The Accuracy of Different Apex Locator Systems in Detecting Root Perforations in the Presence of Different Irrigation Solutions

IF 0.3 Q3 MEDICINE, GENERAL & INTERNAL
Oğuz Burhan Çetinkaya, Emre Çulha, Uğur Aydın
{"title":"The Accuracy of Different Apex Locator Systems in Detecting Root Perforations in the Presence of Different Irrigation Solutions","authors":"Oğuz Burhan Çetinkaya, Emre Çulha, Uğur Aydın","doi":"10.58600/eurjther1936","DOIUrl":null,"url":null,"abstract":"Objective: One of the most studied topics in electronic apex locators (EALs) is the effect of root canal condition on the accuracy of EALs. In this study, the accuracy of Root ZX Mini, Raypex 6, and Apex ID in detecting root perforation was evaluated in a dry environment and in the presence of saline, ethylenediamine tetraacetic acid (EDTA), and sodium hypochlorite (NaOCl) solutions.\nMethods: The mesiobuccal roots of 64 human maxillary first molars were selected for the study. These root canals were perforated from buccal root surface using a #1 Freze Beutherlock Peeso to form a 0.4 mm cavity. After perforation, the area where #40 K type file appeared was recorded as the actual length (AL) and the length measured by EAL devices both in dry canals and in the presence of solutions was recorded as the electronic measurement (EL). All electronic measurements were statistically compared with the actual length.\nResults: All EALs achieved AL-consistent results on EL measures. Consistency was determined using Root ZX Mini measurements in dry canals and canals irrigated with saline. The consistency of EL and AL conducted with Raypex 6 and Apex ID in canals irrigated with NaOCl, saline, and EDTA was found. There was a statistically significant difference in ELs with irrigation solutions among all EALs (p<0.05). In the dry environment, there was no statistically significant difference between the EALs (p>0.05). A consistency was discovered between ELs generated with the Root ZX Mini and ALs in both dry and saline-irrigated canals. Consistency was observed between ELs measured with Raypex 6 and Apex ID and ALs in canals irrigated with NaOCl, saline, and EDTA.\nConclusion: The accuracy rates of the EALs used in this research were 97%–100% in the 1 mm range and 83%–92% in the 0.5 mm range. Despite the fact that ALs and ELs differed statistically significantly at the 0.05 level, these variations weren't thought to be clinically relevant. In the presence of conditions with different electro conductors, EALs from different generations may be used safely, and in a range of canal situations, these devices can yield measurements that are most similar to the AL.","PeriodicalId":42642,"journal":{"name":"European Journal of Therapeutics","volume":null,"pages":null},"PeriodicalIF":0.3000,"publicationDate":"2023-12-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Therapeutics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.58600/eurjther1936","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: One of the most studied topics in electronic apex locators (EALs) is the effect of root canal condition on the accuracy of EALs. In this study, the accuracy of Root ZX Mini, Raypex 6, and Apex ID in detecting root perforation was evaluated in a dry environment and in the presence of saline, ethylenediamine tetraacetic acid (EDTA), and sodium hypochlorite (NaOCl) solutions. Methods: The mesiobuccal roots of 64 human maxillary first molars were selected for the study. These root canals were perforated from buccal root surface using a #1 Freze Beutherlock Peeso to form a 0.4 mm cavity. After perforation, the area where #40 K type file appeared was recorded as the actual length (AL) and the length measured by EAL devices both in dry canals and in the presence of solutions was recorded as the electronic measurement (EL). All electronic measurements were statistically compared with the actual length. Results: All EALs achieved AL-consistent results on EL measures. Consistency was determined using Root ZX Mini measurements in dry canals and canals irrigated with saline. The consistency of EL and AL conducted with Raypex 6 and Apex ID in canals irrigated with NaOCl, saline, and EDTA was found. There was a statistically significant difference in ELs with irrigation solutions among all EALs (p<0.05). In the dry environment, there was no statistically significant difference between the EALs (p>0.05). A consistency was discovered between ELs generated with the Root ZX Mini and ALs in both dry and saline-irrigated canals. Consistency was observed between ELs measured with Raypex 6 and Apex ID and ALs in canals irrigated with NaOCl, saline, and EDTA. Conclusion: The accuracy rates of the EALs used in this research were 97%–100% in the 1 mm range and 83%–92% in the 0.5 mm range. Despite the fact that ALs and ELs differed statistically significantly at the 0.05 level, these variations weren't thought to be clinically relevant. In the presence of conditions with different electro conductors, EALs from different generations may be used safely, and in a range of canal situations, these devices can yield measurements that are most similar to the AL.
不同灌溉溶液条件下不同牙尖定位系统检测牙根穿孔的准确性
目的:电子根尖定位仪(EAL)研究最多的课题之一是根管状况对 EAL 精确度的影响。本研究评估了 Root ZX Mini、Raypex 6 和 Apex ID 在干燥环境和生理盐水、乙二胺四乙酸 (EDTA) 和次氯酸钠 (NaOCl) 溶液中检测根穿孔的准确性:研究选取了 64 颗人类上颌第一磨牙的中颊根。使用 1 号 Freze Beutherlock Peeso 从颊面根表面对这些根管进行穿孔,形成 0.4 毫米的空腔。穿孔后,40 号 K 型锉出现的区域被记录为实际长度 (AL),而在干燥根管和溶液存在的情况下通过 EAL 设备测量的长度被记录为电子测量值 (EL)。所有电子测量值都与实际长度进行了统计比较:结果:所有 EAL 在 EL 测量上都达到了与 AL 一致的结果。使用 Root ZX Mini 对干燥的根管和生理盐水灌溉的根管进行测量,可以确定测量结果的一致性。使用 Raypex 6 和 Apex ID 在用 NaOCl、生理盐水和 EDTA 灌溉的根管中进行的 EL 和 AL 测量结果一致。在所有 EAL 中,使用灌溉溶液的 EL 有显著的统计学差异(P0.05)。使用 Root ZX Mini 生成的 ELs 与干燥和盐水灌溉下的 ALs 之间具有一致性。用 Raypex 6 和 Apex ID 测量的 EL 与用 NaOCl、生理盐水和 EDTA 灌溉的牙管中的 AL 之间也有一致性:本研究中使用的 EAL 在 1 毫米范围内的准确率为 97%-100%,在 0.5 毫米范围内的准确率为 83%-92%。尽管 AL 和 EL 在 0.05 的统计学水平上存在显著差异,但这些差异被认为与临床无关。在存在不同电导体的条件下,可以安全地使用不同世代的EAL,而且在一系列的牙槽骨情况下,这些设备可以获得与AL最相似的测量结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
European Journal of Therapeutics
European Journal of Therapeutics MEDICINE, GENERAL & INTERNAL-
自引率
0.00%
发文量
48
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信