Testing the limits of multiplex respiratory virus assays for SARS-CoV-2 at high cycle threshold values: Comparative performance of cobas 6800/8800 SARS-CoV-2 & Influenza A/B, Xpert Xpress SARS-CoV-2/Flu/RSV, and cobas Liat SARS-CoV-2 & Influenza A/B.
Nancy Matic, Tanya Lawson, Gordon Ritchie, Christopher F Lowe, Marc G Romney
{"title":"Testing the limits of multiplex respiratory virus assays for SARS-CoV-2 at high cycle threshold values: Comparative performance of cobas 6800/8800 SARS-CoV-2 & Influenza A/B, Xpert Xpress SARS-CoV-2/Flu/RSV, and cobas Liat SARS-CoV-2 & Influenza A/B.","authors":"Nancy Matic, Tanya Lawson, Gordon Ritchie, Christopher F Lowe, Marc G Romney","doi":"10.3138/jammi-2022-0039","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Multiplex real-time RT-PCR assays for respiratory pathogens are valuable tools to optimize laboratory workflow and turnaround time. At a time when resurgence of influenza and respiratory syncytial virus (RSV) cases have been widely observed along with continued transmission of SARS-CoV-2, timely identification of all circulating respiratory viruses is crucial. This study evaluates the detection of low viral loads of SARS-CoV-2 by four multiplex molecular assays: Roche cobas 6800/8800 SARS-CoV-2 & Influenza A/B Test, Cepheid Xpert Xpress SARS-CoV-2/Flu/RSV, cobas Liat SARS-CoV-2 & Influenza A/B, and a laboratory-developed test (LDT).</p><p><strong>Methods: </strong>Retrospective upper respiratory tract specimens positive for various respiratory viruses at a range of cycle threshold (Ct) values (18-40) were tested by four multiplex assays. Positive and negative percent agreement (PPA and NPA) with validated RT-PCR assays were calculated.</p><p><strong>Results: </strong>A total of 82 samples were assessed, with discordant results observed in a portion of the samples (10/82, 12.2%) where Ct values were >33. The majority of the discordant results (6/10, 60%) were false negatives. Overall, PPA was 100% (58/58) for cobas 6800, 97.4% (38/39) for GeneXpert, 100% (17/17) for Liat, and 90.5% (57/63) for the LDT. PPA for the LDT increased to 92.1% after manual review of amplification curves.</p><p><strong>Conclusions: </strong>Commercial multiplex respiratory virus assays have good performance for samples with medium to high viral loads (Ct values <33). Laboratories should consider appropriate test result review and confirmation protocols to optimize sensitivity, and may consider reporting samples with additional interpretive comments when low viral loads are detected.</p>","PeriodicalId":36782,"journal":{"name":"JAMMI","volume":"8 4","pages":"328-335"},"PeriodicalIF":0.0000,"publicationDate":"2024-01-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10797767/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JAMMI","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3138/jammi-2022-0039","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Multiplex real-time RT-PCR assays for respiratory pathogens are valuable tools to optimize laboratory workflow and turnaround time. At a time when resurgence of influenza and respiratory syncytial virus (RSV) cases have been widely observed along with continued transmission of SARS-CoV-2, timely identification of all circulating respiratory viruses is crucial. This study evaluates the detection of low viral loads of SARS-CoV-2 by four multiplex molecular assays: Roche cobas 6800/8800 SARS-CoV-2 & Influenza A/B Test, Cepheid Xpert Xpress SARS-CoV-2/Flu/RSV, cobas Liat SARS-CoV-2 & Influenza A/B, and a laboratory-developed test (LDT).
Methods: Retrospective upper respiratory tract specimens positive for various respiratory viruses at a range of cycle threshold (Ct) values (18-40) were tested by four multiplex assays. Positive and negative percent agreement (PPA and NPA) with validated RT-PCR assays were calculated.
Results: A total of 82 samples were assessed, with discordant results observed in a portion of the samples (10/82, 12.2%) where Ct values were >33. The majority of the discordant results (6/10, 60%) were false negatives. Overall, PPA was 100% (58/58) for cobas 6800, 97.4% (38/39) for GeneXpert, 100% (17/17) for Liat, and 90.5% (57/63) for the LDT. PPA for the LDT increased to 92.1% after manual review of amplification curves.
Conclusions: Commercial multiplex respiratory virus assays have good performance for samples with medium to high viral loads (Ct values <33). Laboratories should consider appropriate test result review and confirmation protocols to optimize sensitivity, and may consider reporting samples with additional interpretive comments when low viral loads are detected.