{"title":"Validation of the DBP-1333b upper-arm blood pressure monitor according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020).","authors":"Guocun Hou, Yulu Wu, Jiaying Wang, Jia Zhi","doi":"10.1097/MBP.0000000000000688","DOIUrl":null,"url":null,"abstract":"<p><p>To evaluate the accuracy of the DBP-1333b upper-arm blood pressure (BP) measuring device in the adult population according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020). Subjects were recruited in the adult population. The test device was an arm-type electronic sphygmomanometer (DBP-1333b) and the reference device was a desktop sphygmomanometer (XJ11D). Using the BP data measured by the desktop sphygmomanometer as reference BP, the accuracy of the non-invasive BP module of the test device was evaluated to determine whether it met the requirements. Data from 90 individuals were analysed. According to Criterion 1, the mean difference of SBP between the test and reference device was 0.19 mmHg and the SD was 7.45 mmHg. The mean difference of DBP was -0.59 mmHg and the SD was 6.47 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, SD of SBP was 5.79 mmHg, which was less than 6.95 mmHg and met the requirements. The SD of DBP was 5.58 mmHg, which was less than 6.93 mmHg and met the requirements. It was concluded that the DBP-1333b complies with the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020) and can be recommended for use by the adults.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":"149-155"},"PeriodicalIF":1.2000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11045392/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000688","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/4/24 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
To evaluate the accuracy of the DBP-1333b upper-arm blood pressure (BP) measuring device in the adult population according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020). Subjects were recruited in the adult population. The test device was an arm-type electronic sphygmomanometer (DBP-1333b) and the reference device was a desktop sphygmomanometer (XJ11D). Using the BP data measured by the desktop sphygmomanometer as reference BP, the accuracy of the non-invasive BP module of the test device was evaluated to determine whether it met the requirements. Data from 90 individuals were analysed. According to Criterion 1, the mean difference of SBP between the test and reference device was 0.19 mmHg and the SD was 7.45 mmHg. The mean difference of DBP was -0.59 mmHg and the SD was 6.47 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, SD of SBP was 5.79 mmHg, which was less than 6.95 mmHg and met the requirements. The SD of DBP was 5.58 mmHg, which was less than 6.93 mmHg and met the requirements. It was concluded that the DBP-1333b complies with the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020) and can be recommended for use by the adults.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.