Resonance frequency analysis – indicator of post-implantation morphology of mandibular bone tissue

A. P. Oshurko, I. Yu, C. Oliinyk A, А. П. Ошурко, І. Ю. Олійник, Н. Б. Кузняк, Л. М. Герасим
{"title":"Resonance frequency analysis – indicator of post-implantation morphology of mandibular bone tissue","authors":"A. P. Oshurko, I. Yu, C. Oliinyk A, А. П. Ошурко, І. Ю. Олійник, Н. Б. Кузняк, Л. М. Герасим","doi":"10.34287/mmt.4(59).2023.9","DOIUrl":null,"url":null,"abstract":"This article describes the role of resonance frequency analysis (RFA) as an indicator of the quality of primary implant stability, which is the basis of the goal and confirmed by the results of a clinical case – odontological implantation on edentulous segments of atrophied bone tissue on the right side of the mandible, taking into account the topographic features of its canal, using ultra-short (h = 5.5 mm) implants. \nAim. To highlight the role of resonance frequency analysis during surgical interventions using short (ultrashort) implants in predicting their primary stability and preserving the normal morphology of the mandibular bone tissue. \nMaterials and methods. The primary stability of 17 dental implants in seven patients was determined using the Penguin Instruments (RFA) technique in declarative units of measurement – the implant stability quotient (ISQ). Values in the range of 75 ISQ meant that the implant had adequate primary stability and could receive and distribute the immediate load on the bone tissue of the edentulous segment. \nResults. Following the stages of preparation, 17 ultra-short (h = 5.5 mm × b = 4.0 mm) implants were placed on the edentulous segments of the mandible without an incisional osteotomy path according to the patient rehabilitation protocol. We used the method of resonance frequency analysis to obtain a value of high primary stability of two short subcortical implants in the projection of tooth 4.7 – 92 ISQ and the projection of tooth 4.6 – 90 ISQ. On the 3rd day of the clinical stage of fixation of the supraconstruction, RFA measurements were performed, the values of which showed a significant decrease in the stability quotient of the installed implants in the projection of tooth 4.7 – 61 ISQ and in tooth 4.6 – 74 ISQ. This result made it necessary to revise the rehabilitation plan, adjust the medical prescription sheet and decide to postpone the placement of the all-milled restorative structure on the implants. \nConclusions. The RFA study indicates not only the primary stability of implants but also a full-fledged stage-bystage clinical prognosis of the further functionality of a prosthetic structure based on short implants with proper preservation of the morphological functionality of bone tissue.","PeriodicalId":199868,"journal":{"name":"Modern medical technology","volume":"15 11","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-12-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Modern medical technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.34287/mmt.4(59).2023.9","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

This article describes the role of resonance frequency analysis (RFA) as an indicator of the quality of primary implant stability, which is the basis of the goal and confirmed by the results of a clinical case – odontological implantation on edentulous segments of atrophied bone tissue on the right side of the mandible, taking into account the topographic features of its canal, using ultra-short (h = 5.5 mm) implants. Aim. To highlight the role of resonance frequency analysis during surgical interventions using short (ultrashort) implants in predicting their primary stability and preserving the normal morphology of the mandibular bone tissue. Materials and methods. The primary stability of 17 dental implants in seven patients was determined using the Penguin Instruments (RFA) technique in declarative units of measurement – the implant stability quotient (ISQ). Values in the range of 75 ISQ meant that the implant had adequate primary stability and could receive and distribute the immediate load on the bone tissue of the edentulous segment. Results. Following the stages of preparation, 17 ultra-short (h = 5.5 mm × b = 4.0 mm) implants were placed on the edentulous segments of the mandible without an incisional osteotomy path according to the patient rehabilitation protocol. We used the method of resonance frequency analysis to obtain a value of high primary stability of two short subcortical implants in the projection of tooth 4.7 – 92 ISQ and the projection of tooth 4.6 – 90 ISQ. On the 3rd day of the clinical stage of fixation of the supraconstruction, RFA measurements were performed, the values of which showed a significant decrease in the stability quotient of the installed implants in the projection of tooth 4.7 – 61 ISQ and in tooth 4.6 – 74 ISQ. This result made it necessary to revise the rehabilitation plan, adjust the medical prescription sheet and decide to postpone the placement of the all-milled restorative structure on the implants. Conclusions. The RFA study indicates not only the primary stability of implants but also a full-fledged stage-bystage clinical prognosis of the further functionality of a prosthetic structure based on short implants with proper preservation of the morphological functionality of bone tissue.
共振频率分析--下颌骨组织种植后形态的指标
本文描述了共振频率分析(RFA)作为初级种植体稳定性质量指标的作用,这是目标的基础,并通过一个临床病例的结果证实-下颌右侧萎缩骨组织的无牙段种植,考虑其管的地形特征,使用超短(h = 5.5 mm)种植体。的目标。强调共振频率分析在使用短(超短)种植体的手术干预中预测其初级稳定性和保持下颌骨组织的正常形态的作用。材料和方法。采用企鹅仪器(RFA)技术测定7例患者17颗种植体的初始稳定性,测量单位为种植体稳定商(ISQ)。在75 ISQ范围内的值意味着种植体具有足够的初级稳定性,可以接受和分配无牙节段骨组织上的直接负荷。结果。在准备阶段之后,根据患者康复方案,将17个超短(h = 5.5 mm × b = 4.0 mm)种植体放置在下颌无牙节段,没有切口截骨路径。我们使用共振频率分析的方法得到两个短皮质下种植体在4.7 - 92 ISQ和4.6 - 90 ISQ的牙投影中的高初级稳定性值。在上建物固定临床阶段的第3天,进行RFA测量,其值显示安装的种植体在4.7 - 61 ISQ和4.6 - 74 ISQ的牙投影处的稳定商显著下降。因此,有必要修改康复计划,调整医疗处方,并决定推迟全磨修复结构在种植体上的放置。结论。RFA研究不仅表明了种植体的初步稳定性,而且还表明了基于适当保存骨组织形态功能的短种植体的假体结构进一步功能的成熟分期临床预后。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
0.20
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信