Danyal Z Khan, Kanza Tariq, Keng Siang Lee, Edward W Dyson, Vittorio Russo, Laurence D Watkins, Antonino Russo
{"title":"Patient-specific automated cerebrospinal fluid pressure control to augment spinal wound closure: a case series using the LiquoGuard®.","authors":"Danyal Z Khan, Kanza Tariq, Keng Siang Lee, Edward W Dyson, Vittorio Russo, Laurence D Watkins, Antonino Russo","doi":"10.1080/02688697.2023.2290101","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Spinal cerebrospinal fluid (CSF) leaks are common, and their management is heterogeneous. For high-flow leaks, numerous studies advocate for primary dural repair and CSF diversion. The LiquoGuard7® allows automated and precise pressure and volume control, and calculation of patient-specific CSF production rate (prCSF), which is hypothesized to be increased in the context of durotomies and CSF leaks.</p><p><strong>Methods: </strong>This single-centre illustrative case series included patients undergoing complex spinal surgery where: 1) a high flow intra-operative and/or post-operative CSF leak was expected and 2) lumbar CSF drainage was performed using a LiquoGuard7®. CSF diversion was tailored to prCSF for each patient, combined with layered spinal wound closure.</p><p><strong>Results: </strong>Three patients were included, with a variety of pathologies: T7/T8 disc prolapse, T8-T9 meningioma, and T4-T5 metastatic spinal cord compression. The first two patients underwent CSF diversion to prevent post-op CSF leak, whilst the third required this in response to post-op CSF leak. CSF hyperproduction was evident in all cases (mean >/=140ml/hr). With patient-specific CSF diversion regimes, no cases required further intervention for CSF fistulae repair (including for pleural CSF effusion), wound breakdown or infection.</p><p><strong>Conclusions: </strong>Patient-specific cerebrospinal fluid drainage may be a useful tool in the management of high-flow intra-operative and post-operative CSF leaks during complex spinal surgery. These systems may reduce post-operative CSF leakage from the wound or into adjacent body cavities. Further larger studies are needed to evaluate the comparative benefits and cost-effectiveness of this approach.</p>","PeriodicalId":9261,"journal":{"name":"British Journal of Neurosurgery","volume":" ","pages":"476-484"},"PeriodicalIF":0.8000,"publicationDate":"2025-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11013023/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"British Journal of Neurosurgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/02688697.2023.2290101","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/4 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Spinal cerebrospinal fluid (CSF) leaks are common, and their management is heterogeneous. For high-flow leaks, numerous studies advocate for primary dural repair and CSF diversion. The LiquoGuard7® allows automated and precise pressure and volume control, and calculation of patient-specific CSF production rate (prCSF), which is hypothesized to be increased in the context of durotomies and CSF leaks.
Methods: This single-centre illustrative case series included patients undergoing complex spinal surgery where: 1) a high flow intra-operative and/or post-operative CSF leak was expected and 2) lumbar CSF drainage was performed using a LiquoGuard7®. CSF diversion was tailored to prCSF for each patient, combined with layered spinal wound closure.
Results: Three patients were included, with a variety of pathologies: T7/T8 disc prolapse, T8-T9 meningioma, and T4-T5 metastatic spinal cord compression. The first two patients underwent CSF diversion to prevent post-op CSF leak, whilst the third required this in response to post-op CSF leak. CSF hyperproduction was evident in all cases (mean >/=140ml/hr). With patient-specific CSF diversion regimes, no cases required further intervention for CSF fistulae repair (including for pleural CSF effusion), wound breakdown or infection.
Conclusions: Patient-specific cerebrospinal fluid drainage may be a useful tool in the management of high-flow intra-operative and post-operative CSF leaks during complex spinal surgery. These systems may reduce post-operative CSF leakage from the wound or into adjacent body cavities. Further larger studies are needed to evaluate the comparative benefits and cost-effectiveness of this approach.
期刊介绍:
The British Journal of Neurosurgery is a leading international forum for debate in the field of neurosurgery, publishing original peer-reviewed articles of the highest quality, along with comment and correspondence on all topics of current interest to neurosurgeons worldwide.
Coverage includes all aspects of case assessment and surgical practice, as well as wide-ranging research, with an emphasis on clinical rather than experimental material. Special emphasis is placed on postgraduate education with review articles on basic neurosciences and on the theory behind advances in techniques, investigation and clinical management. All papers are submitted to rigorous and independent peer-review, ensuring the journal’s wide citation and its appearance in the major abstracting and indexing services.