Syed Faisal, Michael Barbour, Eric J Seibel, Alberto Aliseda
{"title":"Hemodynamics of Saline Flushing in Endoscopic Imaging of Partially Occluded Coronary Arteries.","authors":"Syed Faisal, Michael Barbour, Eric J Seibel, Alberto Aliseda","doi":"10.1007/s13239-023-00708-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Intravascular endoscopy can aid in the diagnosis of coronary atherosclerosis by providing direct color images of coronary plaques. The procedure requires a blood-free optical path between the catheter and plaque, and achieving clearance safely remains an engineering challenge. In this study, we investigate the hemodynamics of saline flushing in partially occluded coronary arteries to advance the development of intravascular forward-imaging catheters that do not require balloon occlusion.</p><p><strong>Methods: </strong>In-vitro experiments and CFD simulations are used to quantify the influence of plaque size, catheter stand-off distance, saline injection flowrate, and injection orientation on the time required to achieve blood clearance.</p><p><strong>Results: </strong>Experiments and simulation of saline injection from a dual-lumen catheter demonstrated that flushing times increase both as injection flow rate (Reynolds number) decreases and as the catheter moves distally away from the plaque. CFD simulations demonstrated that successful flushing was achieved regardless of lumen axial orientation in a 95% occluded artery. Flushing time was also found to increase as plaque size decreases for a set injection flowrate, and a lower limit for injection flowrate was found to exist for each plaques size, below which clearance was not achieved. For the three occlusion sizes investigated (90, 95, 97% by area), successful occlusion was achieved in less than 1.2 s. Investigation of the pressure fields developed during injection, highlight that rapid clearance can be achieved while keeping the arterial overpressure to < 1 mmHg.</p><p><strong>Conclusions: </strong>A dual lumen saline injection catheter was shown to produce clearance safely and effectively in models of partially occluded coronary arteries. Clearance was achieved across a range of engineering and clinical parameters without the use of a balloon occlusion, providing development guideposts for a fluid injection system in forward-imaging coronary endoscopes.</p>","PeriodicalId":54322,"journal":{"name":"Cardiovascular Engineering and Technology","volume":" ","pages":"211-223"},"PeriodicalIF":1.6000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiovascular Engineering and Technology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s13239-023-00708-9","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/8 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Intravascular endoscopy can aid in the diagnosis of coronary atherosclerosis by providing direct color images of coronary plaques. The procedure requires a blood-free optical path between the catheter and plaque, and achieving clearance safely remains an engineering challenge. In this study, we investigate the hemodynamics of saline flushing in partially occluded coronary arteries to advance the development of intravascular forward-imaging catheters that do not require balloon occlusion.
Methods: In-vitro experiments and CFD simulations are used to quantify the influence of plaque size, catheter stand-off distance, saline injection flowrate, and injection orientation on the time required to achieve blood clearance.
Results: Experiments and simulation of saline injection from a dual-lumen catheter demonstrated that flushing times increase both as injection flow rate (Reynolds number) decreases and as the catheter moves distally away from the plaque. CFD simulations demonstrated that successful flushing was achieved regardless of lumen axial orientation in a 95% occluded artery. Flushing time was also found to increase as plaque size decreases for a set injection flowrate, and a lower limit for injection flowrate was found to exist for each plaques size, below which clearance was not achieved. For the three occlusion sizes investigated (90, 95, 97% by area), successful occlusion was achieved in less than 1.2 s. Investigation of the pressure fields developed during injection, highlight that rapid clearance can be achieved while keeping the arterial overpressure to < 1 mmHg.
Conclusions: A dual lumen saline injection catheter was shown to produce clearance safely and effectively in models of partially occluded coronary arteries. Clearance was achieved across a range of engineering and clinical parameters without the use of a balloon occlusion, providing development guideposts for a fluid injection system in forward-imaging coronary endoscopes.
期刊介绍:
Cardiovascular Engineering and Technology is a journal publishing the spectrum of basic to translational research in all aspects of cardiovascular physiology and medical treatment. It is the forum for academic and industrial investigators to disseminate research that utilizes engineering principles and methods to advance fundamental knowledge and technological solutions related to the cardiovascular system. Manuscripts spanning from subcellular to systems level topics are invited, including but not limited to implantable medical devices, hemodynamics and tissue biomechanics, functional imaging, surgical devices, electrophysiology, tissue engineering and regenerative medicine, diagnostic instruments, transport and delivery of biologics, and sensors. In addition to manuscripts describing the original publication of research, manuscripts reviewing developments in these topics or their state-of-art are also invited.