening of blood and plasma donations on the Alinity s system and the Alinity i system, two new fully-automated immunoassay systems in Poland
{"title":"Evaluation and experience from routine use of chemiluminescence assays for serological screening of blood and plasma donations on the Alinity s system and the Alinity i system, two new fully-automated immunoassay systems in Poland","authors":"Aneta Kopacz , Dorota Kubicka-Russel , Grzegorz Liszewski , Alicja Bukowska , Sylwia Samek , Dorota Malka , Magdalena Łętowska , Piotr Grabarczyk","doi":"10.1016/j.plabm.2024.e00364","DOIUrl":null,"url":null,"abstract":"<div><p>In Poland, independent evaluations under the auspices of the Institute of Hematology and Transfusion Medicine (IHTM) are mandated for any new device, assay, systems for screening samples from whole blood and plasma donors prior to implementation by Blood Transfusion Center (BTC). In last 5 years, two new systems were introduced to the market by Abbott GmbH, namely the Alinity s and the Alinity i. The evaluations performed for these two systems included the assessment of sensitivity, specificity and precision for each of the four mandatory serological screening markers in Poland: Hepatitis B Surface Antigen (HBsAg), Hepatitis C virus antibodies (Anti-HCV), HIV antibodies (anti-HIV) and Syphilis antibodies (anti-<em>Treponema pallidum,</em> anti-TP). Sensitivity was assessed by testing seroconversion panels, HBsAg international reference standard, well characterized local samples, and dilution panels. Specificity was assessed by testing routine donor samples. The results from Alinity i assays were compared to the results from Abbott ARCHITECT i2000SR and Ortho VITROS 3600 assays, while the results from Alinity s assays were compared to the results of ARCHITECT i2000SR assays. The evaluation of the Alinity s and Alinity i assays for sensitivity (100 %), specificity (99,92–100 %) and precision generated results that were as good as or better than generated by routinely used systems, were within acceptance criteria, and met all requirements for screening blood donor samples in accordance with Polish regulations. The specificity of the assays in routine use by BTCs, analyzed after approximately 150,000 donations on both systems, was comparable to the specificity observed during the evaluations at IHTM.</p></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":"39 ","pages":"Article e00364"},"PeriodicalIF":1.7000,"publicationDate":"2024-01-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2352551724000106/pdfft?md5=2084406dcb4697eab73ce30be4c33222&pid=1-s2.0-S2352551724000106-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352551724000106","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
In Poland, independent evaluations under the auspices of the Institute of Hematology and Transfusion Medicine (IHTM) are mandated for any new device, assay, systems for screening samples from whole blood and plasma donors prior to implementation by Blood Transfusion Center (BTC). In last 5 years, two new systems were introduced to the market by Abbott GmbH, namely the Alinity s and the Alinity i. The evaluations performed for these two systems included the assessment of sensitivity, specificity and precision for each of the four mandatory serological screening markers in Poland: Hepatitis B Surface Antigen (HBsAg), Hepatitis C virus antibodies (Anti-HCV), HIV antibodies (anti-HIV) and Syphilis antibodies (anti-Treponema pallidum, anti-TP). Sensitivity was assessed by testing seroconversion panels, HBsAg international reference standard, well characterized local samples, and dilution panels. Specificity was assessed by testing routine donor samples. The results from Alinity i assays were compared to the results from Abbott ARCHITECT i2000SR and Ortho VITROS 3600 assays, while the results from Alinity s assays were compared to the results of ARCHITECT i2000SR assays. The evaluation of the Alinity s and Alinity i assays for sensitivity (100 %), specificity (99,92–100 %) and precision generated results that were as good as or better than generated by routinely used systems, were within acceptance criteria, and met all requirements for screening blood donor samples in accordance with Polish regulations. The specificity of the assays in routine use by BTCs, analyzed after approximately 150,000 donations on both systems, was comparable to the specificity observed during the evaluations at IHTM.
期刊介绍:
Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.