Robin H.J. Kemperman , Rebecca D. Ganetzky , Stephen R. Master
{"title":"Development and validation of a multiplexed LC-MS/MS ketone body assay for clinical diagnostics","authors":"Robin H.J. Kemperman , Rebecca D. Ganetzky , Stephen R. Master","doi":"10.1016/j.jmsacl.2024.01.004","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><p>Ketone bodies (KBs) serve as important energy sources that spare glucose, providing the primary energy for cardiac muscle, skeletal muscle during aerobic exercise, and the brain during periods of catabolism. The levels and relationships between the KBs are critical indicators of metabolic health and disease. However, challenges in separating isomeric KBs and concerns about sample stability have previously limited their clinical measurement.</p></div><div><h3>Methods</h3><p>A novel 6.5-minute liquid chromatography-mass spectrometry-based assay was developed, enabling the precise measurement of alpha-, beta- and gamma-hydroxybutyrate, beta-hydroxyisobutyrate, and acetoacetate. This method was fully validated for human serum and plasma samples by investigating extraction efficiency, matrix effects, accuracy, recovery, intra- and inter-precision, linearity, lower limit of quantitation (LLOQ), carryover, specificity, stability, and more. From 107 normal samples, reference ranges were established for all analytes and the beta-hydroxybutyrate/acetoacetate ratio.</p></div><div><h3>Results</h3><p>All five analytes were adequately separated chromatographically. An extraction efficiency between 80 and 120 % was observed for all KBs. Accuracy was evaluated through spike and recovery using 10 random patient samples, with an average recovery of 85–115 % for all KBs and a coefficient of variation of ≤ 3 %. Coefficients of variation for intra- and inter-day imprecision were < 5 %, and the total imprecision was < 10 %. No significant interferences were observed. Specimens remained stable for up to 6 h on ice or 2 h at room temperature.</p></div><div><h3>Conclusions</h3><p>The developed method is highly sensitive and robust. It has been validated for use with human serum and plasma, overcoming stability concerns and providing a reliable and efficient quantitative estimation of ketone bodies.</p></div>","PeriodicalId":52406,"journal":{"name":"Journal of Mass Spectrometry and Advances in the Clinical Lab","volume":"31 ","pages":"Pages 49-58"},"PeriodicalIF":3.1000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2667145X2400004X/pdfft?md5=ecb251e99bcd8539ebb37c657c813fcf&pid=1-s2.0-S2667145X2400004X-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Mass Spectrometry and Advances in the Clinical Lab","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2667145X2400004X","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives
Ketone bodies (KBs) serve as important energy sources that spare glucose, providing the primary energy for cardiac muscle, skeletal muscle during aerobic exercise, and the brain during periods of catabolism. The levels and relationships between the KBs are critical indicators of metabolic health and disease. However, challenges in separating isomeric KBs and concerns about sample stability have previously limited their clinical measurement.
Methods
A novel 6.5-minute liquid chromatography-mass spectrometry-based assay was developed, enabling the precise measurement of alpha-, beta- and gamma-hydroxybutyrate, beta-hydroxyisobutyrate, and acetoacetate. This method was fully validated for human serum and plasma samples by investigating extraction efficiency, matrix effects, accuracy, recovery, intra- and inter-precision, linearity, lower limit of quantitation (LLOQ), carryover, specificity, stability, and more. From 107 normal samples, reference ranges were established for all analytes and the beta-hydroxybutyrate/acetoacetate ratio.
Results
All five analytes were adequately separated chromatographically. An extraction efficiency between 80 and 120 % was observed for all KBs. Accuracy was evaluated through spike and recovery using 10 random patient samples, with an average recovery of 85–115 % for all KBs and a coefficient of variation of ≤ 3 %. Coefficients of variation for intra- and inter-day imprecision were < 5 %, and the total imprecision was < 10 %. No significant interferences were observed. Specimens remained stable for up to 6 h on ice or 2 h at room temperature.
Conclusions
The developed method is highly sensitive and robust. It has been validated for use with human serum and plasma, overcoming stability concerns and providing a reliable and efficient quantitative estimation of ketone bodies.